Biosimilars: SOP for Cell Density Monitoring (VCD, OD600) in Bioreactor Operations – V 2.0
Standard Operating Procedure for Cell Density Monitoring (VCD, OD600) in Bioreactor Operations for Biosimilars
| Department |
Biosimilars |
| SOP No. |
SOP/BS/080/2025 |
| Supersedes |
SOP/BS/080/2022 |
| Page No. |
Page 1 of 12 |
| Issue Date |
04/05/2025 |
| Effective Date |
06/05/2025 |
| Review Date |
04/05/2026 |
1. Purpose
To define the standardized procedure for monitoring cell density in bioreactor operations using Viable Cell Density (VCD) via hemocytometer/automated counter and Optical Density (OD600) using a spectrophotometer during biosimilar production.
2. Scope
This SOP applies to upstream bioprocessing units involved in biosimilar production utilizing mammalian or microbial systems requiring real-time cell density monitoring.
3. Responsibilities
- Bioprocess Technician: Collect and measure cell density at defined intervals and maintain logs.
- QC Analyst: Calibrate instruments and validate measurement accuracy.
- QA Personnel: Review data logs and ensure SOP compliance.
4. Accountability
The Upstream Processing Manager is accountable for accurate implementation and compliance with cell density monitoring protocols.
5. Procedure
5.1 Sampling for Cell Density
- Collect ~5 mL of culture fluid aseptically using SOP/BS/078/2025.
- Transfer to labeled tubes and proceed with VCD or OD600 measurement immediately.
5.2 VCD Measurement Using Hemocytometer or Cell Counter
- Mix 100 µL of sample with 100 µL of 0.4% trypan blue solution.
- Load 10 µL into hemocytometer and observe under microscope (manual method) or use automatic cell counter.
- Count viable (unstained) and non-viable (blue) cells in 4 corner squares.
- Calculate VCD using:
- VCD (cells/mL) = Average viable count × Dilution Factor × 104
- Log data in Annexure-1.
5.3 OD600 Measurement (Primarily for Microbial Culture)
- Zero the spectrophotometer with fresh culture medium as blank at 600 nm.
- Measure absorbance of culture sample diluted to fall within 0.1–1.0 OD range.
- Use conversion factor for approximate CFU/mL if required.
- Record OD600 readings and time in Annexure-2.
5.4 Instrument Calibration
- Calibrate cell counter monthly using standard beads or reference slides.
- Verify OD600 accuracy quarterly using traceable reference standards.
5.5 Documentation
- Enter data in Bioreactor Monitoring Log and VCD/OD Log (Annexures 1 & 2).
- Review trends to monitor growth rate and trigger feeding or harvest decisions.
6. Abbreviations
- VCD: Viable Cell Density
- OD600: Optical Density at 600 nm
- CFU: Colony Forming Units
7. Documents
- VCD Measurement Log – Annexure-1
- OD600 Monitoring Sheet – Annexure-2
8. References
- USP <1046> – Cell Viability Assays
- ICH Q9 – Quality Risk Management
9. SOP Version
Version: 2.0
10. Approval Section
|
Prepared By |
Checked By |
Approved By |
| Signature |
|
|
|
| Date |
|
|
|
| Name |
|
|
|
| Designation |
|
|
|
| Department |
|
|
|
11. Annexures
Annexure-1: VCD Measurement Log
| Date |
Time |
Batch No. |
VCD (cells/mL) |
Viability (%) |
Operator |
| 04/05/2025 |
10:00 |
BS-080-B01 |
6.2 × 106 |
94% |
Rajesh Kumar |
Annexure-2: OD600 Monitoring Sheet
| Date |
Time |
Batch No. |
OD600 |
Dilution Factor |
Operator |
| 04/05/2025 |
10:15 |
BS-080-M01 |
0.78 |
10x |
Sunita Reddy |
Revision History:
| Revision Date |
Revision No. |
Revision Details |
Reason for Revision |
Approved By |
| 04/05/2025 |
2.0 |
Added automated counter method and OD600 calibration |
Technology Upgrade |
|