Analytical Method Development: SOP for Preparation and Approval of Working Standards – V 2.0
Standard Operating Procedure for Preparation and Approval of Working Standards in Analytical Method Development
| Department |
Analytical Method Development |
| SOP No. |
SOP/AMD/342/2025 |
| Supersedes |
SOP/AMD/342/2022 |
| Page No. |
Page 1 of 11 |
| Issue Date |
01/06/2025 |
| Effective Date |
03/06/2025 |
| Review Date |
01/06/2026 |
1. Purpose
The purpose of this SOP is to establish a detailed procedure for the preparation, qualification, and approval of working standards used in analytical method development and routine
testing activities within the laboratory.
2. Scope
This SOP is applicable to all working standards derived from certified reference standards, pharmacopeial standards, or in-house standards for use in quality control or analytical method development.
3. Responsibilities
- Analytical Chemist: Prepares, labels, and submits working standard samples for analysis.
- QA Reviewer: Reviews documentation and approves the working standard based on predefined acceptance criteria.
- Documentation Officer: Maintains logs, COA copies, and calibration data for all approved working standards.
4. Accountability
The Head of Analytical Method Development is accountable for the proper implementation of this SOP and ensuring compliance with applicable regulatory and quality requirements for reference material handling.
5. Procedure
5.1 Procurement and Receipt of Reference Standard
- Obtain pharmacopeial reference standards (USP, BP, EP, IP) or certified in-house reference standards.
- Verify COA and expiration date upon receipt.
- Record batch number, potency, and expiry date in Reference Standard Receipt Log (Annexure-1).
5.2 Preparation of Working Standard
- Weigh appropriate quantity of reference standard in a clean and dry container using calibrated balance.
- Mix and homogenize thoroughly to ensure consistency.
- Store the working standard in airtight, labeled amber bottles.
- Label must include:
- Material name
- Batch number
- Prepared by and date
- Potency and storage conditions
5.3 Characterization and Qualification
- Determine the following parameters:
- Identity by IR or UV
- Assay by validated method (minimum 3 replicates)
- Moisture content by KF titration
- Any additional tests as per product-specific requirement
- Calculate potency using assay value and LOD correction, where applicable.
- Enter results in Working Standard Qualification Sheet (Annexure-2).
5.4 Approval and Release
- Submit the qualification sheet and COA to QA for review and approval.
- On approval, assign working standard ID and enter in Working Standard Register (Annexure-3).
- Distribute small aliquots to analysts with usage log and instructions for handling.
5.5 Re-Qualification and Expiry Monitoring
- Working standards should be re-qualified every 6 months or based on usage pattern.
- Record re-qualification data in Annexure-2 under new entry date.
- Destroy expired standards as per SOP for disposal and update the register accordingly.
6. Abbreviations
- COA: Certificate of Analysis
- LOD: Loss on Drying
- KF: Karl Fischer
- USP: United States Pharmacopeia
- IR: Infrared Spectroscopy
7. Documents
- Annexure-1: Reference Standard Receipt Log
- Annexure-2: Working Standard Qualification Sheet
- Annexure-3: Working Standard Register
8. References
- ICH Q6A – Specifications: Test Procedures and Acceptance Criteria
- USP General Chapter <11> – Reference Standards
- WHO TRS 996 – Quality Control of Pharmaceutical Products
9. SOP Version
Version: 2.0
10. Approval Section
|
Prepared By |
Checked By |
Approved By |
| Signature |
|
|
|
| Date |
|
|
|
| Name |
Deepika Sharma |
Neeraj Thakur |
Dr. Rakesh Shetty |
| Designation |
Analytical Chemist |
QA Reviewer |
Head – AMD |
| Department |
Analytical Method Development |
Quality Assurance |
Analytical Method Development |
11. Annexures
Annexure-1: Reference Standard Receipt Log
| Date |
Standard Name |
Source |
Batch No. |
Potency |
Expiry |
Received By |
| 28/05/2025 |
Paracetamol CRS |
USP |
PSM-023 |
99.6% |
31/12/2026 |
D. Sharma |
Annexure-2: Working Standard Qualification Sheet
Includes data on IR spectra, assay result, moisture content, calculated potency, and QA approval signature.
Annexure-3: Working Standard Register
| WS ID |
Material |
Preparation Date |
Potency |
Expiry |
Requalified On |
| WS/AMD/101 |
Paracetamol |
30/05/2025 |
99.4% |
30/11/2025 |
— |
Revision History:
| Revision Date |
Revision No. |
Details |
Reason |
Approved By |
| 01/06/2025 |
2.0 |
Updated requalification interval and moisture analysis requirement |
Annual SOP review |
Dr. Rakesh Shetty |
| 10/08/2022 |
1.0 |
Initial release |
New SOP |
QA Head |