Standard Operating Procedure for Particulate Matter Testing by Light Obscuration in Analytical Method Development
| Department | Analytical Method Development |
|---|---|
| SOP No. | SOP/AMD/268/2025 |
| Supersedes | SOP/AMD/268/2022 |
| Page No. | Page 1 of 11 |
| Issue Date | 21/05/2025 |
| Effective Date | 23/05/2025 |
| Review Date | 21/05/2026 |
1. Purpose
To define a validated and regulatory-compliant method for the determination of particulate matter in injectable formulations using light obscuration particle count test, ensuring
2. Scope
This SOP applies to the Analytical Method Development and Quality Control departments for routine and validation testing of visible and sub-visible particulate matter in large-volume and small-volume parenterals using automatic particle counters.
3. Responsibilities
- Analytical Scientist: Conducts testing using calibrated light obscuration equipment and records results.
- QC Analyst: Prepares test samples, ensures cleanliness of glassware and test environment.
- QA Executive: Verifies results, ensures method alignment with pharmacopeial standards (USP, Ph. Eur.).
4. Accountability
The Head of Analytical Method Development is accountable for ensuring validated methodology, equipment calibration, and compliance with regulatory guidelines on particulate matter control.
5. Procedure
5.1 Equipment and Materials
- Automatic Particle Counter (e.g., HIAC or equivalent)
- Calibrated standards and blank solution (WFI)
- Cleaned sample containers and syringes
- Laminar airflow workstation
5.2 Equipment Calibration
- Calibrate the instrument using polystyrene latex beads traceable to NIST standards.
- Verify detection sensitivity at ≥10 µm and ≥25 µm size ranges.
- Perform system suitability with blank WFI: Should show <10 particles/mL ≥10 µm and <2 particles/mL ≥25 µm.
5.3 Sample Preparation
- Visually inspect vials for visible particles prior to testing.
- Gently invert to homogenize without generating bubbles.
- Withdraw sample aseptically into syringe and transfer to clean test chamber of particle counter.
5.4 Test Conditions and Measurement
- Ensure background particle level in environment is minimal.
- Flush instrument with blank before sample testing.
- Run sample in duplicate; each run must test not less than 5 mL (SVP) or 10 mL (LVP).
- Count particles at ≥10 µm and ≥25 µm thresholds.
5.5 Acceptance Criteria (USP <788>)
- Small-Volume Injections (≤100 mL):
- ≥10 µm: NMT 6000 particles per container
- ≥25 µm: NMT 600 particles per container
- Large-Volume Injections (>100 mL):
- ≥10 µm: NMT 25 particles/mL
- ≥25 µm: NMT 3 particles/mL
5.6 Method Validation Parameters
- Specificity: Blank samples must be free from interference.
- Precision: RSD ≤10% for replicate readings.
- Linearity: Confirm across standard concentrations using PSL beads.
- Robustness: Evaluate instrument performance at varied flow rates and temperatures.
6. Abbreviations
- USP: United States Pharmacopeia
- WFI: Water for Injection
- PSL: Polystyrene Latex
- RSD: Relative Standard Deviation
- QA: Quality Assurance
7. Documents
- Particulate Matter Test Log – Annexure-1
- Calibration Report – Annexure-2
- Validation Summary Report – Annexure-3
8. References
- USP <788>: Particulate Matter in Injections
- Ph. Eur. 2.9.19: Particulate Contamination
- ICH Q6A: Specifications Guidelines
9. SOP Version
Version: 2.0
10. Approval Section
| Prepared By | Checked By | Approved By | |
|---|---|---|---|
| Signature | |||
| Date | |||
| Name | Sandeep Bhaskar | Kavita Jain | Sunita Reddy |
| Designation | Senior Analyst | QA Reviewer | Head – AMD |
| Department | Analytical Method Development | QA | Analytical Method Development |
11. Annexures
Annexure-1: Particulate Matter Test Log
| Sample ID | Volume (mL) | ≥10 µm (Count) | ≥25 µm (Count) | Status |
|---|---|---|---|---|
| INJ-268-01 | 10 | 230 | 12 | Pass |
Annexure-2: Calibration Report
| Date | Standard Size (µm) | Observed Count | Acceptance Range | Status |
|---|---|---|---|---|
| 20/05/2025 | 10.0 | 1480 | 1400–1600 | Pass |
| 20/05/2025 | 25.0 | 245 | 230–270 | Pass |
Annexure-3: Validation Summary Report
The light obscuration method was validated for multiple injection types. Results met USP acceptance limits for particulate counts. Precision (RSD = 5.4%), specificity, and robustness across sample types confirmed. The method is approved for routine release and stability testing.
Revision History:
| Revision Date | Revision No. | Details | Reason | Approved By |
|---|---|---|---|---|
| 21/05/2025 | 2.0 | Aligned with Ph. Eur. 2.9.19 and added new annexures | Annual Review | Sunita Reddy |
| 15/04/2022 | 1.0 | Initial Release | New SOP | QA Head |