is applicable to all new or updated analytical methods developed in the Analytical Method Development (AMD) department for drug substances and drug products that require quantitative determination.
3. Responsibilities
- Analytical Chemist: Executes repeatability and intermediate precision tests, compiles results, and performs statistical evaluations.
- QA Reviewer: Verifies the integrity and compliance of the precision study as per protocol.
- AMD Group Leader: Reviews the precision report and provides final authorization for inclusion in validation documentation.
4. Accountability
The AMD Department Head is accountable for ensuring the proper execution and documentation of precision studies as per ICH Q2(R1), USP <1225>, and applicable GMP requirements.
5. Procedure
5.1 Definitions
- Repeatability: Precision under the same operating conditions over a short interval of time, usually by the same analyst, on the same equipment.
- Intermediate Precision: Variation within a laboratory caused by different analysts, equipment, or days.
5.2 Repeatability Testing
- Prepare six replicates of the test solution at the target assay concentration (Annexure-1).
- Perform analysis using the validated method parameters (HPLC, UV, etc.).
- Record retention times, peak areas, and any relevant system suitability data.
- Calculate mean, standard deviation (SD), and relative standard deviation (RSD).
- Acceptance Criteria: RSD should be ≤ 2% unless otherwise justified by product-specific requirements.
5.3 Intermediate Precision Testing
- Repeat the precision study on a different day, with a different analyst and equipment if available.
- Prepare fresh sample solutions and analyze in triplicate at the same concentration levels.
- Combine data from both runs and calculate pooled SD, pooled RSD, and compare variability.
- Ensure system suitability is established in each run as per validated method criteria.
5.4 Data Interpretation
- Compare intra-assay (repeatability) and inter-assay (intermediate) data.
- Evaluate whether variability is within acceptable limits specified in ICH Q2(R1).
- Plot data using control charts or run plots to visualize consistency and outliers.
- If outliers are identified, conduct investigation and repeat analysis with justification.
5.5 Documentation
- Document each test with unique ID numbers and link to the analytical run log (Annexure-2).
- Compile raw data, system suitability data, calculations, and chromatograms in a precision summary report.
- Complete Annexure-3 for data tabulation and statistical evaluation.
6. Abbreviations
- RSD: Relative Standard Deviation
- SD: Standard Deviation
- HPLC: High Performance Liquid Chromatography
- ICH: International Council for Harmonisation
7. Documents
- Annexure-1: Precision Sample Preparation Record
- Annexure-2: Chromatographic Run Log
- Annexure-3: Statistical Summary Sheet
8. References
- ICH Q2(R1) – Validation of Analytical Procedures
- USP General Chapter <1225> – Validation of Compendial Procedures
- 21 CFR Part 211 – Current Good Manufacturing Practice
9. SOP Version
Version: 2.0
10. Approval Section
|
Prepared By |
Checked By |
Approved By |
| Signature |
|
|
|
| Date |
|
|
|
| Name |
Swati Kulkarni |
Ramesh Patil |
Dr. Neeraj Desai |
| Designation |
Analyst – AMD |
QA Reviewer |
Department Head |
| Department |
Analytical Method Development |
Quality Assurance |
Analytical Method Development |
11. Annexures
Annexure-1: Precision Sample Preparation Record
| Sample ID |
Analyst |
Date |
Target Concentration |
Volume Prepared |
| P-361-R1 to R6 |
Swati Kulkarni |
01/06/2025 |
50 µg/mL |
50 mL |
Annexure-2: Chromatographic Run Log
| Sample |
Retention Time |
Peak Area |
System Suitability Pass/Fail |
| P-361-R1 |
6.53 |
105,430 |
Pass |
| P-361-R2 |
6.55 |
104,988 |
Pass |
| P-361-R3 |
6.52 |
105,001 |
Pass |
| P-361-R4 |
6.56 |
105,350 |
Pass |
| P-361-R5 |
6.54 |
104,970 |
Pass |
| P-361-R6 |
6.57 |
105,200 |
Pass |
Annexure-3: Statistical Summary Sheet
| Level |
Mean Area |
SD |
RSD (%) |
| Repeatability |
105,156 |
178 |
0.17% |
| Intermediate Precision |
105,020 |
196 |
0.19% |
Revision History:
| Revision Date |
Revision No. |
Details |
Reason |
Approved By |
| 01/06/2025 |
2.0 |
Expanded section on intermediate precision, added pooled RSD calculations |
Annual Review |
Dr. Neeraj Desai |
| 10/07/2022 |
1.0 |
Initial Issue |
New SOP |
QA Head |