Standard Operating Procedure for Uniformity of Dosage Units in Semi-Solids in Analytical Method Development


Department Analytical Method Development
SOP No. SOP/AMD/258/2025
Supersedes SOP/AMD/258/2022
Page No. Page 1 of 12
Issue Date 21/05/2025
Effective Date 23/05/2025
Review Date 21/05/2026

1. Purpose

To define the procedure for evaluating the uniformity of dosage units in semi-solid formulations such as creams, ointments, and gels, ensuring

consistency of active ingredient distribution within and across containers.

2. Scope

This SOP applies to the Analytical Method Development (AMD) team responsible for assessing content uniformity and/or weight variation in semi-solid dosage forms during formulation development, batch release, and stability testing.

3. Responsibilities

  • Analytical Scientist: Conducts the test and evaluates data against acceptance criteria.
  • QC Analyst: Prepares individual dosage units, performs sample analysis, and records results.
  • QA Executive: Reviews the results for compliance with internal and regulatory standards.

4. Accountability

The Head of Analytical Method Development is accountable for implementing and validating the dosage uniformity method and ensuring compliance with regulatory guidelines such as USP <905> and ICH Q6A.

5. Procedure

5.1 Definitions

Dosage Unit: A specified quantity of semi-solid formulation (e.g., 1 g or 2 g) withdrawn consistently from different locations of the container.

5.2 Sample Selection

  1. Select 10 dosage units randomly from a single batch (entirely filled containers).
  2. Collect from the top, middle, and bottom to account for any stratification.

5.3 Sample Preparation

  1. Weigh 1.0 g (or labeled unit dose) from each selected container into a suitable volumetric flask.
  2. Add diluent (e.g., methanol:water), sonicate for 10–15 minutes, and dilute to volume.
  3. Filter through 0.45 µm membrane before injection.

5.4 Standard Preparation

  1. Weigh accurately known quantity of reference standard and dilute to match expected concentration of test samples.
  2. Filter through 0.45 µm filter and inject along with test samples.

5.5 Analytical Method

  1. Use validated HPLC or UV method for assay.
  2. Record absorbance or peak area and calculate the percentage of label claim for each unit.

5.6 Acceptance Criteria (USP <905>)

  • 10 units must fall within 85%–115% of label claim.
  • RSD ≤ 6.0% for assay results.
  • If one result is outside 85%–115% but within 75%–125%, test an additional 20 units.
  • Content uniformity passes if not more than 3 of 30 units fall outside 85%–115% and none outside 75%–125%.

5.7 Method Validation

  1. Precision: Intra-day and inter-day RSD ≤ 2.0%
  2. Linearity: r ≥ 0.999 across 50–150% of target concentration
  3. Recovery: 98%–102% across various matrix levels

5.8 Documentation

  1. Record individual sample weights, absorbance/peak area, and calculated assay values.
  2. Attach chromatograms or UV spectra along with calculation sheets in Annexure-1.

6. Abbreviations

  • HPLC: High Performance Liquid Chromatography
  • RSD: Relative Standard Deviation
  • QA: Quality Assurance
  • AMD: Analytical Method Development

7. Documents

  1. Dosage Unit Assay Log – Annexure-1
  2. Validation Summary Report – Annexure-2

8. References

  • USP <905>: Uniformity of Dosage Units
  • ICH Q6A: Specifications for New Drug Substances and Products
  • EP 2.9.40: Uniformity of Dosage Units

9. SOP Version

Version: 2.0

10. Approval Section

Prepared By Checked By Approved By
Signature
Date
Name Neelam Joshi Varun Iyer Sunita Reddy
Designation Analytical Scientist QA Reviewer Head – AMD
Department Analytical Method Development QA Analytical Method Development

11. Annexures

Annexure-1: Dosage Unit Assay Log

Unit No. Weight (g) Peak Area % Label Claim Status
1 1.02 15432 99.2% Pass
2 0.98 15012 98.6% Pass
3 1.01 15500 99.5% Pass

Annexure-2: Validation Summary Report

The method was validated across 3 batches of ointment product O-258. Results met criteria for specificity, linearity, precision, and accuracy. The method is suitable for routine batch release and stability testing.

Revision History:

Revision Date Revision No. Details Reason Approved By
21/05/2025 2.0 Added clarification on extended testing and updated annexures Annual Review Sunita Reddy
01/04/2022 1.0 Initial Release New SOP QA Head