Skip to content
  • Clinical Studies
  • Schedule M
  • Stability Studies
  • Pharma GMP
  • Pharma Tips
  • Pharma Books
  • Pharma Validation
  • Pharma Regulatory

SOP Guide for Pharma

The Ultimate Resource for Pharmaceutical SOPs and Best Practices

  • Home
  • SOP Guidelines
    • SOP Development
    • SOP Training
    • SOP Compliance Monitoring
    • SOP Revision Processes
    • Pharmaceutical SOP templates
    • GMP documentation SOP
    • Data integrity SOP pharma
    • Manufacturing SOP pharmaceutical
    • FDA SOP guidance
    • eQMS SOP workflows
  • Job Safety Analysis (JSA)
    • Oral Dosage Forms (Tablets & Capsules)
    • Oral Liquid Dosage Forms (Syrups, Elixirs, Suspensions, Emulsions)
    • Powder and Granule Dosage Forms
    • Topical Dosage Forms (Creams, Ointments, Gels, Lotions, Pastes)
    • Transdermal Dosage Forms (Patches)
  • Standard Test Procedures (STP)
  • GMP Audit Findings
    • Non-Compliance
      • SOP Absence
      • Revision Control
      • Non-Adherence
      • Poor Writing
      • Training Failure
      • Data Integrity Gaps
      • Mismatch Between SOPs and Practice
      • Critical Operations
      • Regulatory Change
      • Emergency Changes
      • Third-Party SOPs
      • Inspection Readiness
      • Deviation Handling
      • CAPA Integration
      • Validation Alignment
      • System Integration
      • Uncontrolled Distribution
      • SOP Implementation Gaps
      • SOP Alignment with Validation
      • Change Control Linkage
  • Toggle search form

Analytical Method Development: SOP for Documentation of Chromatographic System Suitability Parameters – V 2.0

Posted on By

Analytical Method Development: SOP for Documentation of Chromatographic System Suitability Parameters – V 2.0

Standard Operating Procedure for Documentation of Chromatographic System Suitability Parameters


Department Analytical Method Development
SOP No. SOP/AMD/391/2025
Supersedes SOP/AMD/391/2022
Page No. Page 1 of 12
Issue Date 01/06/2025
Effective Date 03/06/2025
Review Date 01/06/2026

1. Purpose

The purpose of this SOP is to establish a consistent process for documenting chromatographic system suitability parameters, such as theoretical plates, tailing factor, resolution, and repeatability, during analytical method development and validation activities.

2. Scope

This SOP

applies to all personnel performing HPLC, UPLC, or GC-based chromatographic analyses in the Analytical Method Development (AMD) laboratory, where system suitability tests (SST) are mandatory to ensure analytical system performance.

3. Responsibilities

  • Analyst: Executes system suitability runs and records parameters as per method SOP.
  • AMD Scientist: Verifies acceptability of SST results and authorizes method execution.
  • QA Reviewer: Audits documentation for completeness and ensures compliance with regulatory standards.

4. Accountability

The Head of AMD is accountable for implementation and control of this SOP, ensuring all chromatographic SST data are properly reviewed, documented, and retained.

See also  Analytical Method Development: Selection of Analytical Techniques - V 2.0

5. Procedure

5.1 Understanding System Suitability Tests

  1. System suitability tests are conducted to verify that the analytical system is adequate for performing the intended analysis.
  2. Parameters assessed typically include:
    • Theoretical plates (N)
    • Tailing factor (T)
    • Resolution (Rs)
    • Repeatability (%RSD of peak area or retention time)
  3. Acceptance criteria must be predefined in the method SOP based on validation studies or pharmacopeial recommendations.

5.2 Preparation for Chromatographic Runs

  1. Ensure the system (e.g., HPLC/UPLC/GC) is calibrated and qualified.
  2. Use freshly prepared mobile phase and equilibrate the system as per method conditions.
  3. Inject standard solution as defined and allow system stabilization if required.

5.3 Recording System Suitability Parameters

  1. After successful injection of standards:
    • Extract chromatograms and system-generated parameters from the software.
    • Record the following manually or via electronic data:
      • Retention time (Rt)
      • Peak area
      • Tailing factor
      • Theoretical plates
      • Resolution between critical peaks
      • Baseline noise (if applicable)
  2. Enter these parameters into the System Suitability Log Sheet (Annexure-1).
See also  Analytical Method Development: SOP for Apparatus Suitability and RPM Optimization - V 2.0

5.4 Evaluation of System Suitability

  1. Compare recorded SST values against the method-specific acceptance criteria.
  2. If any parameter fails to meet criteria:
    • Do not proceed with sample injections.
    • Notify AMD Scientist and initiate investigation as per SOP for OOS/OOT handling.
  3. If parameters pass:
    • Document ‘System suitability criteria met’ on analytical worksheet.
    • Proceed with analysis of test samples.

5.5 Data Integrity and Audit Trail

  1. All chromatographic data must be stored securely within validated CDS software.
  2. Ensure audit trail captures all modifications and re-integrations.
  3. Printouts or electronic records should include chromatograms and summary reports with SST values.

5.6 Review and Approval

  1. AMD Scientist to review and sign off on SST data and analytical worksheet before submission to QA.
  2. QA to verify SST documentation as part of batch release or validation packet review.

