Gel Manufacturing: SOP for Operation of pH Meters During Gel Production – V 2.0

Gel Manufacturing: SOP for Operation of pH Meters During Gel Production – V 2.0

Standard Operating Procedure for Using pH Meters in Gel Manufacturing Process


Department Gel Manufacturing
SOP No. SOP/GM/044/2025
Supersedes SOP/GM/044/2022
Page No. Page 1 of 12
Issue Date 02/06/2025
Effective Date 04/06/2025
Review Date 02/06/2026

1. Purpose

The purpose of this Standard Operating Procedure (SOP) is to establish a standardized method for operating pH meters during the manufacturing of gel formulations. The SOP ensures accuracy in pH measurement, prevents equipment misuse, and ensures compliance with cGMP guidelines.

2.

Scope

This SOP applies to all personnel involved in using, calibrating, and maintaining pH meters for in-process and final product testing during gel manufacturing activities in the pharmaceutical plant.

3. Responsibilities

  • Production Operator: Performs pH measurements and basic care of the instrument.
  • Production Supervisor: Ensures adherence to SOP and reviews records.
  • QA Executive: Performs periodic checks, reviews calibration records, and verifies compliance.
  • Engineering/Maintenance Team: Responsible for troubleshooting, repairs, and preventive maintenance.

4. Accountability

The Head – Manufacturing shall ensure implementation, training, and strict adherence to this SOP.

5. Procedure

5.1 General Precautions

  • Handle pH electrodes with care; avoid dropping or applying force.
  • Use gloves while handling calibration solutions and gel samples.
  • Calibrate the pH meter daily or before every batch measurement.
  • Ensure that the meter and electrode are properly cleaned before use.

5.2 Pre-Operational Checks

  1. Verify that the pH meter is in “Ready to Use” status.
  2. Ensure buffer solutions (typically pH 4.0, 7.0, and 9.2) are not expired and are at room temperature.
  3. Check for any visible damage or residue on the electrode.

5.3 Calibration Procedure

  1. Switch on the pH meter and allow it to stabilize for 5 minutes.
  2. Rinse the electrode with distilled water and blot gently with a lint-free tissue.
  3. Immerse the electrode into pH 7.0 buffer solution. Allow the reading to stabilize and press “Calibrate”.
  4. Repeat the above steps using pH 4.0 and 9.2 buffer solutions for 2-point or 3-point calibration.
  5. Record calibration details in Annexure-1.

5.4 pH Measurement of Gel Samples

  1. Rinse the electrode with distilled water.
  2. Immerse the electrode in the gel sample in a clean beaker with adequate depth.
  3. Stir the sample gently to avoid entrapment of air bubbles.
  4. Allow the reading to stabilize (usually 30–60 seconds).
  5. Record the pH value in the Batch Manufacturing Record (BMR) and Annexure-2.
  6. Rinse the electrode thoroughly between samples to avoid cross-contamination.

5.5 Post-Measurement Care

  1. Clean the electrode with distilled water and store it in electrode storage solution.
  2. Do not store electrodes dry or submerged in distilled water.
  3. Switch off the pH meter when not in use and cover it with a dust shield.

5.6 Equipment Maintenance

  • Perform monthly maintenance including inspection of cables and connectors.
  • Replace the electrode every 6–12 months or as recommended by the manufacturer.
  • Record maintenance details in equipment logbook.

6. Abbreviations

  • BMR: Batch Manufacturing Record
  • QA: Quality Assurance
  • SOP: Standard Operating Procedure
  • PPE: Personal Protective Equipment

7. Documents

  1. pH Meter Calibration Log – Annexure-1
  2. pH Testing Record – Annexure-2
  3. Equipment Cleaning Log
  4. Batch Manufacturing Record (BMR)

8. References

  • ICH Q7: Good Manufacturing Practice Guide for Active Pharmaceutical Ingredients
  • Manufacturer’s Manual for Digital pH Meter

9. SOP Version

Version: 2.0

10. Approval Section

Prepared By Checked By Approved By
Signature
Date
Name
Designation Jr. Production Chemist QA Executive Head – Manufacturing
Department Gel Manufacturing Quality Assurance Manufacturing

11. Annexures

Annexure-1: pH Meter Calibration Log

Date Equipment ID Buffer pH Used Observed Value Calibrated By Remarks
pH 4.0 / 7.0 / 9.2

Annexure-2: pH Testing Record

Date Batch No. Product Name Sample Type (In-Process/Final) pH Observed Tested By Remarks

Revision History:

Revision Date Revision No. Details Reason Approved By
01/05/2022 1.0 Initial version New SOP QA Head
02/06/2025 2.0 Expanded procedure and annexure details Periodic Review QA Head