Standard Operating Procedure for TGA Method Development for Moisture and Volatile Content Analysis
| Department | Analytical Method Development |
|---|---|
| SOP No. | SOP/AMD/173/2025 |
| Supersedes | SOP/AMD/173/2022 |
| Page No. | Page 1 of 14 |
| Issue Date | 19/05/2025 |
| Effective Date | 20/05/2025 |
| Review Date | 19/05/2026 |
1. Purpose
To describe the procedure for the development and validation of Thermogravimetric Analysis (TGA) methods to quantitatively determine moisture content, volatile solvents, and decomposition profiles in
pharmaceutical materials including APIs, excipients, and finished products.
2. Scope
This SOP applies to the Analytical Method Development (AMD) department for evaluating material stability, hygroscopicity, and thermal degradation using TGA instrumentation in R&D and routine analysis.
3. Responsibilities
- Analytical Scientist: Responsible for method design, thermal profile evaluation, and interpretation.
- Instrument Analyst: Operates and maintains the TGA instrument, performs calibration and sample runs.
- QA Officer: Ensures review and compliance of validation reports and experimental data.
- Head – AMD: Approves the developed TGA method and validates its implementation in QC or stability labs.
4. Accountability
The Head of Analytical Method Development is accountable for ensuring that all TGA methods are scientifically sound, validated, and aligned with ICH and pharmacopoeial standards.
5. Procedure
5.1 Instrument Setup and Calibration
- Ensure the TGA instrument is calibrated for temperature and weight using certified standards (e.g., calcium oxalate monohydrate).
- Document calibration data in Annexure-1: TGA Calibration Log.
- Verify purge gas (e.g., nitrogen or air) is connected and pressure is stable at 50–60 mL/min.
5.2 Sample Preparation
- Weigh 5–10 mg of powdered sample using an analytical balance.
- Use pre-tared, clean platinum or alumina crucibles with lids if applicable.
- Avoid contact with moisture prior to measurement; store hygroscopic samples in desiccators until analysis.
5.3 Method Parameters for Moisture and Volatile Analysis
- Set temperature program:
- Start Temp: 25°C
- End Temp: 150–300°C (depending on analyte)
- Ramp Rate: 10°C/min
- Use isothermal holds at 105°C for accurate moisture quantification if needed.
- Record weight loss and derivative thermogravimetric (DTG) curves.
- Determine:
- % Moisture (loss below 120°C)
- % Volatile (loss between 120°C–200°C)
- Thermal degradation (onset temperature)
- Enter all thermal transitions in Annexure-2: TGA Profile Report.
5.4 Method Validation
- Validate TGA method per ICH Q2(R2) guidelines:
- Precision – Repeatability of % loss (RSD ≤ 3%)
- Accuracy – Recovery studies with known spiked water or solvent samples
- Specificity – Ability to distinguish moisture from solvents
- Robustness – Evaluate different ramp rates or crucible types
- Document validation summary in Annexure-3: Validation Log.
5.5 Safety Considerations
- Ensure the crucibles are cool before removing from furnace.
- Handle decomposing samples in fume hood if toxic gases are expected.
- Dispose of thermally decomposed residues as per hazardous waste policy.
6. Abbreviations
- TGA: Thermogravimetric Analysis
- DTG: Derivative Thermogravimetry
- RSD: Relative Standard Deviation
- ICH: International Council for Harmonisation
- SOP: Standard Operating Procedure
7. Documents
- TGA Calibration Log – Annexure-1
- TGA Profile Report – Annexure-2
- Validation Log – Annexure-3
8. References
- ICH Q2(R2) – Validation of Analytical Procedures
- USP <891> – Thermal Analysis
- European Pharmacopoeia 2.2.34 – Thermogravimetric Analysis
9. SOP Version
Version: 2.0
10. Approval Section
| Prepared By | Checked By | Approved By | |
|---|---|---|---|
| Signature | |||
| Date | |||
| Name | |||
| Designation | |||
| Department |
11. Annexures
Annexure-1: TGA Calibration Log
| Material | Theoretical Temp (°C) | Observed Temp (°C) | Weight Loss (%) | Analyst |
|---|---|---|---|---|
| Calcium Oxalate | 133 | 132.8 | 18.5 | Rajesh Kumar |
Annexure-2: TGA Profile Report
| Sample ID | Moisture (%) | Volatile (%) | Decomposition Onset (°C) | Analyst |
|---|---|---|---|---|
| API-TGA-22 | 2.1 | 3.5 | 218.6 | Sunita Reddy |
Annexure-3: Validation Log
| Parameter | Result | Acceptance Criteria | Status |
|---|---|---|---|
| Repeatability | RSD = 2.4% | ≤ 3% | Pass |
| Accuracy | 98.8% | 95%–105% | Pass |
Revision History:
| Revision Date | Revision No. | Details | Reason | Approved By |
|---|---|---|---|---|
| 04/05/2025 | 2.0 | Added DTG interpretation and expanded volatile content reporting | Annual SOP Review |