Standard Operating Procedure for Monitoring Ambient Conditions of Sample Freezers in BA/BE Studies

Department BA-BE Studies
SOP No. SOP/BA-BE/076/2025
Supersedes SOP/BA-BE/076/2022
Page No. Page 1 of 9
Issue Date 17/04/2025
Effective Date 20/04/2025
Review Date 17/04/2026

1. Purpose

To establish a consistent and reliable procedure for monitoring the ambient temperature and relative humidity conditions of rooms where -20°C and -70°C sample freezers are installed, thereby ensuring optimal operating conditions and sample integrity in BA/BE studies.

2. Scope

This SOP applies to all areas within the clinical site or laboratory where biological sample freezers used in BA/BE studies are located and ambient condition monitoring is required.

3. Responsibilities

4. Accountability

The Facility Manager is accountable for ensuring ambient conditions are monitored and maintained within the specified limits and that proper documentation is retained.

5. Procedure

5.1 Installation and Calibration of Thermo-Hygrometers

  1. Install validated and calibrated digital thermo-hygrometers in each freezer room.
  2. Place sensors at a height of approximately 1.5 meters from the floor, away from direct vents and heat sources.
  3. Calibrate instruments every 6 months or as per manufacturer’s recommendation.

5.2 Monitoring Frequency

  1. Record ambient temperature and humidity:
    • Twice daily (e.g., 9:00 AM and 5:00 PM)
    • In Annexure-1: Ambient Condition Monitoring Log
  2. Acceptable ambient conditions:
    • Temperature: 18°C to 25°C
    • Relative Humidity (RH): 40% to 60%

5.3 Alarm and Alert System

  1. Set upper and lower alert limits in devices to notify staff in case of deviation.
  2. Ensure audio or SMS alerts are enabled for remote monitoring.

5.4 Deviation Handling

  1. In case of deviation:
    • Record in Annexure-2: Ambient Condition Deviation Log
    • Investigate cause and take immediate corrective action (e.g., adjust HVAC, limit access)
    • Inform QA and document CAPA

5.5 Preventive Maintenance

  1. Perform monthly checks of air circulation, vents, and calibration due dates of instruments.
  2. Document maintenance activities in Annexure-3: Preventive Maintenance Log.

6. Abbreviations

7. Documents

  1. Ambient Condition Monitoring Log – Annexure-1
  2. Ambient Condition Deviation Log – Annexure-2
  3. Preventive Maintenance Log – Annexure-3

8. References

9. SOP Version

Version: 2.0

10. Approval Section

Prepared By Checked By Approved By
Signature
Date
Name
Designation
Department

11. Annexures

Annexure-1: Ambient Condition Monitoring Log

Date Time Temperature (°C) Humidity (%) Checked By Remarks
17/04/2025 09:00 22.1 45 Ravi Nair Normal

Annexure-2: Ambient Condition Deviation Log

Date Time Observed Value Limit Breached Corrective Action Reviewed By
16/04/2025 17:00 27.5°C Temp > 25°C HVAC serviced, air duct opened Dr. Arvind Shah

Annexure-3: Preventive Maintenance Log

Date Activity Performed By Remarks
10/04/2025 HVAC filter cleaning Maintenance Team Completed

Revision History:

Revision Date Revision No. Details Reason Approved By
10/01/2022 1.0 Initial version Operational SOP QA Head
17/04/2025 2.0 Added alarm triggers and preventive maintenance log Audit enhancement QA Head