Standard Operating Procedure for Labeling and Identification of Analytical Batches in BA/BE Studies
| Department | BA-BE Studies |
|---|---|
| SOP No. | SOP/BA-BE/169/2025 |
| Supersedes | SOP/BA-BE/169/2022 |
| Page No. | Page 1 of 9 |
| Issue Date | 17/04/2025 |
| Effective Date | 20/04/2025 |
| Review Date | 17/04/2026 |
1. Purpose
To define a standard procedure for the labeling and unique identification of analytical batches in bioanalytical studies supporting BA/BE clinical trials to ensure traceability, prevent sample mix-ups, and comply with regulatory standards.
2. Scope
This SOP is applicable to all analysts and supervisors involved in labeling, identifying, and documenting analytical batches prepared for LC-MS/MS analysis of plasma samples in BA/BE studies.
3. Responsibilities
- Analyst: Prepares and labels analytical batches as per SOP guidelines and logs them in batch identification records.
- Supervisor: Reviews and verifies correctness of labels and documentation.
- QA: Performs periodic audits to ensure compliance with labeling conventions and traceability standards.
4. Accountability
The Bioanalytical Laboratory Manager is accountable for the implementation of consistent labeling formats and batch identification systems to support GLP and GCP practices.
5. Procedure
5.1 Batch Numbering Convention
- Each analytical batch shall be assigned a unique ID following the format:
- Format: [StudyCode]-[AnalyteCode]-[BatchNumber]
- Example: BE1234-MET-01
- The batch number shall be sequential and reset for each new analyte in the same study.
- Document all batch numbers in Annexure-1: Analytical Batch Identification Log.
5.2 Labeling of Vials and Trays
- All vials (calibrators, QCs, and samples) shall be labeled with:
- Study Code
- Sample ID
- Batch ID
- Vial position (e.g., B5)
- Trays must also carry tray number, run ID, and analyst initials.
- Use tamper-evident and water-resistant printed barcode labels wherever applicable.
5.3 Sequence File Naming and Metadata
- Each instrument sequence file shall be named according to:
- [StudyCode]_Seq_[BatchNumber]_YYYYMMDD
- Example: BE1234_Seq_01_20250417
- Save sequence files in a structured folder system with study-wise and batch-wise segregation.
5.4 Label Verification and Documentation
- After labeling, a second analyst or supervisor must cross-verify the labels and sign in Annexure-2: Label Verification Checklist.
- Any corrections must be made before sample loading into the autosampler tray.
5.5 Deviations and Corrections
- In case of incorrect labeling:
- Segregate affected vials immediately
- Replace labels using reprint procedure and document in Annexure-3: Labeling Deviation Log
- Cross-check sample IDs with CRF and sample receipt log before correction.
6. Abbreviations
- BA/BE: Bioavailability/Bioequivalence
- QC: Quality Control
- CRF: Case Record Form
- GLP: Good Laboratory Practice
- GCP: Good Clinical Practice
7. Documents
- Analytical Batch Identification Log â Annexure-1
- Label Verification Checklist â Annexure-2
- Labeling Deviation Log â Annexure-3
8. References
- ICH M10: Bioanalytical Method Validation
- FDA Bioanalytical Method Validation Guidance
- Internal Labeling SOP Policy
9. SOP Version
Version: 2.0
10. Approval Section
| Prepared By | Checked By | Approved By | |
|---|---|---|---|
| Signature | |||
| Date | |||
| Name | |||
| Designation | |||
| Department |
11. Annexures
Annexure-1: Analytical Batch Identification Log
| Batch ID | Study Code | Analyte | Date Created | Prepared By |
|---|---|---|---|---|
| BE1234-MET-01 | BE1234 | Metformin | 17/04/2025 | Sunita Reddy |
Annexure-2: Label Verification Checklist
| Sample ID | Label Content Verified | Initials | Date |
|---|---|---|---|
| BE101-TP3 | Yes | RK | 17/04/2025 |
Annexure-3: Labeling Deviation Log
| Date | Sample ID | Issue | Corrective Action | Verified By |
|---|---|---|---|---|
| 17/04/2025 | BE101-TP5 | Missing Vial Position | Reprinted label and updated log | QA Officer |
Revision History:
| Revision Date | Revision No. | Details | Reason | Approved By |
|---|---|---|---|---|
| 01/01/2022 | 1.0 | Initial release | New SOP | QA Head |
| 17/04/2025 | 2.0 | Inclusion of barcode labels and deviation log | Compliance update | QA Head |