SOP for UV Spectrophotometric Method Development in AMD Laboratory
Department | Analytical Method Development |
---|---|
SOP No. | SOP/AMD/054/2025 |
Supersedes | SOP/AMD/054/2022 |
Page No. | Page 1 of 14 |
Issue Date | 19/05/2025 |
Effective Date | 20/05/2025 |
Review Date | 19/05/2026 |
1. Purpose
The purpose of this SOP is to describe the standard procedure for developing, optimizing, and validating UV spectrophotometric methods for the quantitative analysis of active pharmaceutical ingredients (APIs), intermediates, and finished dosage forms using a UV-visible spectrophotometer.
2. Scope
This procedure applies to the Analytical Method Development (AMD) laboratory and includes all steps from wavelength scanning, method optimization, calibration curve generation, and full validation according to ICH Q2(R1) and pharmacopeial guidelines.
3. Responsibilities
- Analytical Scientist: Responsible for scanning, method development, standardization, and performing validation parameters.
- Reviewer: Verifies raw data, calculations, and validation outcomes for compliance.
- QA Personnel: Reviews final method reports and ensures SOP compliance with regulatory standards.
- Head – AMD: Oversees the method development process and approves final documentation for use in QC or regulatory filings.
4. Accountability
The Head of AMD is accountable for ensuring UV spectrophotometric methods are scientifically sound, validated, and regulatory-compliant for intended pharmaceutical use.
5. Procedure
5.1 Selection of Analytical Wavelength
- Perform a UV spectrum scan of a known concentration of analyte between 200 nm and 400 nm.
- Identify the wavelength of maximum absorbance (λmax).
- Ensure the λmax is not overlapped by known excipients or impurities.
- Record the λmax and corresponding absorbance in Annexure-1: Wavelength Selection Log.
5.2 Preparation of Standard Stock and Working Solutions
- Prepare a standard stock solution of the analyte using a certified reference standard in a suitable solvent (typically water, methanol, or buffer).
- Dilute the stock to prepare 5–7 concentrations over the intended working range (e.g., 5 ppm to 50 ppm).
- Use the same diluent as the sample matrix to avoid matrix mismatch.
- Record details in Annexure-2: Standard Preparation Sheet.
5.3 Calibration Curve Construction
- Measure absorbance of each standard solution at the selected λmax using a blank as reference.
- Plot absorbance vs. concentration using appropriate software or manually.
- Calculate the regression line equation and determine correlation coefficient (R²).
- Acceptable linearity: R² ≥ 0.999.
- Record in Annexure-3: Calibration Curve Log.
5.4 Sample Preparation
- Prepare sample solutions in the same matrix/diluent as standards.
- Filter or centrifuge if required to remove insoluble matter.
- Ensure concentration of test solution falls within standard curve range.
- Document results in Annexure-4: Sample Analysis Report.
5.5 Method Validation
- Validate the method as per ICH Q2(R1) and pharmacopeial expectations.
- Specificity: Ensure blank and excipients do not absorb at λmax.
- Linearity: Prepare at least 5 concentrations (80–120%) and confirm R² ≥ 0.999.
- Accuracy (Recovery): Spike placebo at 80%, 100%, and 120% levels. Recovery should be 98–102%.
- Precision:
- Repeatability (intra-day): 6 replicates at 100% level, RSD ≤ 2.0%
- Intermediate precision (inter-day/different analyst): RSD ≤ 2.0%
- LOD/LOQ: Use calibration curve method or signal-to-noise (S/N) method (LOD S/N = 3.3, LOQ S/N = 10).
- Robustness: Vary wavelength (±1 nm), temperature (±2°C), and diluent ratio (±5%).
- Solution Stability: Measure absorbance over 6 hours at room temperature. Acceptable deviation ≤ 2%.
- All validation data must be documented in Annexure-5: Method Validation Summary.
5.6 Final Method Documentation
- Compile complete method including:
- Title and scope
- Reagents and solutions
- Instrument settings
- Calculation formulae
- Acceptance criteria
- System suitability tests
- Ensure peer-review and QA approval before use in stability or routine analysis.
6. Abbreviations
- UV: Ultraviolet
- SOP: Standard Operating Procedure
- RSD: Relative Standard Deviation
- ICH: International Council for Harmonisation
- QA: Quality Assurance
- λmax: Wavelength of maximum absorbance
7. Documents
- Wavelength Selection Log – Annexure-1
- Standard Preparation Sheet – Annexure-2
- Calibration Curve Log – Annexure-3
- Sample Analysis Report – Annexure-4
- Method Validation Summary – Annexure-5
8. References
- ICH Q2(R1): Validation of Analytical Procedures
- USP <857>: Ultraviolet-Visible Spectroscopy
- IP 2022: Appendix – UV Methods
- Ph. Eur. 2.2.25: UV-Vis Spectrophotometry
9. SOP Version
Version: 2.0
10. Approval Section
Prepared By | Checked By | Approved By | |
---|---|---|---|
Signature | |||
Date | |||
Name | |||
Designation | |||
Department |
11. Annexures
Annexure-1: Wavelength Selection Log
Sample | λmax (nm) | Absorbance | Selected By |
---|---|---|---|
API-A | 272 | 1.215 | Ajay Mehra |
Annexure-2: Standard Preparation Sheet
Stock Conc. (ppm) | Dilution Volume | Final Conc. (ppm) | Prepared By |
---|---|---|---|
100 | 10 mL in 100 mL | 10 | Sunita Reddy |
Annexure-3: Calibration Curve Log
Conc. (ppm) | Absorbance |
---|---|
5 | 0.255 |
10 | 0.512 |
20 | 1.021 |
30 | 1.527 |
40 | 2.038 |
Annexure-4: Sample Analysis Report
Sample | Absorbance | Calculated Conc. (ppm) | Analyst |
---|---|---|---|
Formulation X | 0.984 | 19.3 | Rajesh Kumar |
Annexure-5: Method Validation Summary
Parameter | Criteria | Result | Status |
---|---|---|---|
Linearity | R² ≥ 0.999 | 0.9996 | Pass |
Accuracy | 98–102% | 99.2% | Pass |
Precision | RSD ≤ 2.0% | 1.1% | Pass |
LOD | Calculated | 1.5 ppm | Pass |
LOQ | Calculated | 4.8 ppm | Pass |
Revision History:
Revision Date | Revision No. | Details | Reason | Approved By |
---|---|---|---|---|
04/05/2025 | 2.0 | Expanded validation parameters and added solution stability study | Audit Preparedness |