1) Purpose

The purpose of this SOP is to establish a standardized procedure for documenting risk management activities for medical devices in compliance with ISO 14971. Proper documentation ensures systematic identification, assessment, control, and monitoring of risks associated with medical devices throughout their lifecycle.

2) Scope

This SOP applies to all activities related to risk management for medical devices, including design, development, manufacturing, post-market monitoring, and process changes. It is relevant to quality assurance, regulatory affairs, product development, and risk management teams.

3) Responsibilities

– Risk Management Team: Conducts risk assessments, implements control measures, and maintains the risk management file.
– Quality Assurance (QA): Reviews and approves risk management documentation to ensure compliance with ISO 14971.
– Regulatory Affairs: Ensures risk management documentation aligns with regulatory requirements for specific markets.
– Product Development Team: Provides input on potential risks during the design and development phases.
– Post-Market Surveillance Team: Monitors field data and updates risk management documentation based on feedback.

4) Procedure

4.1 Initiation of Risk Management Process
4.1.1 Creation of Risk Management Plan
– Develop a Risk Management Plan (RMP) outlining:
– Scope of risk

management activities.
– Roles and responsibilities.
– Risk criteria, including acceptable risk levels.
– Methods for risk assessment, control, and verification.
– Ensure the RMP is reviewed and approved by QA and regulatory affairs.

4.1.2 Risk Management File
– Establish and maintain a Risk Management File (RMF) for each medical device.
– Include all documents related to risk assessment, control measures, and verification activities in the RMF.

4.2 Risk Identification
4.2.1 Hazard Identification
– Identify potential hazards associated with the medical device, including:
– Physical, chemical, and biological hazards.
– Use-related and environmental hazards.
– Regulatory and compliance risks.
– Use brainstorming sessions, historical data, and failure analysis techniques.

4.2.2 Documentation
– Document all identified hazards in the Risk Assessment Worksheet, specifying:
– Hazard description.
– Cause of hazard.
– Associated harm or potential impact.

4.3 Risk Analysis
4.3.1 Severity and Probability
– Assess the severity of potential harm using predefined categories (e.g., minor, serious, critical).
– Estimate the probability of occurrence based on historical data, testing results, or expert judgment.

4.3.2 Risk Evaluation
– Calculate the risk level by combining severity and probability scores.
– Use a risk matrix or scoring system to categorize risks as acceptable, unacceptable, or requiring further evaluation.

4.3.3 Documentation
– Record risk analysis results in the Risk Assessment Worksheet, including risk level and rationale for classification.

4.4 Risk Control
4.4.1 Implementation of Control Measures
– Identify and implement appropriate control measures to reduce risk, such as:
– Inherently safe design.
– Protective measures (e.g., alarms, shields).
– User training and instructions.
– Prioritize control measures based on the effectiveness hierarchy.

4.4.2 Verification of Control Measures
– Verify the effectiveness of implemented control measures through testing, simulations, or audits.
– Record verification results and supporting evidence in the Risk Management File.

4.4.3 Residual Risk Evaluation
– Reassess risks after control measures are implemented to ensure residual risks are within acceptable levels.
– Document residual risk analysis and justification for acceptance.

4.5 Risk Benefit Analysis
4.5.1 Evaluation of Benefits
– Assess the medical device’s benefits, including improved patient outcomes and enhanced safety.
– Compare benefits to residual risks to determine if risks are justified.

4.5.2 Documentation
– Record the risk-benefit analysis in the Risk Management Report, including supporting data and conclusions.

4.6 Risk Monitoring and Review
4.6.1 Post-Market Surveillance
– Monitor post-market data, including complaints, adverse events, and field reports, to identify emerging risks.
– Update the Risk Management File based on new data or feedback.

4.6.2 Periodic Reviews
– Conduct periodic reviews of the Risk Management File to ensure it reflects current risks and control measures.
– Record review findings and updates in the RMF.

4.7 Documentation and Record Keeping
4.7.1 Risk Management Report
– Compile a Risk Management Report summarizing:
– Risk assessments.
– Implemented control measures and their effectiveness.
– Residual risk evaluations and risk-benefit analyses.
– Ensure the report is reviewed and approved by QA and regulatory affairs.

4.7.2 Risk Assessment Worksheet
– Maintain detailed risk assessment worksheets documenting hazard identification, risk analysis, and control measures.

4.7.3 Record Retention
– Retain risk management documentation for the duration specified by regulatory agencies (e.g., the product lifecycle or as mandated).

5) Abbreviations

– ISO: International Organization for Standardization
– RMP: Risk Management Plan
– RMF: Risk Management File
– QA: Quality Assurance
– SOP: Standard Operating Procedure

6) Documents

– Risk Management Plan (RMP)
– Risk Assessment Worksheet
– Risk Management Report
– Post-Market Surveillance Reports
– Verification and Validation Records

7) Reference

– ISO 14971: Application of Risk Management to Medical Devices
– FDA CFR Title 21, Part 820.30: Design Controls
– ISO 13485: Medical devices – Quality management systems

8) SOP Version

– Version: 1.0
– Effective Date: DD/MM/YYYY
– Approved by: [Name/Title]

Annexure

Annexure 1: Risk Assessment Worksheet Template

Hazard ID Description Cause Severity Probability Risk Level Control Measures
H001 Sharp edges Poor manufacturing process Critical Low Moderate Revised process design

Annexure 2: Risk Management Plan Template

Plan Section Details
Scope Risk management activities for Device X
Criteria Acceptable risk levels defined as low to moderate
Methods ISO 14971-compliant risk assessment tools