SOP Guide for Pharma

Analytical Method Development: SOP for Calibration of UV-Visible Spectrophotometer – V 2.0

Analytical Method Development: SOP for Calibration of UV-Visible Spectrophotometer – V 2.0

Standard Operating Procedure for UV-Visible Spectrophotometer Calibration in Analytical Method Development


Department Analytical Method Development
SOP No. SOP/AMD/333/2025
Supersedes SOP/AMD/333/2022
Page No. Page 1 of 14
Issue Date 01/06/2025
Effective Date 03/06/2025
Review Date 01/06/2026

1. Purpose

This SOP establishes a standardized method for calibrating the UV-Visible Spectrophotometer to ensure that it functions accurately and consistently, in compliance with cGMP and ICH guidelines.

2. Scope

This procedure applies to all UV-Visible spectrophotometers used in

the Analytical Method Development laboratory for pharmaceutical API and formulation analysis.

3. Responsibilities

  • Analytical Chemist: Perform calibration as per this SOP and maintain records.
  • Instrument Coordinator: Ensure calibration schedule and standards are up-to-date.
  • QA Officer: Review and approve calibration results and ensure instrument compliance.

4. Accountability

The Head of Analytical Method Development is accountable for implementation, review, and control of calibration activities for the UV-Visible spectrophotometer.

5. Procedure

5.1 Calibration Frequency

  1. Calibration should be carried out:
    • Upon installation or relocation
    • After major repairs or software upgrade
    • Routinely at monthly intervals

5.2 Calibration Parameters

  1. Wavelength Accuracy using Holmium Perchlorate solution
  2. Photometric Accuracy using Potassium Dichromate solution
  3. Stray Light using 1.2% Potassium Chloride solution
  4. Resolution Power using Toluene in Hexane
  5. Baseline Flatness and Noise Level using empty cell holders

5.3 Preparation of Calibration Standards

Prepare all calibration solutions using Class A glassware and analytical grade reagents as per Annexure-1. Use freshly prepared solutions on the day of calibration.

5.4 Calibration Execution

  1. Switch on the instrument and allow it to warm up for 30 minutes.
  2. Set baseline using matched quartz cells filled with distilled water.
  3. Run individual tests as per the calibration checklist and record observations.

5.5 Acceptance Criteria

  • Wavelength Accuracy: ±1.0 nm
  • Photometric Accuracy: ±0.010 absorbance units
  • Stray Light: Absorbance ≥ 2.0 at 198 nm
  • Resolution: Ratio of Absorbance at 269 nm / 266 nm ≥ 1.5
  • Baseline Flatness: ≤ ±0.002 Abs

5.6 Data Documentation

  1. Record results in Calibration Report (Annexure-2).
  2. Any deviation must be investigated and recorded using Deviation Form (Annexure-3).

5.7 Review and Approval

  1. Submit calibration report to QA for review.
  2. Ensure signatures of analyst, reviewer, and approver with date and time.

5.8 Labeling and Status Update

  1. Affix “CALIBRATED” label with calibration and next due date.
  2. Update instrument status log and master calibration schedule.

6. Abbreviations

  • UV: Ultraviolet
  • QA: Quality Assurance
  • ICH: International Council for Harmonisation
  • GMP: Good Manufacturing Practice

7. Documents

  1. Annexure-1: Preparation of Calibration Standards
  2. Annexure-2: Calibration Report
  3. Annexure-3: Deviation Form

8. References

  • ICH Q2(R1) – Validation of Analytical Procedures
  • USP <857> – Ultraviolet-Visible Spectroscopy
  • 21 CFR Part 211 – Subpart I: Laboratory Controls

9. SOP Version

Version: 2.0

10. Approval Section

Prepared By Checked By Approved By
Signature
Date
Name Ankita Sharma Priya Mehra Dr. Ramesh Iyer
Designation Analyst QA Executive Head – AMD
Department Analytical Method Development Quality Assurance Analytical Method Development

11. Annexures

Annexure-1: Calibration Standard Preparations

Detailed procedures for preparation of standard solutions for Holmium Perchlorate, Potassium Dichromate, Potassium Chloride, and Toluene in Hexane.

Annexure-2: UV Calibration Report

Parameter Observed Value Acceptance Criteria Pass/Fail
Wavelength Accuracy ±0.6 nm ±1.0 nm Pass
Photometric Accuracy ±0.005 AU ±0.010 AU Pass
Stray Light 2.1 Abs ≥2.0 Abs Pass

Annexure-3: Deviation Form

Template for deviation reporting, root cause analysis, CAPA, and final disposition of calibration failure.

Revision History:

Revision Date Revision No. Details Reason Approved By
01/06/2025 2.0 Updated stray light criteria and annexure formats Annual Review Dr. Ramesh Iyer
15/04/2022 1.0 Initial SOP release New SOP QA Head
Exit mobile version