Standard Operating Procedure for Qualification and Performance Verification of Analytical Balances in AMD
| Department | Analytical Method Development |
|---|---|
| SOP No. | SOP/AMD/316/2025 |
| Supersedes | SOP/AMD/316/2022 |
| Page No. | Page 1 of 11 |
| Issue Date | 01/06/2025 |
| Effective Date | 03/06/2025 |
| Review Date | 01/06/2026 |
1. Purpose
The purpose of this SOP is to outline the qualification process and periodic performance verification procedures for analytical balances to ensure they provide accurate, consistent, and
2. Scope
This SOP applies to all analytical balances used within the Analytical Method Development (AMD) department for sample and standard preparation requiring high precision weighing.
3. Responsibilities
- Analytical Chemist: Performs daily checks, qualification support, and documents readings.
- Engineering/Maintenance: Provides technical support during installation qualification and preventive maintenance.
- QA Personnel: Verifies qualification documents and ensures protocol compliance.
4. Accountability
The Head of AMD is accountable for ensuring that all analytical balances undergo proper qualification and periodic verification in accordance with the approved schedule and regulatory expectations.
5. Procedure
5.1 Qualification Phases
- Installation Qualification (IQ):
- Ensure placement on vibration-free surface in a controlled environment.
- Verify calibration certificates from manufacturer.
- Connect instrument to stable power source with proper grounding.
- Operational Qualification (OQ):
- Test linearity using certified standard weights (e.g., 1 mg to 200 g).
- Conduct repeatability and drift checks by weighing same mass multiple times.
- Verify leveling, display resolution, and functionality of tare/zero buttons.
- Performance Qualification (PQ):
- Evaluate routine use conditions with typical loads.
- Verify performance over extended weighing cycles.
- Include environmental monitoring during PQ (temperature/humidity).
5.2 Daily Performance Verification
- Verify balance leveling using built-in indicators.
- Record balance ID, date, time, and weight readings in logbook (Annexure-1).
- Use certified weights (e.g., 10 mg, 100 mg, 1 g, 10 g) traceable to NABL/NIST standards.
- Perform weighing in triplicate and document average deviation and % error.
5.3 Acceptance Criteria
- Repeatability: RSD should be ≤ 0.1% for five consecutive measurements.
- Accuracy: Weight difference should not exceed ±0.1 mg for weights ≤ 20 g, and ±0.2 mg for weights above 20 g.
5.4 Preventive Maintenance and Re-Qualification
- Conduct preventive maintenance every 6 months or as per manufacturer guidance.
- Post-maintenance, perform full OQ and PQ before resuming analytical use.
5.5 Deviation Handling
- Record any failed results in the deviation log (Annexure-2).
- Initiate root cause analysis and implement CAPA measures.
- Do not use the balance until successful re-verification is completed.
6. Abbreviations
- IQ: Installation Qualification
- OQ: Operational Qualification
- PQ: Performance Qualification
- RSD: Relative Standard Deviation
- CAPA: Corrective and Preventive Action
7. Documents
- Balance Qualification Protocol – Annexure-1
- Daily Performance Log – Annexure-2
- Deviation & CAPA Register – Annexure-3
8. References
- ICH Q2(R1) – Validation of Analytical Procedures
- USP Chapter <41> Weights and Balances
- 21 CFR Part 211 – GMP Requirements
- GAMP5 – Good Automated Manufacturing Practice
9. SOP Version
Version: 2.0
10. Approval Section
| Prepared By | Checked By | Approved By | |
|---|---|---|---|
| Signature | |||
| Date | |||
| Name | Vikram Joshi | Anita Shah | Dr. Harshita Goyal |
| Designation | Analyst – AMD | QA Reviewer | Department Head |
| Department | Analytical Method Development | Quality Assurance | Analytical Method Development |
11. Annexures
Annexure-1: Balance Qualification Protocol
Includes sections for IQ, OQ, and PQ with checklists, observations, and sign-offs.
Annexure-2: Daily Performance Log
| Date | Balance ID | Weight Used | Reading 1 | Reading 2 | Reading 3 | Avg | RSD (%) | Status |
|---|---|---|---|---|---|---|---|---|
| 01/06/2025 | BL-102 | 100 mg | 100.003 | 100.002 | 100.001 | 100.002 | 0.001 | Pass |
Annexure-3: Deviation & CAPA Register
Logs any failure, deviation number, description, root cause, corrective actions, and final verification status.
Revision History:
| Revision Date | Revision No. | Details | Reason | Approved By |
|---|---|---|---|---|
| 01/06/2025 | 2.0 | Added PQ verification and updated accuracy criteria | Annual Review | Dr. Harshita Goyal |
| 12/08/2022 | 1.0 | Initial SOP Release | New SOP | QA Head |