SOP for Calibration of Nebulizer Pressure Relief Valves

Posted on By

SOP for Calibration of Nebulizer Pressure Relief Valves

Standard Operating Procedure for Calibrating Nebulizer Pressure Relief Valves

1) Purpose

The purpose of this SOP is to define the procedure for calibrating pressure relief valves used in nebulizer systems to ensure their reliability, safety, and compliance with operational standards.

2) Scope

This SOP applies to all pressure relief valves used in nebulizer manufacturing systems within the facility.

3) Responsibilities

Operators: Assist in calibration setup and document readings.
Maintenance Team: Perform calibration and ensure proper functioning of the valves.
Quality Assurance (QA): Validate calibration results and approve calibration records.
Supervisors: Ensure compliance with this SOP and oversee calibration activities.

4) Procedure

4.1 Pre-Calibration Checks

  • Ensure the calibration equipment, such as pressure gauges and reference valves, is within its calibration validity period.
  • Inspect the pressure relief valve for physical damage, dirt, or wear.
  • Verify that the valve is correctly installed in the system and isolated for calibration purposes.
  • Document the valve details, such as model and serial number, in the Calibration Log.
See also  SOP for Maintenance of Air Filtration Systems in Nebulizer Manufacturing Areas

4.2 Calibration Setup

  • Connect the pressure relief valve to the calibration test bench or setup as specified in the manufacturer’s guidelines.
  • Attach the pressure source to the valve inlet and the pressure gauge to monitor the applied pressure.
  • Ensure all connections
are secure and free of leaks.

4.3 Calibration Procedure

  • Gradually increase the pressure using the test setup until the valve begins to release pressure (the set-point).
  • Record the pressure at which the valve opens and compare it with the manufacturer’s specifications.
  • Adjust the valve set-point if it deviates from the acceptable range using the adjustment mechanism provided by the manufacturer.
  • Repeat the process three times to confirm consistent performance.
  • Document the results of each trial in the Calibration Report.

4.4 Post-Calibration Activities

  • Reinstall the calibrated pressure relief valve onto the nebulizer system.
  • Label the valve with the calibration status, date, and next due date.
  • Submit the calibration records to QA for validation and approval.
  • Document the calibration activity in the Calibration Log.

4.5 Frequency of Calibration

  • Routine Calibration: Annually or as per the manufacturer’s recommendations.
  • Post-Maintenance Calibration: After repairs, replacements, or major adjustments.

4.6 Handling Calibration Failures

  • Isolate the valve and report the issue to the maintenance team if calibration fails.
  • Document the failure details in the Calibration Failure Report and notify QA.
  • Replace or repair faulty valves and repeat the calibration process.

4.7 Safety Precautions

  • Wear appropriate personal protective equipment (PPE), including gloves and safety goggles, during calibration activities.
  • Handle pressurized systems carefully to avoid accidents or equipment damage.
  • Ensure proper grounding and follow lockout/tagout procedures where necessary.

5) Abbreviations

  • QA: Quality Assurance
  • PPE: Personal Protective Equipment

6) Documents

The following documents should be maintained:

  • Calibration Log
  • Calibration Report
  • Calibration Failure Report

7) References

Relevant regulatory guidelines and references include:

  • ISO 13485: Medical Devices Quality Management Systems
  • Good Manufacturing Practices (GMP) Guidelines
  • Equipment Manufacturer’s Manual

8) SOP Version

Version: 1.0

Annexure

Annexure Title: Calibration Log

 
Date Valve ID Set-Point Pressure Measured Pressure Performed By Remarks
DD/MM/YYYY Valve Identifier Specified Value Measured Value Technician Name Details of calibration
           

Annexure Title: Calibration Report

 
Date Valve ID Pre-Calibration Reading Post-Calibration Reading Performed By QA Approval
DD/MM/YYYY Valve Identifier Pre-Calibration Value Post-Calibration Value Technician Name QA Name
           

Annexure Title: Calibration Failure Report

 
Date Valve ID Issue Identified Corrective Action Performed By QA Approval
DD/MM/YYYY Valve Identifier Details of issue Details of corrective action Technician Name QA Name
           
See also  SOP for Use of Automated Nebulizer Inspection Systems
Nebulizers Tags:Nebulizer calibration steps, Nebulizer cleaning protocol, Nebulizer compliance standards, Nebulizer disinfection procedure, Nebulizer equipment inspection, Nebulizer handling procedures, Nebulizer infection control, Nebulizer maintenance guidelines, Nebulizer medication administration, Nebulizer operation manual, Nebulizer operational checklist, Nebulizer parts cleaning, Nebulizer patient instructions, Nebulizer performance testing, Nebulizer preventive maintenance, Nebulizer quality assurance, Nebulizer safety precautions, Nebulizer setup instructions, Nebulizer SOP, Nebulizer standard operating procedure, Nebulizer sterilization process, Nebulizer storage guidelines, Nebulizer troubleshooting steps, Nebulizer usage instructions, Nebulizer user training

Post navigation

Previous Post: SOP for In Silico Docking Studies
Next Post: SOP for QSAR Modeling in Drug Discovery

Standard Operating Procedures V 1.0

  • Aerosols
  • Analytical Method Development
  • Bioequivalence Bioavailability Study
  • Capsule Formulation
  • Clinical Studies
  • Creams
  • Data Integrity
  • Dental Dosage Forms
  • Drug Discovery
  • Environment, Health and Safety
  • Formulation Development
  • Gels
  • Good Distribution Practice
  • Good Warehousing Practices
  • In-Process Control
  • Injectables
  • Liquid Orals
  • Liposome and Emulsion Formulations
  • Lotions
  • Lyophilized Products
  • Maintenance Dept.
  • Medical Devices
  • Metered-Dose Inhaler
  • Microbiology Testing
  • Nanoparticle Formulation
  • Nasal Spray Formulations
  • Nebulizers
  • Ocular (Eye) Dosage Forms
  • Ointments
  • Otic (Ear) Dosage Forms
  • Pharmacovigilance
  • Powder & Granules
  • Purchase Departments
  • Quality Assurance
  • Quality Control
  • Raw Material Stores
  • Regulatory Affairs
  • Tablet Manufacturing
  • Rectal Dosage Forms
  • Transdermal Patches
  • Vaginal Dosage Forms
  • Validations and Qualifications

Read SOPs in your Language:

 - 
Bengali
 - 
bn
English
 - 
en
Gujarati
 - 
gu
Hindi
 - 
hi
Malayalam
 - 
ml
Marathi
 - 
mr
Punjabi
 - 
pa
Tamil
 - 
ta
Telugu
 - 
te

NEW! Revised SOPs – V 2.0

  • Aerosols V 2.0
  • Analytical Method Development V 2.0
  • API Manufacturing V 2.0
  • BA-BE Studies V 2.0
  • Biosimilars V 2.0
  • Capsules V 2.0
  • Creams V 2.0
  • Elixers V 2.0
  • Ointments V 2.0
  • Raw Material Warehouse V 2.0
  • Tablet Manufacturing V2.0

New Publication: A must for All.

Copyright © 2025 SOP Guide for Pharma.

Powered by PressBook WordPress theme

Go to mobile version