in the Analytical Method Development (AMD) laboratory. System suitability tests (SST) confirm that the system is operating correctly before and during analytical runs to ensure method reliability and data integrity.
2. Scope
This procedure applies to all HPLC and GC methods being developed, validated, transferred, or used for routine quality control within the AMD function for raw materials, APIs, and finished products.
3. Responsibilities
- Analyst: Prepares system suitability standards, initiates runs, calculates system suitability parameters.
- AMD Scientist: Defines acceptance criteria based on method development studies and literature references.
- QA: Reviews system suitability trends and approves SST criteria during validation and transfer.
4. Accountability
The Head of Analytical Method Development is accountable for ensuring compliance with SST protocols in accordance with ICH, USP, and regulatory guidelines.
5. Procedure
5.1 Identification of SST Parameters
Select system suitability parameters relevant to the method type and critical performance attributes:
- Retention Time (RT)
- Theoretical Plates (N)
- Resolution (Rs)
- Tailoring Factor (T)
- Repeatability (%RSD) of peak area from multiple injections
5.2 Preparation of SST Standard
- Prepare standard solution at defined concentration according to the analytical method.
- Filter and sonicate the solution if required.
- Ensure that SST standard contains analyte and impurities where applicable.
5.3 Instrumental Setup and Injections
- Use validated or qualified instruments (HPLC/GC) for all SST evaluations.
- Inject 5–6 replicates of SST standard solution to evaluate repeatability and other criteria.
- Ensure column equilibrium and consistent baseline before starting injections.
5.4 Acceptance Criteria
| Parameter |
Typical Acceptance Range |
| % RSD (Peak Area) |
Not more than 2.0% |
| Resolution (Rs) between main peak and impurity |
Not less than 2.0 |
| Theoretical Plates (N) |
As per method development; typically > 2000 |
| Tailoring Factor |
1.0–2.0 |
5.5 Troubleshooting
- If SST criteria fail, do not analyze test samples.
- Investigate root cause (e.g., column condition, mobile phase, injection volume, temperature).
- Document any deviation and recondition the system or replace column if necessary.
5.6 Documentation
- Record SST results in the System Suitability Record Sheet (Annexure-1).
- Attach chromatograms of all injections used for SST calculation.
- Include a justification section in validation or routine reports referring to SST results.
6. Abbreviations
- SST: System Suitability Test
- RT: Retention Time
- N: Theoretical Plates
- Rs: Resolution
- RSD: Relative Standard Deviation
7. Documents
- Annexure-1: System Suitability Record Sheet
- Annexure-2: SST Deviation Log
- Instrument Logbook
8. References
- ICH Q2(R1): Validation of Analytical Procedures
- USP General Chapter <621> Chromatography
- 21 CFR Part 211 – cGMP for Finished Pharmaceuticals
- Internal SOPs on HPLC/GC Qualification
9. SOP Version
Version: 2.0
10. Approval Section
|
Prepared By |
Checked By |
Approved By |
| Signature |
|
|
|
| Date |
|
|
|
| Name |
Anil Jha |
Pooja Mahadik |
Dr. Shravan Kumar |
| Designation |
Senior Analyst |
QA Reviewer |
Head – AMD |
| Department |
Analytical Method Development |
Quality Assurance |
Analytical Method Development |
11. Annexures
Annexure-1: System Suitability Record Sheet
| Injection No. |
Peak Area |
RT (min) |
Resolution |
Tailing Factor |
Plates (N) |
| 1 |
2501 |
5.65 |
2.5 |
1.15 |
3400 |
| 2 |
2503 |
5.64 |
2.6 |
1.14 |
3420 |
| 3 |
2498 |
5.63 |
2.5 |
1.13 |
3390 |
Annexure-2: SST Deviation Log
| Date |
Instrument ID |
Deviation Observed |
Action Taken |
Remarks |
| 31/05/2025 |
HPLC-007 |
RSD above 2.5% |
Reconditioning of column |
Criteria met post re-run |
Revision History
| Revision Date |
Revision No. |
Details |
Reason |
Approved By |
| 01/06/2025 |
2.0 |
Expanded procedure for troubleshooting and tabular criteria |
Annual SOP Review |
Dr. Shravan Kumar |
| 05/06/2022 |
1.0 |
First Release |
New SOP |
QA Head |