Standard Operating Procedure for Solid Phase Extraction (SPE) Method in BA/BE Bioanalytical Studies
| Department | BA-BE Studies |
|---|---|
| SOP No. | SOP/BA-BE/157/2025 |
| Supersedes | SOP/BA-BE/157/2022 |
| Page No. | Page 1 of 11 |
| Issue Date | 17/04/2025 |
| Effective Date | 20/04/2025 |
| Review Date | 17/04/2026 |
1. Purpose
To outline the procedure for performing Solid Phase Extraction (SPE) for the isolation and concentration of analytes from biological matrices such as plasma or serum during BA/BE studies.
2. Scope
This SOP is applicable to all bioanalytical personnel performing SPE in regulated BA/BE studies where the method of extraction has been validated and approved for use.
3. Responsibilities
- Bioanalytical Analyst:
Performs SPE according to validated procedures and documents all steps accurately.
Reviewer/QA: Reviews documentation, checks for deviations, and ensures compliance.
Laboratory Supervisor: Ensures availability of consumables and verifies equipment suitability.
4. Accountability
The Head of Bioanalytical Laboratory is accountable for ensuring SPE procedures are followed, validated, and meet regulatory expectations for accuracy and reproducibility.
5. Procedure
5.1 Equipment and Materials
- Vacuum manifold
- SPE cartridges (type specified in validated method)
- Collection tubes, centrifuge tubes
- Solvents: Methanol, water, buffer, elution solvent
- Pipettes and vortex mixer
- Nitrogen evaporator (if required)
5.2 Sample Preparation
- Thaw and vortex plasma samples as per SOP/BA-BE/155/2025.
- Label all SPE cartridges and collection tubes clearly with Sample ID.
- Add internal standard (IS) to each plasma sample and vortex for 15 seconds.
- If required, dilute samples with buffer before SPE loading step.
5.3 Conditioning of SPE Cartridge
- Place cartridge in SPE manifold.
- Condition the cartridge by passing:
- 1 mL methanol
- 1 mL water
- Ensure complete wetting of sorbent material for efficient binding.
5.4 Sample Loading
- Transfer plasma sample onto conditioned SPE cartridge.
- Allow it to pass through slowly under gravity or gentle vacuum (not exceeding 5–10 psi).
- Discard the filtrate unless method specifies collection of flow-through.
5.5 Washing Step
- Wash the cartridge with 1 mL of wash solution (e.g., water or aqueous buffer) to remove impurities.
- Follow with a second wash of organic solvent if specified (e.g., 5% methanol).
- Allow wash to drain completely before elution.
5.6 Elution of Analyte
- Elute the retained analyte by adding 1 mL of elution solvent (e.g., methanol or acetonitrile).
- Collect the eluate into pre-labeled collection tubes.
- If specified, repeat elution step to improve recovery.
5.7 Post-Elution Processing
- Evaporate eluates to dryness under nitrogen at 35–40°C.
- Reconstitute the residue with appropriate mobile phase (e.g., 200 µL).
- Transfer to autosampler vials and cap securely.
5.8 Precautions
- Do not allow SPE cartridges to run dry between steps.
- Use fresh pipette tips and tubes for each sample to avoid cross-contamination.
- Document each step and ensure traceability using Annexure-1: SPE Extraction Log.
6. Abbreviations
- BA/BE: Bioavailability/Bioequivalence
- SPE: Solid Phase Extraction
- IS: Internal Standard
- LC-MS/MS: Liquid Chromatography–Mass Spectrometry
7. Documents
- SPE Extraction Log – Annexure-1
- SPE Cartridge Batch Record – Annexure-2
- Solvent Preparation Log – Annexure-3
8. References
- ICH M10: Bioanalytical Method Validation
- US FDA and EMA Bioanalytical Method Validation Guidelines
- Internal validated method documentation
9. SOP Version
Version: 2.0
10. Approval Section
| Prepared By | Checked By | Approved By | |
|---|---|---|---|
| Signature | |||
| Date | |||
| Name | |||
| Designation | |||
| Department |
11. Annexures
Annexure-1: SPE Extraction Log
| Date | Sample ID | IS Added | Wash Volume | Elution Volume | Analyst |
|---|---|---|---|---|---|
| 17/04/2025 | PLS-1005 | Yes | 2 mL | 1 mL | Rajesh Kumar |
Annexure-2: SPE Cartridge Batch Record
| Cartridge Type | Lot No. | Expiry | Conditioning Verified | Checked By |
|---|---|---|---|---|
| Oasis HLB | HLB1023 | 12/2025 | Yes | Sunita Reddy |
Annexure-3: Solvent Preparation Log
| Solvent | Composition | Preparation Date | Prepared By |
|---|---|---|---|
| Elution Solvent | 100% Methanol | 16/04/2025 | Vinay Pawar |
Revision History:
| Revision Date | Revision No. | Details | Reason | Approved By |
|---|---|---|---|---|
| 01/01/2022 | 1.0 | Initial Release | New SOP | QA Head |
| 17/04/2025 | 2.0 | Added cartridge log and elution volume checks | Regulatory Alignment | QA Head |