6. Abbreviations

  • SST: System Suitability Test
  • HPLC: High-Performance Liquid Chromatography
  • GC: Gas Chromatography
  • UPLC: Ultra Performance Liquid Chromatography
  • RSD: Relative Standard Deviation

7. Documents

  1. System Suitability Log Sheet – Annexure-1
  2. Chromatographic Data Review Checklist – Annexure-2
  3. Instrument Audit Trail Verification Log – Annexure-3
See also  Analytical Method Development: SOP for Dissolution Method for Coated Beads in Capsules - V 2.0

8. References

  • ICH Q2(R1) – Validation of Analytical Procedures
  • 21 CFR Part 211 – Current Good Manufacturing Practices
  • USP General Chapter <621> – Chromatography

9. SOP Version

Version: 2.0

10. Approval Section

Prepared By Checked By Approved By
Signature
Date
Name Ankita Sharma Dr. Ravi Chandra Dr. Harshita Goyal
Designation Analyst QA Reviewer Head – AMD
Department Analytical Method Development Quality Assurance Analytical Method Development

11. Annexures

Annexure-1: System Suitability Log Sheet

Template to record retention time, theoretical plates, resolution, tailing factor, and RSD for standard injections.

Annexure-2: Chromatographic Data Review Checklist

Checklist includes items such as proper labeling, complete chromatograms, SST values, injection sequence, and calibration curve verification.

Annexure-3: Instrument Audit Trail Verification Log

Record audit trail checks performed before and after analysis, including reviewer name, date, observations, and comments.

Revision History:

Revision Date Revision No. Details Reason Approved By
01/06/2025 2.0 Updated for audit trail and integrity checks Annual SOP review Dr. Harshita Goyal
10/08/2022 1.0 Initial release New SOP QA Head
Analytical Method Development V 2.0 Tags:Analytical Method Development SOP, SOP for analytical calculations, SOP for analytical documentation, SOP for analytical instrument calibration, SOP for analytical instrument qualification, SOP for analytical method change control, SOP for analytical method deviation handling, SOP for analytical method documentation practices, SOP for analytical method lifecycle management, SOP for analytical method optimization, SOP for analytical method risk assessment, SOP for analytical method training and competency, SOP for analytical method transfer protocol, SOP for analytical method validation, SOP for forced degradation studies, SOP for HPLC method development, SOP for limit of detection (LOD) determination, SOP for limit of quantitation (LOQ) determination, SOP for method accuracy testing, SOP for method linearity assessment, SOP for method precision evaluation, SOP for method reproducibility, SOP for method robustness testing, SOP for method transfer, SOP for method verification, SOP for qualification of working standards, SOP for reference standard handling, SOP for specificity testing in analytical methods, SOP for stability indicating methods, SOP for system suitability testing

Post navigation

Previous Post: Analytical Method Development: SOP for Handling Out of Trend (OOT) Results – V 2.0
Next Post: Analytical Method Development: SOP for Verification of Analytical Calculations – V 2.0

Standard Operating Procedures V 1.0

  • Aerosols
  • Analytical Method Development
  • Bioequivalence Bioavailability Study
  • Capsule Formulation
  • Clinical Studies
  • Creams
  • Data Integrity
  • Dental Dosage Forms
  • Drug Discovery
  • Environment, Health and Safety
  • Formulation Development
  • Gels
  • Good Distribution Practice
  • Good Warehousing Practices
  • In-Process Control
  • Injectables
  • Liquid Orals
  • Liposome and Emulsion Formulations
  • Lotions
  • Lyophilized Products
  • Maintenance Dept.
  • Medical Devices
  • Metered-Dose Inhaler
  • Microbiology Testing
  • Nanoparticle Formulation
  • Nasal Spray Formulations
  • Nebulizers
  • Ocular (Eye) Dosage Forms
  • Ointments
  • Otic (Ear) Dosage Forms
  • Pharmacovigilance
  • Powder & Granules
  • Purchase Departments
  • Quality Assurance
  • Quality Control
  • Raw Material Stores
  • Regulatory Affairs
  • Tablet Manufacturing
  • Rectal Dosage Forms
  • Transdermal Patches
  • Vaginal Dosage Forms
  • Validations and Qualifications

Read SOPs in your Language:

 - 
Bengali
 - 
bn
English
 - 
en
Gujarati
 - 
gu
Hindi
 - 
hi
Malayalam
 - 
ml
Marathi
 - 
mr
Punjabi
 - 
pa
Tamil
 - 
ta
Telugu
 - 
te

NEW! Revised SOPs – V 2.0

  • Aerosols V 2.0
  • Analytical Method Development V 2.0
  • API Manufacturing V 2.0
  • BA-BE Studies V 2.0
  • Biosimilars V 2.0
  • Capsules V 2.0
  • Creams V 2.0
  • Elixers V 2.0
  • Gels V 2.0
  • Injectables V 2.0
  • Ointments V 2.0
  • Raw Material Warehouse V 2.0
  • Tablet Manufacturing V2.0

New Publication: A must for All.

Copyright © 2025 SOP Guide for Pharma.

Powered by PressBook WordPress theme

Go to mobile version