SOP Guide for Pharma

Analytical Method Development: SOP for Qualification of Analytical Balances – V 2.0

Analytical Method Development: SOP for Qualification of Analytical Balances – V 2.0

Standard Operating Procedure for Qualification of Analytical Balances in Method Development


Department Analytical Method Development
SOP No. SOP/AMD/347/2025
Supersedes SOP/AMD/347/2022
Page No. Page 1 of 14
Issue Date 01/06/2025
Effective Date 03/06/2025
Review Date 01/06/2026

1. Purpose

The purpose of this SOP is to describe the procedure for the qualification of analytical balances used in the Analytical Method Development (AMD) laboratory. This includes Installation Qualification (IQ), Operational Qualification (OQ), and Performance Qualification

(PQ) to ensure balance reliability, accuracy, and GMP compliance.

2. Scope

This SOP applies to all new or relocated analytical balances intended for use in the AMD department, and for balances undergoing requalification after maintenance or calibration deviations.

3. Responsibilities

  • AMD Analyst: Initiates the qualification protocol and assists in execution.
  • Engineering/Maintenance: Performs installation and provides utility confirmation.
  • QA Representative: Reviews and approves qualification documents.
  • External Vendor (if applicable): Supports traceable calibration and validation services.

4. Accountability

The AMD Manager is accountable for ensuring all balances used in method development are properly qualified, maintained, and documented according to cGMP requirements.

5. Procedure

5.1 Installation Qualification (IQ)

  1. Verify physical condition of the balance upon delivery (model, serial number, accessories).
  2. Install in vibration-free, draft-protected location with calibrated spirit level.
  3. Document environmental conditions such as temperature, humidity, and electromagnetic interference.
  4. Check electrical connections, printer ports, and level bubble alignment.
  5. Affix ‘Qualified Equipment’ label post verification.

5.2 Operational Qualification (OQ)

  1. Ensure internal and external calibration options function as per manufacturer’s instructions.
  2. Verify repeatability using standard certified weights (e.g., 10 mg, 100 mg, 1 g, 10 g).
  3. Perform eccentricity testing by placing weights at different positions on the pan.
  4. Test linearity across the range using certified weights in ascending/descending order.
  5. Document readings, tolerance limits, and observations in Annexure-1.

5.3 Performance Qualification (PQ)

  1. Evaluate day-to-day accuracy using internal calibration and two external certified weights.
  2. Run PQ for three non-consecutive days using same weights by different analysts.
  3. Acceptable criteria: % deviation ≤ ±0.1% from standard weight value.
  4. Ensure environmental stability during performance checks.

5.4 Post Qualification Steps

  1. Compile IQ, OQ, PQ results into a Qualification Summary Report (Annexure-2).
  2. Forward the report to QA for review and sign-off.
  3. Update Equipment Master List and Calibration Plan Register.
  4. Label the balance with “Qualified”, “Calibrated” and “Next Due” stickers.

5.5 Requalification Criteria

  • Relocation to another room/facility
  • Post-maintenance or software upgrade
  • Major deviation in calibration or failed test during routine checks

6. Abbreviations

  • IQ: Installation Qualification
  • OQ: Operational Qualification
  • PQ: Performance Qualification
  • cGMP: Current Good Manufacturing Practice

7. Documents

  1. Annexure-1: OQ Data Sheet
  2. Annexure-2: Qualification Summary Report
  3. Annexure-3: Calibration Certificate (External Vendor)

8. References

  • ICH Q9 – Quality Risk Management
  • USP General Chapter < 41 > Weights and Balances
  • 21 CFR Part 211 – Current GMP for Finished Pharmaceuticals

9. SOP Version

Version: 2.0

10. Approval Section

Prepared By Checked By Approved By
Signature
Date
Name Sneha Joshi Rohit Sharma Dr. Kavita Mehra
Designation AMD Analyst QA Executive Head – AMD
Department Analytical Method Development Quality Assurance Analytical Method Development

11. Annexures

Annexure-1: OQ Data Sheet

Weight Used Nominal Value Observed Value % Deviation Status
Certified Weight 1 10 mg 9.998 mg -0.02% Pass
Certified Weight 2 100 mg 100.001 mg +0.001% Pass

Annexure-2: Qualification Summary Report

Summarizes IQ, OQ, and PQ completion dates, responsible personnel, observations, deviations (if any), and final approval.

Annexure-3: Calibration Certificate

Provided by NABL-accredited third-party calibration agency with traceability details and validity period.

Revision History:

Revision Date Revision No. Details Reason Approved By
01/06/2025 2.0 Added detailed PQ procedure and documentation controls Annual SOP Review Dr. Kavita Mehra
01/04/2022 1.0 Initial Release New SOP QA Head