Standard Operating Procedure for Review of Calibration Curve and QC Acceptance in BA/BE Studies
| Department | BA-BE Studies |
|---|---|
| SOP No. | SOP/BA-BE/172/2025 |
| Supersedes | SOP/BA-BE/172/2022 |
| Page No. | Page 1 of 9 |
| Issue Date | 17/04/2025 |
| Effective Date | 20/04/2025 |
| Review Date | 17/04/2026 |
1. Purpose
To establish a standard procedure for the review and acceptance of calibration curves and quality control (QC) samples generated during LC-MS/MS analysis in bioavailability and bioequivalence (BA/BE) studies.
2. Scope
This SOP applies to all bioanalytical runs generated as part of BA/BE studies where calibration standards and QC samples are used for quantitative
analysis of plasma samples using LC-MS/MS instrumentation.
3. Responsibilities
- Analyst: Reviews system suitability, evaluates calibration and QC performance, and documents results.
- Reviewer/Supervisor: Verifies correctness of curve parameters, QC recovery, and overall run validity.
- QA: Performs audits on run acceptance decisions and verifies compliance with SOP and validation protocols.
4. Accountability
The Head of Bioanalytical Laboratory is accountable for ensuring that only runs meeting the acceptance criteria are used for reporting pharmacokinetic data.
5. Procedure
5.1 Calibration Curve Review
- Each run must include at least six non-zero calibration standards covering the validated range.
- Plot peak area ratio (analyte/internal standard) against nominal concentration.
- Evaluate curve using appropriate regression (linear with 1/x² weighting preferred unless otherwise validated).
- Acceptance Criteria:
- ≥75% of non-zero calibrators should be within ±15% of nominal (±20% for LLOQ).
- Correlation coefficient (r) should be ≥ 0.99.
- Document the regression equation, r² value, and residuals in Annexure-1: Calibration Curve Summary Log.
5.2 QC Sample Review
- Each batch must contain QC samples at three levels (LQC, MQC, HQC) in duplicate or more.
- Acceptance Criteria:
- At least 67% of QCs should be within ±15% of nominal concentration.
- At least 50% of QCs at each level must meet the above criteria.
- Calculate accuracy and precision for each QC level and document in Annexure-2: QC Acceptance Summary.
- If criteria are not met, consider:
- Repeating the run
- Investigating cause of deviation
- Documenting result as a failed run
5.3 System Suitability Checks
- Confirm that:
- Retention times are within ±2% of validated range
- Peak shape is consistent
- Blank samples show no carryover
- Document performance in Annexure-3: System Suitability Log.
5.4 Documentation and Approval
- All results must be reviewed by the supervisor and QA prior to data reporting.
- Approved curves and QCs are stored in electronic and physical format.
- Any deviation from acceptance criteria must be logged in Annexure-4: Deviation Justification Record.
6. Abbreviations
- BA/BE: Bioavailability/Bioequivalence
- QC: Quality Control
- LLOQ: Lower Limit of Quantification
- LC-MS/MS: Liquid Chromatography with Tandem Mass Spectrometry
7. Documents
- Calibration Curve Summary Log – Annexure-1
- QC Acceptance Summary – Annexure-2
- System Suitability Log – Annexure-3
- Deviation Justification Record – Annexure-4
8. References
- ICH M10: Bioanalytical Method Validation
- US FDA Guidance for Industry: Bioanalytical Method Validation (2022)
- Sponsor Bioanalytical Method Validation Protocol
9. SOP Version
Version: 2.0
10. Approval Section
| Prepared By | Checked By | Approved By | |
|---|---|---|---|
| Signature | |||
| Date | |||
| Name | |||
| Designation | |||
| Department |
11. Annexures
Annexure-1: Calibration Curve Summary Log
| Batch ID | Equation | r² Value | Acceptance | Reviewer |
|---|---|---|---|---|
| BE1278-02 | y = 0.0043x + 0.002 | 0.9987 | Accepted | Sunita Reddy |
Annexure-2: QC Acceptance Summary
| QC Level | Replicate No. | Measured Value | Nominal | % Difference | Acceptance |
|---|---|---|---|---|---|
| LQC | 1 | 18.5 | 20.0 | -7.5% | Accepted |
Annexure-3: System Suitability Log
| Run Date | Retention Time (RT) | RT Acceptance | Carryover | Reviewer |
|---|---|---|---|---|
| 17/04/2025 | 3.45 | Yes | No | Rajesh Kumar |
Annexure-4: Deviation Justification Record
| Date | Batch | Issue | Justification | Approved By |
|---|---|---|---|---|
| 17/04/2025 | BE1278-03 | One QC failed | Reinjected and verified | QA Manager |
Revision History:
| Revision Date | Revision No. | Details | Reason | Approved By |
|---|---|---|---|---|
| 01/01/2022 | 1.0 | Initial version | New SOP | QA Head |
| 17/04/2025 | 2.0 | Added deviation documentation and system suitability logging | Regulatory update | QA Head |