Standard Operating Procedure for Operation of FTIR Spectrometer in API Manufacturing
| Department | API Manufacturing |
|---|---|
| SOP No. | SOP/API/148/2025 |
| Supersedes | SOP/API/148/2022 |
| Page No. | Page 1 of 10 |
| Issue Date | 13/04/2025 |
| Effective Date | 15/04/2025 |
| Review Date | 13/04/2026 |
1. Purpose
To provide detailed instructions for the proper and consistent use of Fourier Transform Infrared (FTIR) Spectrometer for analysis of raw materials, intermediates, and APIs in the Quality Control laboratory of the API manufacturing unit.
2. Scope
This SOP applies to the FTIR spectrometer installed in the QC laboratory used for identity confirmation, functional group analysis, and comparison of sample spectra with reference standards
in API manufacturing.
3. Responsibilities
- QC Analyst: Operate the FTIR spectrometer as per this SOP and record observations accurately.
- QC Reviewer: Review and approve the FTIR results and ensure compliance with specification and method.
- Instrument Coordinator: Ensure calibration, maintenance, and troubleshooting of FTIR spectrometer.
4. Accountability
The Quality Control Head is accountable for ensuring compliance with the SOP and maintaining FTIR system readiness for routine and validation analysis.
5. Procedure
5.1 Instrument Setup
- Switch on the main power supply and instrument PC.
- Launch the FTIR software (e.g., OMNIC, Spectrum, etc.).
- Allow the instrument to stabilize for a minimum of 30 minutes.
- Check the instrument logbook for any ongoing maintenance or deviations.
5.2 Background Scan
- Ensure the sample compartment is empty and clean.
- Select “Collect Background” from the software.
- Use a suitable number of scans (e.g., 32) and appropriate resolution (e.g., 4 cm⁻¹).
- Save the background file with date and analyst initials.
5.3 Sample Preparation
- Prepare sample based on material type:
- Solid samples: Prepare KBr pellets or use ATR (Attenuated Total Reflectance) technique.
- Liquid samples: Use appropriate liquid cells or ATR with a clean crystal.
- Ensure that the reference standard is similarly prepared for comparison.
- Clean the sampling accessories before and after use.
5.4 Spectrum Collection
- Place the sample on the ATR crystal or in the sample holder.
- Click “Collect Sample” in the software and run at least 32 scans.
- Ensure a proper signal-to-noise ratio and well-defined peaks.
- Compare with standard spectrum and overlay both for confirmation.
5.5 Interpretation and Reporting
- Assign functional group peaks (e.g., O-H, N-H, C=O, C-H, etc.) according to literature.
- Print spectrum and attach to worksheet along with:
- Sample name, batch number
- Reference used
- Analyst name, date
- Document results in the FTIR Analysis Log (Annexure-1).
5.6 Calibration and Verification
- Perform wavelength accuracy check using polystyrene film once per week or before use.
- Record calibration in the FTIR Calibration Log (Annexure-2).
- Ensure absorbance accuracy and repeatability are verified monthly.
5.7 Maintenance and Cleaning
- Clean the sample area daily using lint-free tissue and ethanol.
- Close software and shut down instrument properly after use.
- Log maintenance in the Instrument Maintenance Record (Annexure-3).
6. Abbreviations
- FTIR: Fourier Transform Infrared
- QC: Quality Control
- KBr: Potassium Bromide
- ATR: Attenuated Total Reflectance
- SOP: Standard Operating Procedure
7. Documents
- FTIR Analysis Log (Annexure-1)
- FTIR Calibration Log (Annexure-2)
- Instrument Maintenance Record (Annexure-3)
- Spectra Printouts and Interpretation Sheets
8. References
- ICH Q2(R1): Validation of Analytical Procedures
- Instrument Manufacturer Manual
- 21 CFR Part 211 (cGMP for Analytical Equipment)
9. SOP Version
Version: 2.0
10. Approval Section
| Prepared By | Checked By | Approved By | |
|---|---|---|---|
| Signature | |||
| Date | |||
| Name | |||
| Designation | |||
| Department |
11. Annexures
Annexure-1: FTIR Analysis Log
| Date | Sample ID | Reference | Instrument ID | Analyst | Result |
|---|---|---|---|---|---|
| 13/04/2025 | SMP-148 | API-STD-001 | FTIR-01 | Sunita Reddy | Complies |
Annexure-2: FTIR Calibration Log
| Date | Calibration Performed | Polystyrene Film Used | Wavelength Accuracy | Performed By |
|---|---|---|---|---|
| 10/04/2025 | Yes | PS-2025 | Within 1 cm⁻¹ | Ravi Kumar |
Annexure-3: Instrument Maintenance Record
| Date | Activity | Performed By | Remarks |
|---|---|---|---|
| 13/04/2025 | Cleaned ATR Crystal | Rajesh Kumar | No residue observed |
Revision History:
| Revision Date | Revision No. | Details | Reason | Approved By |
|---|---|---|---|---|
| 01/01/2022 | 1.0 | Initial issue | Instrument Operation SOP | QA Head |
| 13/04/2025 | 2.0 | Format standardization and annexure addition | Internal Audit Requirement | QA Head |