SOP for Calibration of Conductivity Meters

Posted on By

SOP for Calibration of Conductivity Meters

Standard Operating Procedure for Calibration of Conductivity Meters

1) Purpose

The purpose of this SOP is to establish the procedures for calibrating conductivity meters to ensure accuracy and reliability in conductivity measurements as per regulatory requirements in the pharmaceutical industry.

2) Scope

This SOP applies to all conductivity meters used in the pharmaceutical manufacturing process.

3) Responsibilities

The Quality Assurance Department is responsible for implementing and adhering to this SOP. All personnel involved in the calibration process must be trained and qualified.

4) Procedure

  1. Calibration Schedule Establishment
    1. Develop a calibration schedule based on meter criticality, manufacturer recommendations, and regulatory requirements.
    2. Assign unique identification numbers or labels to each conductivity meter for traceability.
    3. Document calibration frequency and tolerance limits for each conductivity meter.
  2. Calibration Process
    1. Retrieve the conductivity meter to be calibrated from its designated location.
    2. Verify the meter’s identity and check its condition for any physical damage.
    3. Prepare calibration standards of known conductivity values.
    4. Calibrate the meter using certified calibration standards and procedures.
    5. Record calibration results including before and after values, adjustments made, and calibration date.
    6. Apply calibration labels indicating next due date and calibration status.
  3. Out-of-Tolerance Conditions
    1. If a conductivity meter is found out-of-tolerance, quarantine it immediately and perform investigation to determine impact.
    2. Take corrective action, recalibrate if possible, or arrange for repair by authorized personnel.
    3. Document all actions taken, including investigation findings and corrective measures.
  4. Record Keeping
    1. Maintain accurate records of all calibration activities, including calibration certificates, calibration history, and maintenance logs.
    2. Archive records for easy retrieval during audits or inspections.
See also  SOP for Maintenance of Stability Chambers

5) Abbreviations, if any

None

6) Documents, if any

Calibration Certificates, Calibration Logs, Maintenance Records

7) Reference, if any

Relevant regulatory guidelines such as FDA, EMA, and WHO GMP requirements for calibration of conductivity meters.

8) SOP Version

Version 1.0

Maintenance Dept. Tags:Balance scale calibration, Cleanroom maintenance SOP, Compressed air system maintenance, Conductivity meter calibration, Corrective maintenance procedures, Electrical system maintenance SOP, Emergency maintenance procedures, Equipment calibration SOP, Fire alarm system maintenance, Flow meter calibration SOP, HPLC GC instrument maintenance, HVAC system maintenance SOP, IQ OQ PQ validation SOP, Laboratory equipment maintenance, Maintenance schedule planning, Measuring instruments calibration, Packaging equipment maintenance, pH meter calibration SOP, Pressure gauge calibration SOP, Preventive maintenance SOP, Production equipment maintenance, Security system maintenance SOP, SOP pharmaceutical industry, SOP template HTML, Spare parts inventory management SOP, Spectrophotometer calibration, Temperature sensor calibration, UPS power backup maintenance, Water system maintenance SOP, Work order management SOP

Post navigation

Previous Post: SOP for IT Systems Used in Procurement
Next Post: Quality Assurance: SOP for Out of Specification (OOS) Handling in Gels Production

Standard Operating Procedures V 1.0

  • Aerosols
  • Analytical Method Development
  • Bioequivalence Bioavailability Study
  • Capsule Formulation
  • Clinical Studies
  • Creams
  • Data Integrity
  • Dental Dosage Forms
  • Drug Discovery
  • Environment, Health and Safety
  • Formulation Development
  • Gels
  • Good Distribution Practice
  • Good Warehousing Practices
  • In-Process Control
  • Injectables
  • Liquid Orals
  • Liposome and Emulsion Formulations
  • Lotions
  • Lyophilized Products
  • Maintenance Dept.
  • Medical Devices
  • Metered-Dose Inhaler
  • Microbiology Testing
  • Nanoparticle Formulation
  • Nasal Spray Formulations
  • Nebulizers
  • Ocular (Eye) Dosage Forms
  • Ointments
  • Otic (Ear) Dosage Forms
  • Pharmacovigilance
  • Powder & Granules
  • Purchase Departments
  • Quality Assurance
  • Quality Control
  • Raw Material Stores
  • Regulatory Affairs
  • Tablet Manufacturing
  • Rectal Dosage Forms
  • Transdermal Patches
  • Vaginal Dosage Forms
  • Validations and Qualifications

Read SOPs in your Language:

 - 
Bengali
 - 
bn
English
 - 
en
Gujarati
 - 
gu
Hindi
 - 
hi
Malayalam
 - 
ml
Marathi
 - 
mr
Punjabi
 - 
pa
Tamil
 - 
ta
Telugu
 - 
te

NEW! Revised SOPs – V 2.0

  • Aerosols V 2.0
  • Analytical Method Development V 2.0
  • API Manufacturing V 2.0
  • BA-BE Studies V 2.0
  • Biosimilars V 2.0
  • Capsules V 2.0
  • Creams V 2.0
  • Elixers V 2.0
  • Ointments V 2.0
  • Raw Material Warehouse V 2.0
  • Tablet Manufacturing V2.0

New Publication: A must for All.

Copyright © 2025 SOP Guide for Pharma.

Powered by PressBook WordPress theme

Go to mobile version