Standard Operating Procedure for Documentation of Sample Storage Conditions in BA/BE Studies
| Department | BA-BE Studies |
|---|---|
| SOP No. | SOP/BA-BE/194/2025 |
| Supersedes | SOP/BA-BE/194/2022 |
| Page No. | Page 1 of 8 |
| Issue Date | 17/04/2025 |
| Effective Date | 20/04/2025 |
| Review Date | 17/04/2026 |
1. Purpose
To define the procedure for documenting and maintaining the storage conditions of biological samples such as plasma, serum, and urine during BA/BE studies to ensure sample integrity and compliance with regulatory requirements.
2. Scope
This SOP applies to all biological samples stored in freezers or refrigerators at the bioanalytical facility throughout the duration of BA/BE studies.
3. Responsibilities
- Sample Custodian: Ensures sample storage
conditions are monitored and documented.
QA Personnel: Reviews storage logs and validates compliance during internal audits.
Lab Technicians: Monitor temperatures and notify for excursions.
Bioanalytical Head: Ensures SOP adherence and CAPA implementation when required.
4. Accountability
The Head of the Bioanalytical Department is accountable for maintaining compliant and auditable records of sample storage conditions for all biological samples.
5. Procedure
5.1 Designated Storage Areas
- Samples are stored in designated:
- Deep freezers (-70°C or lower)
- Freezers (-20°C)
- Refrigerators (2–8°C)
- Each unit must have a unique identifier (e.g., FZ-001, RF-002).
5.2 Temperature Monitoring Devices
- Equip each unit with a calibrated temperature data logger or digital thermometer.
- Devices must be:
- Calibrated annually
- Placed at the center or most critical zone
- Attach display readers outside for quick visual check.
5.3 Documentation of Temperature
- Record temperatures twice daily (start and end of shift) in Annexure-1: Sample Storage Temperature Log.
- Entries must include:
- Date and time
- Storage unit ID
- Temperature (°C)
- Initials of personnel
- Print and archive logger graphs weekly in a binder specific to each unit.
5.4 Handling Temperature Excursions
- If temperature is out of acceptable range:
- Label the storage unit as “Under Observation.”
- Move samples to backup storage.
- Document excursion in Annexure-2: Temperature Excursion Log.
- Inform QA immediately.
- Initiate CAPA as needed and document resolution.
5.5 Backup Power and Alarm Systems
- Ensure:
- Uninterrupted Power Supply (UPS) or generator support is available.
- Freezer units are connected to temperature alarms with 24×7 alert notification.
5.6 Sample Inventory Mapping
- Each freezer must maintain an updated Sample Inventory Sheet (Annexure-3).
- Include:
- Study code
- Rack and box number
- Sample IDs and date of placement
- Reconcile inventory monthly with receipt logs.
6. Abbreviations
- SOP: Standard Operating Procedure
- QA: Quality Assurance
- UPS: Uninterrupted Power Supply
- CAPA: Corrective and Preventive Action
- °C: Degrees Celsius
7. Documents
- Sample Storage Temperature Log – Annexure-1
- Temperature Excursion Log – Annexure-2
- Sample Inventory Sheet – Annexure-3
8. References
- ICH GCP E6(R2)
- OECD GLP Principles
- 21 CFR Part 58 – FDA GLP Regulations
9. SOP Version
Version: 2.0
10. Approval Section
| Prepared By | Checked By | Approved By | |
|---|---|---|---|
| Signature | |||
| Date | |||
| Name | |||
| Designation | |||
| Department |
11. Annexures
Annexure-1: Sample Storage Temperature Log
| Date | Time | Unit ID | Temperature (°C) | Initials |
|---|---|---|---|---|
| 17/04/2025 | 09:00 | FZ-003 | -22.3 | AV |
Annexure-2: Temperature Excursion Log
| Date | Unit ID | Observed Temp (°C) | Duration | Corrective Action |
|---|---|---|---|---|
| 16/04/2025 | FZ-002 | -14.2 | 1 hour | Transferred to FZ-001 |
Annexure-3: Sample Inventory Sheet
| Study Code | Rack No. | Box No. | Sample IDs | Date Placed |
|---|---|---|---|---|
| BE/XYZ/2025/05 | Rack-2 | Box-3 | SUB001 to SUB010 | 12/04/2025 |
Revision History:
| Revision Date | Revision No. | Details | Reason | Approved By |
|---|---|---|---|---|
| 01/01/2022 | 1.0 | Initial SOP | New Procedure | QA Head |
| 17/04/2025 | 2.0 | Included digital logging and excursion protocol | Audit Enhancement | QA Head |