Standard Operating Procedure for Temperature and Humidity Monitoring during Production


Department Manufacturing
SOP No. SOP/MFG/177/2025
Supersedes SOP/MFG/177/2022
Page No. Page 1 of 13
Issue Date 24/06/2025
Effective Date 25/06/2025
Review Date 24/06/2026

1. Purpose

To lay down a procedure for monitoring and controlling temperature and humidity in sterile and non-sterile manufacturing areas to ensure compliance with cGMP, product stability, and personnel comfort.

2. Scope

This SOP applies to all production and packaging areas, including Grade A/B/C/D zones, solution preparation rooms, filling

areas, quarantine zones, and warehouses involved in sterile injectable manufacturing.

3. Responsibilities

  • Production Operator: Monitors readings and reports deviations.
  • QA Personnel: Verifies environmental log data and evaluates trends.
  • Engineering Team: Ensures calibration and maintenance of HVAC and monitoring devices.

4. Accountability

The Head of Manufacturing is accountable for ensuring consistent environmental monitoring and timely resolution of excursions beyond defined limits.

5. Procedure

5.1 Monitoring Devices

  1. Use calibrated digital temperature and humidity monitors or validated data loggers.
  2. Ensure sensors are installed at representative locations (workstations, equipment areas, corners).
  3. Calibration must be performed every 6 months or as per the calibration schedule.

5.2 Monitoring Frequency

  1. Record temperature and humidity readings at:
    • Beginning of shift
    • Every 4 hours during operation
    • End of shift
  2. For automated systems, data should be stored electronically with periodic printouts for record-keeping.
  3. Manual entries shall be made in the Environmental Monitoring Log (Annexure-1).

5.3 Acceptable Ranges

  1. Grade A/B: Temperature 18–22°C, Relative Humidity (RH) 40–60%
  2. Grade C/D: Temperature 20–25°C, RH 40–65%
  3. Warehouse/Storage: 25°C ± 2°C, RH NMT 60%

5.4 Action on Deviations

  1. If temperature or humidity exceeds action limits:
    • Immediately inform the engineering department.
    • Initiate deviation report and document in Annexure-2.
    • Stop critical operations if environment affects product quality.
  2. Resume operation only after conditions return within acceptable limits and QA authorization is obtained.

5.5 Trend Review and Reporting

  1. QA shall compile environmental trends monthly for each area.
  2. Unusual spikes or drift patterns must be investigated.
  3. Summarized reports shall be included in monthly GMP review meetings.

5.6 Preventive Maintenance

  1. Engineering shall service air handling units (AHUs), dehumidifiers, and chillers quarterly.
  2. Logs of HVAC equipment maintenance shall be reviewed monthly.

5.7 Data Integrity and Record Retention

  1. Ensure that data (manual or electronic) is legible, contemporaneous, attributable, and permanent.
  2. Records must be retained for not less than five years from date of batch release.

6. Abbreviations

  • SOP: Standard Operating Procedure
  • RH: Relative Humidity
  • QA: Quality Assurance
  • HVAC: Heating, Ventilation and Air Conditioning
  • AHU: Air Handling Unit
  • GMP: Good Manufacturing Practices

7. Documents

  1. Environmental Monitoring Log – Annexure-1
  2. Temperature/Humidity Deviation Log – Annexure-2
  3. Preventive Maintenance Checklist – Annexure-3

8. References

  • WHO TRS 961 Annex 6 – GMP for sterile pharmaceutical products
  • EU GMP Annex 1 – Manufacture of Sterile Medicinal Products
  • 21 CFR Part 211 – cGMP for Finished Pharmaceuticals

9. SOP Version

Version: 2.0

10. Approval Section

Prepared By Checked By Approved By
Signature
Date
Name
Designation Manufacturing Executive QA Executive Head QA
Department Production Quality Assurance Quality Assurance

11. Annexures

Annexure-1: Environmental Monitoring Log

Date Area Time Temperature (°C) Humidity (%) Operator
24/06/2025 Filling Room 08:00 AM 20.5 50 Rajesh Kumar

Annexure-2: Temperature/Humidity Deviation Log

Date Area Observed Value Limit Action Taken QA Sign
22/06/2025 Solution Room 26°C 25°C Notified Engineering; Temperature restored Sunita Reddy

Annexure-3: Preventive Maintenance Checklist

Date Equipment Maintenance Activity Status Engineer
15/06/2025 AHU-1 Filter Change, Coil Cleaning Completed Vinay Pawar

Revision History:

Revision Date Revision No. Details Reason Approved By
01/01/2022 1.0 Initial Release New SOP QA Head
24/06/2025 2.0 Updated HVAC reference and annexures Annual Review QA Head