Standard Operating Procedure for Calibration of Balances Used in Material Dispensing
Department | Elixir Department |
---|---|
SOP No. | SOP/ELX/033/2025 |
Supersedes | SOP/ELX/033/2022 |
Page No. | Page 1 of 8 |
Issue Date | 11/04/2025 |
Effective Date | 15/04/2025 |
Review Date | 11/04/2026 |
1. Purpose
To define a systematic procedure for the calibration of balances used during the dispensing of raw materials and excipients in the Elixir Department. This ensures weighing accuracy, minimizes errors, and maintains compliance with GMP.
2. Scope
This SOP applies to all analytical, precision, and top-loading balances used in the dispensing areas (including laminar airflow cabinets and weighing rooms) of the Elixir Department.
3. Responsibilities
- Dispensing Operator:
- Perform daily calibration checks before start of operation.
- Record calibration data and report deviations if any.
- Engineering Department:
- Conduct periodic preventive maintenance and external calibration.
- QA Department:
- Verify calibration logs and approve equipment for use.
4. Accountability
The Head of the Elixir Department is accountable for ensuring that all balances are calibrated, documented, and maintained in a validated state as per GMP.
5. Procedure
5.1 Types
of Calibration
- Daily Calibration: Performed by the operator using certified standard weights.
- Monthly Internal Calibration: Conducted by the Engineering team.
- Annual External Calibration: Performed by an external authorized calibration agency.
5.2 Daily Calibration Procedure
- Switch on the balance and allow it to warm up for at least 30 minutes.
- Ensure the balance is clean, levelled, and placed on a vibration-free surface.
- Use calibrated standard weights (Class F2 or better) with valid calibration certificates.
- Perform calibration at two points:
- Low range (e.g., 10 g)
- High range (e.g., 500 g or 1000 g depending on balance capacity)
- Record the observed reading and compare it against the standard weight. Acceptable tolerance: ±0.1% or as specified in the balance SOP.
- If reading is outside the tolerance limit, report immediately to the Engineering Department and QA.
5.3 Monthly and Annual Calibration
- Engineering shall perform internal calibration checks and adjust balance if deviation is observed.
- Annual calibration must be done by a NABL-accredited agency and traceable to national standards.
- Calibration certificates must be filed and referenced in the equipment qualification record.
5.4 Labeling and Equipment Status
- All balances must have a calibration status label affixed indicating:
- Equipment ID
- Last Calibration Date
- Next Calibration Due Date
- Engineer Initials
- Balances without valid calibration label must not be used and should be tagged “Under Maintenance.”
5.5 Documentation and Deviation Management
- Enter daily calibration results in the Balance Calibration Log (Annexure-1).
- Attach calibration certificates in the Equipment Qualification File.
- Report any deviation or failure to QA using a deviation report form. QA will investigate and approve further action.
6. Abbreviations
- QA: Quality Assurance
- SOP: Standard Operating Procedure
- NABL: National Accreditation Board for Testing and Calibration Laboratories
- GMP: Good Manufacturing Practice
7. Documents
- Balance Calibration Log (Annexure-1)
- Calibration Certificate (Annexure-2)
- Deviation Report Form (Annexure-3)
8. References
- ICH Q7: GMP for Active Pharmaceutical Ingredients
- 21 CFR Part 211 – Current Good Manufacturing Practice
- IS/ISO/IEC 17025 – Calibration and Testing Laboratories
9. SOP Version
Version: 2.0
10. Approval Section
Prepared By | Checked By | Approved By | |
---|---|---|---|
Signature | |||
Date | |||
Name | |||
Designation | |||
Department |
11. Annexures
Annexure-1: Balance Calibration Log
Date | Balance ID | Weight Used (g) | Observed Value (g) | Deviation (%) | Operator |
---|---|---|---|---|---|
11/04/2025 | BLX-01 | 500.00 | 499.85 | 0.03% | Rajesh Kumar |
Annexure-2: Calibration Certificate
Balance ID | Calibrated By | Date | Valid Till | Certificate No. |
---|---|---|---|---|
BLX-01 | Precision Cal Lab | 01/01/2025 | 31/12/2025 | PCL/025/2025 |
Annexure-3: Deviation Report Form
Date | Equipment ID | Observed Issue | Reported By | Action Taken | QA Approval |
---|---|---|---|---|---|
10/04/2025 | BLX-02 | Weight deviation > 0.5% | Sunita Reddy | Sent for recalibration | Approved by QA |
Revision History:
Revision Date | Revision No. | Revision Details | Reason for Revision | Approved By |
---|---|---|---|---|
01/01/2024 | 1.0 | Initial SOP Release | New Implementation | QA Head |
11/04/2025 | 2.0 | Revised Format, Added External Calibration Guidelines | Standardization & Compliance | QA Head |