SOP Guide for Pharma

Elixir Department: SOP for Water Content Determination (Karl Fischer) – V 2.0

Elixir Department: SOP for Water Content Determination (Karl Fischer) – V 2.0

Standard Operating Procedure for Water Content Determination in Elixirs Using Karl Fischer Titration

Department Elixir Department
SOP No. SOP/ELX/228/2025
Supersedes SOP/ELX/228/2022
Page No. Page 1 of 8
Issue Date 11/04/2025
Effective Date 15/04/2025
Review Date 11/04/2026

1. Purpose

To establish a standard procedure for the determination of water content in elixir formulations using the Karl Fischer (KF) titration method, ensuring accurate and consistent moisture analysis in accordance with pharmacopeial and regulatory standards.

2. Scope

This SOP is applicable to the Quality Control Department for routine testing of water content in elixirs during release and stability testing using volumetric or coulometric KF titration.

3. Responsibilities

  • QC Analyst:
    • Prepare and analyze samples using the Karl Fischer method.
    • Ensure all data is accurately recorded in the logbook and BMR.
  • QC Supervisor:
    • Verify calibration status of the KF instrument and review results.
  • QA Officer:
    • Ensure test compliance with specifications and regulatory requirements.

4. Accountability

The QC Head is accountable for execution and validation of the KF analysis. The QA Head is responsible for ensuring regulatory compliance and final disposition based on moisture

results.

5. Procedure

5.1 Equipment and Materials

  • Karl Fischer Titrator (volumetric or coulometric type)
  • KF reagent (methanol-based or compatible with solvent system)
  • KF oven (for samples with interference, if required)
  • Syringe with needle for injection
  • Analytical balance (calibrated)

5.2 System Setup and Calibration

  1. Ensure the instrument is calibrated using sodium tartrate dihydrate or water standards.
  2. Check the reagent factor prior to each batch analysis.
  3. Perform blank titration and set to zero before analysis.

5.3 Sample Preparation

  1. Weigh accurately about 1–2 g of elixir formulation into a clean, dry syringe.
  2. Inject directly into the KF cell if compatible. For samples with high alcohol or sugar content, use KF oven technique as required.

5.4 Titration Procedure

  1. Inject sample and start the titration automatically.
  2. Allow the endpoint to be detected automatically.
  3. Record the amount of water detected in mg or percentage.
  4. Repeat in duplicate. Report the average value.

5.5 Calculations

% Water Content = (mg H2O / Sample Weight in mg) × 100

5.6 Acceptance Criteria

  1. The water content must fall within the specified limit in the product specification (e.g., NMT 5.0%).
  2. Any value beyond specification should trigger an investigation.

5.7 Documentation

  1. Record all readings in the Water Content Test Logbook (Annexure-1).
  2. Attach titration printouts and maintain with the respective BMR.
  3. Enter KF reagent batch number, expiry, and calibration details.

6. Abbreviations

  • SOP: Standard Operating Procedure
  • KF: Karl Fischer
  • QC: Quality Control
  • QA: Quality Assurance
  • BMR: Batch Manufacturing Record

7. Documents

  1. Water Content Test Logbook (Annexure-1)
  2. Instrument Calibration Log
  3. Batch Manufacturing Record (BMR)

8. References

  • USP Chapter <921> – Water Determination
  • Ph. Eur. 2.5.32 – Water: Karl Fischer Method
  • ICH Q2(R1) – Validation of Analytical Procedures

9. SOP Version

Version: 2.0

10. Approval Section

Prepared By Checked By Approved By
Signature
Date
Name
Designation
Department

11. Annexures

Annexure-1: Water Content Test Logbook

Date Batch No. Sample Weight (mg) Water Found (mg) % Water Reagent Batch No. Tested By Verified By
11/04/2025 ELX-0425-046 1000 34.2 3.42 KF-9823 Rajesh Kumar Sunita Reddy

Revision History:

Revision Date Revision No. Revision Details Reason for Revision Approved By
01/01/2024 1.0 Initial SOP Release New SOP Creation QA Head
11/04/2025 2.0 Included oven method and calculation clarity Method Enhancement QA Head
Exit mobile version