Standard Operating Procedure for Batch Rejection and Repeat Analysis in BA/BE Studies
| Department | BA-BE Studies |
|---|---|
| SOP No. | SOP/BA-BE/197/2025 |
| Supersedes | SOP/BA-BE/197/2022 |
| Page No. | Page 1 of 9 |
| Issue Date | 17/04/2025 |
| Effective Date | 20/04/2025 |
| Review Date | 17/04/2026 |
1. Purpose
This SOP outlines the systematic procedure to identify criteria for rejecting analytical batches and the process to repeat the analysis, ensuring that data quality and regulatory compliance are maintained during bioanalytical evaluations in BA/BE studies.
2. Scope
This SOP applies to all bioanalytical runs performed in the quantification of drug concentrations in biological matrices like plasma, serum, or urine under BA/BE studies conducted
at the bioanalytical laboratory.
3. Responsibilities
- Analyst: Performs initial batch review and flags issues requiring batch rejection.
- QA Officer: Reviews justification for rejection and ensures documentation is complete.
- Bioanalytical Reviewer: Approves repeat analysis plan and verifies data compliance post-repeat.
- Bioanalytical Head: Authorizes final decision on batch rejection and approves repeat analysis.
4. Accountability
The Bioanalytical Head is accountable for maintaining data integrity by ensuring rejected batches are properly documented and repeat analyses are justified and conducted according to applicable guidelines.
5. Procedure
5.1 Criteria for Batch Rejection
- An analytical batch shall be rejected if:
- QC sample failure exceeds the acceptable limits as per ICH M10 (more than 33% of total or 50% at any level).
- Calibration curve fails to meet accuracy or linearity criteria.
- System suitability parameters (e.g., peak shape, retention time) are out of range.
- Chromatographic issues such as interference, co-elution, or carryover.
- Deviation in run sequence, labeling, or sample position errors.
5.2 Documentation of Batch Rejection
- Record rejection details in Annexure-1: Batch Rejection Log including:
- Run ID
- Study code
- Date of analysis
- Reason for rejection
- Authorized by
- Annotate chromatograms and data files clearly with “REJECTED” watermark and maintain in the archival file.
5.3 Planning Repeat Analysis
- Prepare a Repeat Analysis Request Form (Annexure-2) with:
- List of samples to be repeated
- Justification
- Proposed method and timeline
- QA and Bioanalytical Head must jointly approve the repeat analysis plan before initiation.
- Repeat analysis must use:
- Fresh calibration and QC samples
- Same method and validated parameters
5.4 Conduct of Repeat Analysis
- Follow same procedures as in the original batch preparation and processing.
- Maintain records in the Repeat Analysis Log (Annexure-3).
- Tag repeat runs with a unique batch ID and reference original run ID.
5.5 Review and Reporting
- Compare original and repeat results and justify any variation in values exceeding ±20%.
- Include repeat run data in the bioanalytical report with annotations and explanation for original run failure.
- Ensure that the final reported result is from the repeat batch unless otherwise justified.
6. Abbreviations
- BA/BE: Bioavailability/Bioequivalence
- SOP: Standard Operating Procedure
- QC: Quality Control
- QA: Quality Assurance
- ICH: International Council for Harmonisation
7. Documents
- Batch Rejection Log – Annexure-1
- Repeat Analysis Request Form – Annexure-2
- Repeat Analysis Log – Annexure-3
8. References
- ICH M10 Bioanalytical Method Validation Guideline
- FDA Guidance on Bioanalytical Method Validation (2022)
- OECD GLP Principles
9. SOP Version
Version: 2.0
10. Approval Section
| Prepared By | Checked By | Approved By | |
|---|---|---|---|
| Signature | |||
| Date | |||
| Name | |||
| Designation | |||
| Department |
11. Annexures
Annexure-1: Batch Rejection Log
| Date | Run ID | Study Code | Reason | Approved By |
|---|---|---|---|---|
| 17/04/2025 | RUN-058 | BE/ABC/2025 | QC Failure | Sunita Reddy |
Annexure-2: Repeat Analysis Request Form
| Sample IDs | Original Run ID | Justification | Approval Date | Authorized By |
|---|---|---|---|---|
| SUB005, SUB006 | RUN-058 | Peak anomaly | 17/04/2025 | Rajesh Kumar |
Annexure-3: Repeat Analysis Log
| Date | Repeat Run ID | Original Run ID | Analyst | Remarks |
|---|---|---|---|---|
| 18/04/2025 | RUN-058R | RUN-058 | Ajay Verma | QC Passed |
Revision History:
| Revision Date | Revision No. | Details | Reason | Approved By |
|---|---|---|---|---|
| 01/01/2022 | 1.0 | Initial SOP | New Process Implementation | QA Head |
| 17/04/2025 | 2.0 | Included CAPA tracking and reinjection references | Audit Compliance | QA Head |