SOP Guide for Pharma

BA-BE Studies: SOP for Auto-sampler Stability of Samples – V 2.0

BA-BE Studies: SOP for Auto-sampler Stability of Samples – V 2.0

Standard Operating Procedure for Auto-sampler Stability of Analyte in BA/BE Studies

Department BA-BE Studies
SOP No. SOP/BA-BE/178/2025
Supersedes SOP/BA-BE/178/2022
Page No. Page 1 of 9
Issue Date 17/04/2025
Effective Date 20/04/2025
Review Date 17/04/2026

1. Purpose

To define the process for evaluating the auto-sampler stability of analytes in processed plasma samples stored in the instrument tray during LC-MS/MS runs in bioavailability/bioequivalence (BA/BE) studies.

2. Scope

This SOP is applicable to all validated bioanalytical methods used in BA/BE studies, specifically for assessing the stability of analytes held at auto-sampler temperature (usually 4–10°C) for prolonged durations before injection.

3. Responsibilities

  • Bioanalytical Analyst: Prepares and analyzes auto-sampler stability samples and maintains logs.
  • Supervisor: Ensures compliance with time schedules and evaluates results.
  • QA Officer: Reviews documentation and approves inclusion in validation or study report.

4. Accountability

The Bioanalytical Head is accountable for ensuring the auto-sampler stability studies are conducted according to validation protocols and documented accurately for regulatory compliance.

5. Procedure

5.1 Sample Preparation

  1. Prepare a minimum of six replicates each for Low QC (LQC) and High QC (HQC) samples.
  2. Label each vial clearly with:
    • Sample ID
    • QC Level
    • Analyte Name
    • Preparation Date & Time
    • Auto-sampler Stability
  3. Record sample details in Annexure-1: Auto-sampler Stability Log.

5.2 Exposure in Auto-sampler

  1. Place processed QC samples in the auto-sampler tray set to validated temperature (e.g., 4°C).
  2. Expose samples for a predefined duration (e.g., 24, 48, or 72 hours) as specified in the method validation protocol.
  3. Ensure samples remain capped and undisturbed throughout the holding period.

5.3 Sample Analysis

  1. Inject stability samples along with fresh calibration curve and system suitability QC samples.
  2. Use validated LC-MS/MS method and record analyte concentrations.

5.4 Acceptance Criteria

  1. Mean concentrations of LQC and HQC must be within ±15% of their nominal values.
  2. %CV for each QC level must be ≤15%.
  3. Summarize findings in Annexure-2: Auto-sampler Stability Summary.

5.5 Documentation and Reporting

  1. Compile chromatograms, raw data printouts, and analysis reports.
  2. Prepare and submit a report to QA for approval.
  3. Include the approved report in the study validation or final submission dossier.

6. Abbreviations

  • BA/BE: Bioavailability/Bioequivalence
  • LQC: Low Quality Control
  • HQC: High Quality Control
  • %CV: Percent Coefficient of Variation
  • QA: Quality Assurance

7. Documents

  1. Auto-sampler Stability Log – Annexure-1
  2. Auto-sampler Stability Summary – Annexure-2

8. References

  • ICH M10: Bioanalytical Method Validation
  • US FDA Guidance for Industry: Bioanalytical Method Validation (2022)
  • Internal Method Validation Protocol

9. SOP Version

Version: 2.0

10. Approval Section

Prepared By Checked By Approved By
Signature
Date
Name
Designation
Department

11. Annexures

Annexure-1: Auto-sampler Stability Log

Sample ID QC Level Prep Date Tray Temp Start Time End Time Analyst
BE178-AS-01 LQC 16/04/2025 4°C 10:00 10:00 (18/04/2025) Rajesh Kumar

Annexure-2: Auto-sampler Stability Summary

QC Level Nominal (ng/mL) Mean Measured %CV % Difference Status
LQC 20.0 19.2 4.3 -4.0% Accepted
HQC 1500.0 1496.5 3.1 -0.2% Accepted

Revision History:

Revision Date Revision No. Details Reason Approved By
01/01/2022 1.0 Initial SOP Issuance New SOP QA Head
17/04/2025 2.0 Updated cycle durations, annexures added Method Compliance QA Head