BA-BE Studies: SOP for Sample Storage at -20°C and -70°C – V 2.0

BA-BE Studies: SOP for Sample Storage at -20°C and -70°C – V 2.0

Standard Operating Procedure for Sample Storage at -20°C and -70°C in BA/BE Studies

Department BA-BE Studies
SOP No. SOP/BA-BE/075/2025
Supersedes SOP/BA-BE/075/2022
Page No. Page 1 of 9
Issue Date 17/04/2025
Effective Date 20/04/2025
Review Date 17/04/2026

1. Purpose

To define a standardized procedure for the proper storage of biological samples, including plasma, serum, and whole blood, at -20°C and -70°C in Bioavailability/Bioequivalence (BA/BE) studies, ensuring stability and traceability of samples until analysis.

2. Scope

This SOP applies to all study personnel involved in transferring, placing, monitoring, and maintaining biological samples in temperature-controlled storage units at -20°C and -70°C at the clinical site or bioanalytical lab.

3. Responsibilities

4. Accountability

The Bioanalytical Department Head is accountable for ensuring that all biological samples are stored under validated conditions and documentation is maintained for audit and regulatory purposes.

5. Procedure

5.1 Freezer Qualification and Monitoring

  1. Ensure that freezers are:
    • Qualified and calibrated as per equipment validation protocol
    • Maintained between -18°C to -22°C (for -20°C freezer)
    • Maintained between -68°C to -75°C (for -70°C freezer)
  2. Install 24×7 temperature monitoring system with alarm notifications.
  3. Document daily temperatures in Annexure-1: Freezer Temperature Monitoring Log.

5.2 Sample Placement

  1. Verify that samples are:
    • Correctly labelled with subject ID, timepoint, and aliquot number
    • Pre-frozen at controlled conditions before transfer (if applicable)
  2. Log freezer ID, rack number, and sample code in Annexure-2: Sample Storage Log.

5.3 Storage Layout and Access

  1. Assign fixed racks and box positions for each study to avoid misplacement.
  2. Restrict access to trained lab staff and record entries in Annexure-3: Freezer Access Log.

5.4 Sample Retrieval and Transfer

  1. During sample transfer to bioanalytical labs:
    • Log retrieval time and temperature condition
    • Ensure transport containers are pre-cooled
  2. Update Annexure-2 and generate chain-of-custody documentation.

5.5 Backup Plan for Power Failure

  1. Ensure each freezer is connected to a UPS or backup generator.
  2. Keep dry ice and portable -70°C freezer ready for emergency transfer.

5.6 Deviation and Alarm Management

  1. Record any out-of-limit temperature excursions in Annexure-4: Temperature Deviation Log.
  2. Notify QA and PI for impact assessment and resolution.

6. Abbreviations

7. Documents

  1. Freezer Temperature Monitoring Log – Annexure-1
  2. Sample Storage Log – Annexure-2
  3. Freezer Access Log – Annexure-3
  4. Temperature Deviation Log – Annexure-4

8. References

9. SOP Version

Version: 2.0

10. Approval Section

Prepared By Checked By Approved By
Signature
Date
Name
Designation
Department

11. Annexures

Annexure-1: Freezer Temperature Monitoring Log

Date Freezer ID Temperature (°C) Time Checked By Remarks
17/04/2025 FZ-075 -70.2 08:00 Ravi Nair Normal

Annexure-2: Sample Storage Log

Subject ID Sample Code Freezer ID Rack No. Stored By Storage Time
VOL-075 VOL075/P1/2h/A1 FZ-02 R2 Sunita Reddy 09:10

Annexure-3: Freezer Access Log

Date Time Accessed By Purpose Freezer ID
17/04/2025 11:00 Nisha Verma Sample retrieval FZ-075

Annexure-4: Temperature Deviation Log

Date Time Observed Temp Duration Cause Action Taken Reviewed By
16/04/2025 03:30 -64.5°C 25 min Power issue Transferred to backup freezer Dr. Arvind Shah

Revision History:

Revision Date Revision No. Details Reason Approved By
12/01/2022 1.0 Initial release Process definition QA Head
17/04/2025 2.0 Expanded deviation management and access logs Audit requirement QA Head