SOP Guide for Pharma

Analytical Method Development: SOP for Qualification and Verification of Weighing Balances – V 2.0

Analytical Method Development: SOP for Qualification and Verification of Weighing Balances – V 2.0

Standard Operating Procedure for Qualification and Verification of Weighing Balances in Analytical Method Development


Department Analytical Method Development
SOP No. SOP/AMD/395/2025
Supersedes SOP/AMD/395/2022
Page No. Page 1 of 14
Issue Date 01/06/2025
Effective Date 03/06/2025
Review Date 01/06/2026

1. Purpose

The purpose of this SOP is to define the procedure for the qualification (IQ, OQ, PQ) and routine verification of analytical weighing balances used

in the Analytical Method Development laboratory. This ensures that all balances meet regulatory requirements and provide accurate and precise weighing results.

2. Scope

This SOP applies to all analytical balances, microbalances, and precision balances used for weighing raw materials, standards, samples, and reagents within the AMD department.

3. Responsibilities

  • Analytical Personnel: Perform daily verification before use and document results appropriately.
  • Engineering/Calibration Team: Conduct qualification and scheduled calibrations as per plan.
  • QA: Review qualification documents and ensure compliance with calibration standards.

4. Accountability

The AMD Head and QA Head are accountable for ensuring that all balances in use are qualified, verified, and calibrated per this SOP and documented accordingly.

5. Procedure

5.1 Installation Qualification (IQ)

  1. Ensure the balance is delivered with all accessories and documentation (user manual, calibration certificate, etc.).
  2. Install the balance on a vibration-free surface in a designated weighing area with controlled environmental conditions.
  3. Record environmental parameters (temperature, humidity, airflow) in Annexure-1.
  4. Verify the availability of power supply and leveling of the instrument.

5.2 Operational Qualification (OQ)

  1. Use certified standard weights (Class E2 or equivalent) for testing balance performance.
  2. Perform accuracy, linearity, and repeatability tests as per Annexure-2.
  3. Record observations for each parameter and compare against acceptance criteria.

5.3 Performance Qualification (PQ)

  1. Conduct routine usage simulations with weights similar to actual laboratory use conditions.
  2. Ensure consistent results for the same weights across multiple replicates.
  3. Document PQ results in Annexure-3.

5.4 Daily Verification Procedure

  1. Verify zero function and taring before each use.
  2. Use standard weights (preferably mid-range of the balance capacity).
  3. Document weight verification on Daily Balance Check Log (Annexure-4).
  4. Do not use the balance if deviations exceed ±0.1% of the standard weight.

5.5 Calibration and Recalibration

  1. Ensure external calibration is performed by a NABL-certified agency as per schedule.
  2. Recalibration is triggered in cases of:
    • Balance relocation
    • Deviation or malfunction
    • After preventive maintenance

5.6 Handling Out-of-Tolerance Conditions

  1. Stop using the balance immediately.
  2. Quarantine and tag the instrument as “Out of Service.”
  3. Report to Engineering and QA for investigation.
  4. Perform impact assessment of previously weighed materials using the suspect balance.

6. Abbreviations

  • IQ: Installation Qualification
  • OQ: Operational Qualification
  • PQ: Performance Qualification
  • NABL: National Accreditation Board for Testing and Calibration Laboratories

7. Documents

  1. IQ Checklist – Annexure-1
  2. OQ Test Sheet – Annexure-2
  3. PQ Validation Report – Annexure-3
  4. Daily Verification Log – Annexure-4

8. References

  • USP General Chapter <41> – Balances
  • ICH Q9 – Quality Risk Management
  • 21 CFR Part 211 – cGMP for Finished Pharmaceuticals

9. SOP Version

Version: 2.0

10. Approval Section

Prepared By Checked By Approved By
Signature
Date
Name Kavita Rane Mohit Raj Dr. Harshita Goyal
Designation AMD Analyst QA Reviewer Head – AMD
Department Analytical Method Development Quality Assurance Analytical Method Development

11. Annexures

Annexure-1: IQ Checklist

Parameter Observation Complies (Yes/No)
Level surface Yes Yes
Temperature control 25°C ±2°C Yes
Power supply 230V Yes

Annexure-2: OQ Test Sheet

Weight Used Nominal Value Observed Value % Deviation Status
Std. Wt 100 mg 100.000 mg 99.987 mg -0.013% Pass

Annexure-3: PQ Validation Report

Summary of balance performance with real sample weights during actual laboratory usage simulation across 5 replicates. Data falls within acceptance limits.

Annexure-4: Daily Balance Check Log

Date Balance ID Weight Used Observed Value Verified By Status
01/06/2025 BL-202 200 mg 199.995 mg Kavita Rane Pass

Revision History:

Revision Date Revision No. Details Reason Approved By
01/06/2025 2.0 Expanded qualification sections and clarified recalibration criteria Annual review Dr. Harshita Goyal
10/08/2022 1.0 Initial SOP implementation New equipment commissioning QA Head
Exit mobile version