Biosimilars: SOP for Chromatogram Review and Analysis – V 2.0
Standard Operating Procedure for Chromatogram Review and Analysis in Biosimilar Manufacturing
| Department |
Biosimilars |
| SOP No. |
SOP/BS/189/2025 |
| Supersedes |
SOP/BS/189/2022 |
| Page No. |
Page 1 of 9 |
| Issue Date |
04/05/2025 |
| Effective Date |
06/05/2025 |
| Review Date |
04/05/2026 |
1. Purpose
To establish a standard procedure for reviewing and analyzing chromatograms obtained from biosimilar purification processes and quality control testing, ensuring accurate interpretation and GMP-compliant data reporting.
2. Scope
This SOP applies to all chromatographic data generated from High-Performance Liquid Chromatography (HPLC), Ultra-High Performance Liquid Chromatography (UHPLC), and other related techniques used in biosimilar manufacturing and QC operations.
3. Responsibilities
- QC Analyst: Review chromatograms, integrate peaks, and assess system suitability per method validation criteria.
- QA Reviewer: Cross-check reviewed chromatograms for compliance with specifications and approve or reject based on results.
- Instrumentation Specialist: Maintain chromatographic systems and verify calibration status before use.
4. Accountability
The Quality Control Manager is accountable for ensuring that chromatographic data are reviewed in a timely, accurate, and traceable manner in compliance with GMP and data integrity guidelines.
5. Procedure
5.1 Pre-Review Checklist
- Ensure chromatographic system is qualified and in calibrated state.
- Confirm analytical method number, SOP reference, column type, and mobile phase batch number.
- Verify analyst, date, and sample ID recorded in the raw data file.
5.2 System Suitability Check
- Review system suitability chromatogram for parameters such as:
- Resolution between peaks
- Tailing factor (should be < 2.0)
- Theoretical plates (N)
- Retention time consistency
- If suitability fails, do not proceed with sample analysis. Investigate per deviation SOP.
5.3 Chromatogram Evaluation
- Open raw data in validated chromatography software (e.g., Empower, Chromeleon).
- Verify baseline placement and correct peak integration.
- Ensure consistent integration settings across samples and standards.
- Assign peak identities using retention time and method-defined criteria.
5.4 Quantitative Analysis
- Confirm correct use of calibration curve and response factor.
- Ensure sample dilution, injection volume, and integration method match the BMR or analytical method.
- Calculate analyte concentrations and report with units (e.g., mg/mL, % purity).
5.5 Documentation and Approval
- Print chromatograms with audit trail and relevant overlays (standard/sample).
- Sign and date chromatogram, attach review comments.
- Submit for QA review and archive according to data integrity policy.
6. Abbreviations
- HPLC: High-Performance Liquid Chromatography
- QC: Quality Control
- RT: Retention Time
- SST: System Suitability Test
7. Documents
- Chromatogram Review Log – Annexure-1
- System Suitability Checklist – Annexure-2
- Integration and Calculation Record – Annexure-3
8. References
- ICH Q2(R2) – Validation of Analytical Procedures
- FDA Guidance – Data Integrity and Compliance with cGMP
- WHO TRS 1004 – GMP for Biotherapeutics
9. SOP Version
Version: 2.0
10. Approval Section
|
Prepared By |
Checked By |
Approved By |
| Signature |
|
|
|
| Date |
|
|
|
| Name |
|
|
|
| Designation |
|
|
|
| Department |
|
|
|
11. Annexures
Annexure-1: Chromatogram Review Log
| Date |
Sample ID |
Reviewer |
Software |
Comments |
| 04/05/2025 |
BS-CHR-011 |
Ajay Verma |
Empower |
Peak tailing acceptable |
Annexure-2: System Suitability Checklist
| Parameter |
Acceptance Criteria |
Observed |
Status |
| Resolution |
> 2.0 |
2.4 |
Pass |
| Tailing Factor |
< 2.0 |
1.5 |
Pass |
| Retention Time |
±2% |
1.98% |
Pass |
Annexure-3: Integration and Calculation Record
| Sample ID |
Peak Area |
Response Factor |
Concentration |
Reviewed By |
| BS-SM-121 |
843256 |
1.12 |
9.8 mg/mL |
Sunita Reddy |
Revision History:
| Revision Date |
Revision No. |
Details |
Reason |
Approved By |
| 04/05/2025 |
2.0 |
Updated system suitability limits and chromatographic software references |
Compliance Alignment |
|