Standard Operating Procedure for Delivered Dose Uniformity (DDU) Testing Method in Analytical Method Development


Department Analytical Method Development
SOP No. SOP/AMD/242/2025
Supersedes SOP/AMD/242/2022
Page No. Page 1 of 13
Issue Date 21/05/2025
Effective Date 23/05/2025
Review Date 21/05/2026

1. Purpose

To define the standardized procedure for performing Delivered Dose Uniformity (DDU) testing of nasal sprays, metered-dose inhalers (MDIs), and other unit-dose aerosol products to

ensure dose consistency within and between units.

2. Scope

This SOP applies to all analytical scientists within the Analytical Method Development (AMD) department conducting uniformity testing of inhalation and nasal spray dosage forms during development, validation, and batch release activities.

3. Responsibilities

  • Analytical Scientist: Executes the DDU testing using validated methods and approved apparatus.
  • QC Analyst: Assists in sample preparation, weighing, and solution analysis.
  • QA Executive: Reviews test data, validates results, and ensures compliance with regulatory limits.

4. Accountability

The Head of Analytical Method Development is accountable for the validity, reproducibility, and regulatory compliance of the DDU testing procedure.

5. Procedure

5.1 Apparatus and Equipment

  1. DDU Sampling Apparatus (e.g., Dose Uniformity Sampling Apparatus – DUSA)
  2. Vacuum source with flow regulator and trap
  3. Analytical balance (0.1 mg sensitivity)
  4. Volumetric flasks, HPLC vials, and appropriate diluent

5.2 Sample Selection and Preparation

  1. Select 10–20 units from the batch based on statistical design.
  2. Shake each unit as per manufacturer’s instructions (e.g., 10 times vertically).
  3. Prime the device as recommended (e.g., 3 sprays to waste).

5.3 Dose Collection Procedure

  1. Actuate device into DUSA for defined number of sprays (typically 1 spray per test per unit).
  2. Rinse collection chamber with diluent and transfer to volumetric flask.
  3. Make up to volume and analyze by validated HPLC or UV method.

5.4 Analytical Testing

  1. Prepare standard solution of known concentration.
  2. Run both standard and sample on validated HPLC system.
  3. Calculate delivered dose using peak area ratio:

    Delivered Dose (mg) = (Sample Peak Area / Standard Peak Area) × Standard Concentration × Dilution Factor

5.5 Acceptance Criteria

  1. All units must fall within 80–120% of label claim.
  2. % RSD of delivered doses ≤ 6.0% for 10 units.
  3. Any outlier to be justified or repeated as per protocol.

5.6 Validation Requirements

  1. Validate method for:
    • Linearity (r ≥ 0.99)
    • Accuracy (90–110% recovery)
    • Precision (intra- and inter-day)
    • Robustness (variation in actuation, temperature)

5.7 Documentation

  1. Maintain raw data printouts, chromatograms, instrument logs, and data sheets in Annexures.
  2. Summarize data in DDU Validation Report with graphical representation.

6. Abbreviations

  • DDU: Delivered Dose Uniformity
  • DUSA: Dose Uniformity Sampling Apparatus
  • HPLC: High Performance Liquid Chromatography
  • RSD: Relative Standard Deviation

7. Documents

  1. DDU Data Log – Annexure-1
  2. DDU Sample Preparation Sheet – Annexure-2
  3. DDU Validation Report – Annexure-3

8. References

  • USP <601>: Aerosols, Nasal Sprays, Metered-Dose Inhalers
  • FDA Guidance for Industry: Nasal Spray and Inhalation Solution Testing
  • ICH Q2(R1): Validation of Analytical Procedures

9. SOP Version

Version: 2.0

10. Approval Section

Prepared By Checked By Approved By
Signature
Date
Name Nikita Shah Rahul Chauhan Sunita Reddy
Designation Analytical Scientist QA Reviewer Head – AMD
Department Analytical Method Development QA Analytical Method Development

11. Annexures

Annexure-1: DDU Data Log

Unit No. Delivered Dose (mg) % of Label Claim Status
1 0.101 101% Pass
2 0.099 99% Pass
3 0.103 103% Pass

Annexure-2: Sample Preparation Sheet

Sample ID Device Number of Sprays Diluent Dilution Volume (mL)
DDU-NS001 NASAL-03 1 Water + MeOH 25

Annexure-3: DDU Validation Report

Delivered dose uniformity method for nasal spray metered device validated using HPLC at 220 nm. 10 units tested showed dose range 97%–104% of label claim, RSD 3.2%. Method considered suitable for routine and release testing.

Revision History:

Revision Date Revision No. Details Reason Approved By
21/05/2025 2.0 Added robustness and validation criteria per FDA guidance Annual Review and Regulatory Update Sunita Reddy
15/04/2022 1.0 Initial SOP Release New SOP QA Head