Standard Operating Procedure for UV-Visible Spectrophotometer Calibration in Analytical Method Development
| Department | Analytical Method Development |
|---|---|
| SOP No. | SOP/AMD/333/2025 |
| Supersedes | SOP/AMD/333/2022 |
| Page No. | Page 1 of 14 |
| Issue Date | 01/06/2025 |
| Effective Date | 03/06/2025 |
| Review Date | 01/06/2026 |
1. Purpose
This SOP establishes a standardized method for calibrating the UV-Visible Spectrophotometer to ensure that it functions accurately and consistently, in compliance with cGMP and ICH guidelines.
2. Scope
This procedure applies to all UV-Visible spectrophotometers used in
3. Responsibilities
- Analytical Chemist: Perform calibration as per this SOP and maintain records.
- Instrument Coordinator: Ensure calibration schedule and standards are up-to-date.
- QA Officer: Review and approve calibration results and ensure instrument compliance.
4. Accountability
The Head of Analytical Method Development is accountable for implementation, review, and control of calibration activities for the UV-Visible spectrophotometer.
5. Procedure
5.1 Calibration Frequency
- Calibration should be carried out:
- Upon installation or relocation
- After major repairs or software upgrade
- Routinely at monthly intervals
5.2 Calibration Parameters
- Wavelength Accuracy using Holmium Perchlorate solution
- Photometric Accuracy using Potassium Dichromate solution
- Stray Light using 1.2% Potassium Chloride solution
- Resolution Power using Toluene in Hexane
- Baseline Flatness and Noise Level using empty cell holders
5.3 Preparation of Calibration Standards
Prepare all calibration solutions using Class A glassware and analytical grade reagents as per Annexure-1. Use freshly prepared solutions on the day of calibration.
5.4 Calibration Execution
- Switch on the instrument and allow it to warm up for 30 minutes.
- Set baseline using matched quartz cells filled with distilled water.
- Run individual tests as per the calibration checklist and record observations.
5.5 Acceptance Criteria
- Wavelength Accuracy: ±1.0 nm
- Photometric Accuracy: ±0.010 absorbance units
- Stray Light: Absorbance ≥ 2.0 at 198 nm
- Resolution: Ratio of Absorbance at 269 nm / 266 nm ≥ 1.5
- Baseline Flatness: ≤ ±0.002 Abs
5.6 Data Documentation
- Record results in Calibration Report (Annexure-2).
- Any deviation must be investigated and recorded using Deviation Form (Annexure-3).
5.7 Review and Approval
- Submit calibration report to QA for review.
- Ensure signatures of analyst, reviewer, and approver with date and time.
5.8 Labeling and Status Update
- Affix “CALIBRATED” label with calibration and next due date.
- Update instrument status log and master calibration schedule.
6. Abbreviations
- UV: Ultraviolet
- QA: Quality Assurance
- ICH: International Council for Harmonisation
- GMP: Good Manufacturing Practice
7. Documents
- Annexure-1: Preparation of Calibration Standards
- Annexure-2: Calibration Report
- Annexure-3: Deviation Form
8. References
- ICH Q2(R1) – Validation of Analytical Procedures
- USP <857> – Ultraviolet-Visible Spectroscopy
- 21 CFR Part 211 – Subpart I: Laboratory Controls
9. SOP Version
Version: 2.0
10. Approval Section
| Prepared By | Checked By | Approved By | |
|---|---|---|---|
| Signature | |||
| Date | |||
| Name | Ankita Sharma | Priya Mehra | Dr. Ramesh Iyer |
| Designation | Analyst | QA Executive | Head – AMD |
| Department | Analytical Method Development | Quality Assurance | Analytical Method Development |
11. Annexures
Annexure-1: Calibration Standard Preparations
Detailed procedures for preparation of standard solutions for Holmium Perchlorate, Potassium Dichromate, Potassium Chloride, and Toluene in Hexane.
Annexure-2: UV Calibration Report
| Parameter | Observed Value | Acceptance Criteria | Pass/Fail |
|---|---|---|---|
| Wavelength Accuracy | ±0.6 nm | ±1.0 nm | Pass |
| Photometric Accuracy | ±0.005 AU | ±0.010 AU | Pass |
| Stray Light | 2.1 Abs | ≥2.0 Abs | Pass |
Annexure-3: Deviation Form
Template for deviation reporting, root cause analysis, CAPA, and final disposition of calibration failure.
Revision History:
| Revision Date | Revision No. | Details | Reason | Approved By |
|---|---|---|---|---|
| 01/06/2025 | 2.0 | Updated stray light criteria and annexure formats | Annual Review | Dr. Ramesh Iyer |
| 15/04/2022 | 1.0 | Initial SOP release | New SOP | QA Head |