Analytical Method Development: SOP for Qualification and Verification of Weighing Balances – V 2.0
Standard Operating Procedure for Qualification and Verification of Weighing Balances in Analytical Method Development
| Department |
Analytical Method Development |
| SOP No. |
SOP/AMD/395/2025 |
| Supersedes |
SOP/AMD/395/2022 |
| Page No. |
Page 1 of 14 |
| Issue Date |
01/06/2025 |
| Effective Date |
03/06/2025 |
| Review Date |
01/06/2026 |
1. Purpose
The purpose of this SOP is to define the procedure for the qualification (IQ, OQ, PQ) and routine verification of analytical weighing balances used
in the Analytical Method Development laboratory. This ensures that all balances meet regulatory requirements and provide accurate and precise weighing results.
2. Scope
This SOP applies to all analytical balances, microbalances, and precision balances used for weighing raw materials, standards, samples, and reagents within the AMD department.
3. Responsibilities
- Analytical Personnel: Perform daily verification before use and document results appropriately.
- Engineering/Calibration Team: Conduct qualification and scheduled calibrations as per plan.
- QA: Review qualification documents and ensure compliance with calibration standards.
4. Accountability
The AMD Head and QA Head are accountable for ensuring that all balances in use are qualified, verified, and calibrated per this SOP and documented accordingly.
5. Procedure
5.1 Installation Qualification (IQ)
- Ensure the balance is delivered with all accessories and documentation (user manual, calibration certificate, etc.).
- Install the balance on a vibration-free surface in a designated weighing area with controlled environmental conditions.
- Record environmental parameters (temperature, humidity, airflow) in Annexure-1.
- Verify the availability of power supply and leveling of the instrument.
5.2 Operational Qualification (OQ)
- Use certified standard weights (Class E2 or equivalent) for testing balance performance.
- Perform accuracy, linearity, and repeatability tests as per Annexure-2.
- Record observations for each parameter and compare against acceptance criteria.
5.3 Performance Qualification (PQ)
- Conduct routine usage simulations with weights similar to actual laboratory use conditions.
- Ensure consistent results for the same weights across multiple replicates.
- Document PQ results in Annexure-3.
5.4 Daily Verification Procedure
- Verify zero function and taring before each use.
- Use standard weights (preferably mid-range of the balance capacity).
- Document weight verification on Daily Balance Check Log (Annexure-4).
- Do not use the balance if deviations exceed ±0.1% of the standard weight.
5.5 Calibration and Recalibration
- Ensure external calibration is performed by a NABL-certified agency as per schedule.
- Recalibration is triggered in cases of:
- Balance relocation
- Deviation or malfunction
- After preventive maintenance
5.6 Handling Out-of-Tolerance Conditions
- Stop using the balance immediately.
- Quarantine and tag the instrument as “Out of Service.”
- Report to Engineering and QA for investigation.
- Perform impact assessment of previously weighed materials using the suspect balance.
6. Abbreviations
- IQ: Installation Qualification
- OQ: Operational Qualification
- PQ: Performance Qualification
- NABL: National Accreditation Board for Testing and Calibration Laboratories
7. Documents
- IQ Checklist – Annexure-1
- OQ Test Sheet – Annexure-2
- PQ Validation Report – Annexure-3
- Daily Verification Log – Annexure-4
8. References
- USP General Chapter <41> – Balances
- ICH Q9 – Quality Risk Management
- 21 CFR Part 211 – cGMP for Finished Pharmaceuticals
9. SOP Version
Version: 2.0
10. Approval Section
|
Prepared By |
Checked By |
Approved By |
| Signature |
|
|
|
| Date |
|
|
|
| Name |
Kavita Rane |
Mohit Raj |
Dr. Harshita Goyal |
| Designation |
AMD Analyst |
QA Reviewer |
Head – AMD |
| Department |
Analytical Method Development |
Quality Assurance |
Analytical Method Development |
11. Annexures
Annexure-1: IQ Checklist
| Parameter |
Observation |
Complies (Yes/No) |
| Level surface |
Yes |
Yes |
| Temperature control |
25°C ±2°C |
Yes |
| Power supply |
230V |
Yes |
Annexure-2: OQ Test Sheet
| Weight Used |
Nominal Value |
Observed Value |
% Deviation |
Status |
| Std. Wt 100 mg |
100.000 mg |
99.987 mg |
-0.013% |
Pass |
Annexure-3: PQ Validation Report
Summary of balance performance with real sample weights during actual laboratory usage simulation across 5 replicates. Data falls within acceptance limits.
Annexure-4: Daily Balance Check Log
| Date |
Balance ID |
Weight Used |
Observed Value |
Verified By |
Status |
| 01/06/2025 |
BL-202 |
200 mg |
199.995 mg |
Kavita Rane |
Pass |
Revision History:
| Revision Date |
Revision No. |
Details |
Reason |
Approved By |
| 01/06/2025 |
2.0 |
Expanded qualification sections and clarified recalibration criteria |
Annual review |
Dr. Harshita Goyal |
| 10/08/2022 |
1.0 |
Initial SOP implementation |
New equipment commissioning |
QA Head |