methods to ensure product stability and regulatory compliance.
2. Scope
This SOP applies to the Analytical Method Development and Quality Control departments responsible for moisture determination in raw materials, intermediates, and finished powder formulations.
3. Responsibilities
- QC Analyst: Performs moisture content testing using LOD or Karl Fischer titration.
- Analytical Scientist: Develops and validates moisture determination methods.
- QA Executive: Reviews data and ensures compliance with pharmacopeial standards.
4. Accountability
The Head of Analytical Method Development is accountable for ensuring accurate and validated moisture content testing, proper instrument calibration, and adherence to specifications.
5. Procedure
5.1 Method Selection
- Loss on Drying (LOD): Suitable for thermally stable compounds; used for routine analysis.
- Karl Fischer Titration: Preferred for thermolabile or hygroscopic materials and for water-specific quantification.
5.2 Loss on Drying (LOD) Procedure
- Switch on the LOD oven and set temperature as per specification (usually 105°C unless stated otherwise).
- Weigh 1–2 g of sample into a pre-dried and tared LOD dish.
- Dry in the oven for 3 hours or as validated.
- Cool in desiccator and reweigh.
- Calculate % moisture using the formula:
Moisture (%) = [(Initial weight – Final weight) / Initial weight] × 100
5.3 Karl Fischer Titration Procedure
- Use volumetric or coulometric KF titrator based on sample moisture content.
- Weigh 50–300 mg of sample directly into titration vessel (or use methanol solution).
- Titrate with Karl Fischer reagent until endpoint is reached (stable drift).
- Calculate moisture content in % using software or manual calculation based on titrant volume and sample weight.
5.4 Validation Parameters
- Specificity: Ensure no interference from excipients or volatile solvents.
- Precision: RSD ≤ 2.0% for 6 replicates.
- Accuracy: Recovery 98%–102% using water-spiked samples.
- Linearity: For Karl Fischer method over 0.1%–10% moisture.
5.5 Acceptance Criteria
- Moisture content should comply with pharmacopeial or in-house specification (e.g., NMT 5.0%).
6. Abbreviations
- LOD: Loss on Drying
- KF: Karl Fischer
- RSD: Relative Standard Deviation
- QA: Quality Assurance
7. Documents
- Moisture Content Raw Data Sheet – Annexure-1
- LOD Weight Log – Annexure-2
- Karl Fischer Titration Record – Annexure-3
8. References
- USP <921>: Water Determination
- ICH Q2(R1): Validation of Analytical Procedures
- Ph. Eur. 2.5.32: Loss on Drying
9. SOP Version
Version: 2.0
10. Approval Section
|
Prepared By |
Checked By |
Approved By |
| Signature |
|
|
|
| Date |
|
|
|
| Name |
Aditi Thakur |
Rajiv Menon |
Sunita Reddy |
| Designation |
QC Analyst |
QA Reviewer |
Head – AMD |
| Department |
Analytical Method Development |
QA |
Analytical Method Development |
11. Annexures
Annexure-1: Moisture Content Raw Data Sheet
| Sample ID |
Initial Wt. (g) |
Final Wt. (g) |
% Moisture |
Method |
Status |
| MC-289-01 |
1.256 |
1.210 |
3.67% |
LOD |
Pass |
Annexure-2: LOD Weight Log
Includes calibration, blank trial, and sample weights for each run. Oven temperature set and monitored at 105°C ± 2°C.
Annexure-3: Karl Fischer Titration Record
| Sample ID |
Wt. (mg) |
Volume of KF (mL) |
% Water |
Result |
| MC-289-KF01 |
210 |
0.45 |
2.14% |
Pass |
Revision History:
| Revision Date |
Revision No. |
Details |
Reason |
Approved By |
| 21/05/2025 |
2.0 |
Added dual-method approach and updated validation criteria |
Annual Review |
Sunita Reddy |
| 01/12/2022 |
1.0 |
Initial SOP Release |
New SOP |
QA Head |