Standard Operating Procedure for Temperature Mapping in Drying Chambers in API Manufacturing

Department API Manufacturing
SOP No. SOP/API/072/2025
Supersedes SOP/API/072/2022
Page No. Page 1 of 10
Issue Date 13/04/2025
Effective Date 15/04/2025
Review Date 13/04/2026

1. Purpose

To establish a validated procedure for performing temperature mapping in drying chambers, including tray dryers, vacuum ovens, and fluid bed dryers, to confirm uniform temperature distribution during operation.

2. Scope

This SOP applies to all drying chambers used in API manufacturing that require temperature validation or requalification following installation, major maintenance, or at predefined intervals.

3. Responsibilities

4. Accountability

The Validation Head is accountable for ensuring correct execution of temperature mapping. The QA Head is responsible for verifying compliance and approving validation documentation.

5. Procedure

5.1 Pre-Mapping Preparation

  1. Develop and approve a validation protocol describing:
    • Number and location of sensors (minimum 9-point grid for small chambers, 15–21 for larger)
    • Duration of mapping
    • Acceptance criteria (±3°C unless otherwise justified)
    • Calibration details of sensors/data loggers
  2. Ensure the chamber is clean and empty before mapping unless performing loaded qualification.
  3. Label each sensor with location and record serial number.

5.2 Sensor Placement

  1. Place temperature sensors:
    • Top-left, top-center, top-right
    • Middle-left, middle-center, middle-right
    • Bottom-left, bottom-center, bottom-right
  2. In large chambers, additional sensors should be placed at airflow entry/exit and dead zones identified in prior studies.

5.3 Data Collection

  1. Set chamber to standard operating temperature (e.g., 60°C for vacuum dryers).
  2. Begin mapping when steady state is achieved (stabilization time ~30–60 minutes).
  3. Record temperature readings at 1–5 minute intervals for a minimum of 2 hours or as per protocol.
  4. Download and save data from each logger in electronic format.

5.4 Data Analysis and Reporting

  1. Analyze temperature variation across all sensors.
  2. Calculate maximum, minimum, and average temperature.
  3. Identify cold/hot spots and verify all readings fall within acceptance limits.
  4. Prepare and submit “Temperature Mapping Report” (Annexure-1) including graphs and summary.

5.5 Post-Mapping Activities

  1. Remove all sensors and restore equipment to operational readiness.
  2. Label equipment as “Validated” and update equipment validation status log.
  3. If mapping fails, perform investigation and corrective actions before repeat mapping.

6. Abbreviations

7. Documents

  1. Temperature Mapping Report (Annexure-1)
  2. Validation Protocol and Summary
  3. Sensor Calibration Certificate

8. References

9. SOP Version

Version: 2.0

10. Approval Section

Prepared By Checked By Approved By
Signature
Date
Name
Designation
Department

11. Annexures

Annexure-1: Temperature Mapping Report Summary

Date Equipment ID Set Temp (°C) Max (°C) Min (°C) Avg (°C) Pass/Fail
13/04/2025 TRAY-DRY-01 60 62.1 58.3 60.2 Pass

Revision History:

Revision Date Revision No. Details Reason Approved By
01/01/2022 1.0 Initial SOP Release New Process Equipment Commissioning QA Head
13/04/2025 2.0 Updated sensor layout and acceptance criteria GMP Audit Compliance QA Head