Biosimilars: SOP for pH Monitoring and Adjustment in Bioreactor Operations – V 2.0
Standard Operating Procedure for pH Monitoring and Adjustment in Bioreactor Operations for Biosimilars
Department |
Biosimilars |
SOP No. |
SOP/BS/069/2025 |
Supersedes |
SOP/BS/069/2022 |
Page No. |
Page 1 of 12 |
Issue Date |
04/05/2025 |
Effective Date |
06/05/2025 |
Review Date |
04/05/2026 |
1. Purpose
To provide a standardized procedure for monitoring and adjusting pH in bioreactors during biosimilar manufacturing processes, ensuring optimal cell growth conditions and GMP compliance.
2. Scope
This SOP applies to all production bioreactors used in upstream processing of biosimilars, including stainless steel and single-use systems ranging from 10L to 2000L.
3. Responsibilities
- Process Operators: Monitor pH in real time and perform adjustments as per protocol.
- Maintenance Team: Calibrate pH probes and ensure operational status of pH control loops.
- QA Personnel: Review logs and verify adherence to validated pH ranges.
4. Accountability
The Head of Upstream Production is accountable for maintaining pH control within the defined range and ensuring appropriate documentation and calibration status.
5. Procedure
5.1 Calibration of pH Probes
- Calibrate pH probes using 2-point calibration with certified buffer solutions (pH 4.0 and 7.0) prior to batch start (Annexure-1).
- Ensure calibration is logged and probe tags match bioreactor configuration.
5.2 Setpoint Configuration
- Set desired pH range on the bioreactor control panel (typically 6.8–7.2 for mammalian cells).
- Verify PID settings for acid/base addition systems are in automatic mode.
5.3 Real-Time pH Monitoring
- Monitor pH via bioreactor software interface every 2 hours.
- Log actual values, trends, and any deviation in the pH Monitoring Log (Annexure-2).
- If automated system is offline, perform manual sampling and check with handheld pH meter.
5.4 pH Adjustment Procedure
- If pH falls below setpoint:
- Use sterile 1N NaOH for upward correction through automated or manual ports.
- If pH exceeds setpoint:
- Use sterile 1N HCl or CO2 sparging for downward correction.
- Adjust slowly and monitor impact in 10-minute intervals to avoid overshoot.
5.5 Alarm Management
- In case of pH deviation beyond ±0.2 units:
- Trigger alarm and initiate deviation handling (SOP/BS/050/2025).
- Verify if probe fouling or drift occurred; perform re-calibration if required.
5.6 End-of-Run Procedure
- Record final pH and probe status in End-of-Run Summary (Annexure-3).
- Clean pH sensor as per SOP/BS/098/2025 and store in pH 4.0 buffer.
6. Abbreviations
- PID: Proportional-Integral-Derivative
- QA: Quality Assurance
- GMP: Good Manufacturing Practice
7. Documents
- pH Probe Calibration Log – Annexure-1
- pH Monitoring Log Sheet – Annexure-2
- End-of-Run pH Summary – Annexure-3
8. References
- ICH Q7 – Good Manufacturing Practice for Active Pharmaceutical Ingredients
- WHO TRS 1010 – Guidelines on Biotherapeutics
- SOP/BS/050/2025 – Deviation Handling
- SOP/BS/098/2025 – pH Sensor Maintenance
9. SOP Version
Version: 2.0
10. Approval Section
|
Prepared By |
Checked By |
Approved By |
Signature |
|
|
|
Date |
|
|
|
Name |
|
|
|
Designation |
|
|
|
Department |
|
|
|
11. Annexures
Annexure-1: pH Probe Calibration Log
Probe ID |
Calibration Date |
Buffer Used |
Calibrated By |
PH-BR-01 |
03/05/2025 |
pH 4.0 & 7.0 |
Sunita Reddy |
Annexure-2: pH Monitoring Log Sheet
Date |
Time |
Setpoint |
Actual pH |
Remarks |
04/05/2025 |
12:00 |
7.0 |
7.02 |
Within limit |
Annexure-3: End-of-Run pH Summary
Batch ID |
Final pH |
Deviation Noted |
Corrective Action |
BR-109 |
6.98 |
No |
N/A |
Revision History:
Revision Date |
Revision No. |
Revision Details |
Reason for Revision |
Approved By |
04/05/2025 |
2.0 |
Added Annexure-3 and CO₂ adjustment for high pH |
Process Optimization |
|