Standard Operating Procedure for Calibration of UV-Visible Spectrophotometer in Analytical Method Development
| Department | Analytical Method Development |
|---|---|
| SOP No. | SOP/AMD/315/2025 |
| Supersedes | SOP/AMD/315/2022 |
| Page No. | Page 1 of 11 |
| Issue Date | 01/06/2025 |
| Effective Date | 03/06/2025 |
| Review Date | 01/06/2026 |
1. Purpose
The purpose of this SOP is to provide detailed instructions for the calibration of UV-Visible spectrophotometers to ensure accurate wavelength, photometric, and stray light performance for reliable analytical testing in method development.
2. Scope
This SOP applies
to all UV-Visible spectrophotometers used in the Analytical Method Development (AMD) department for both qualitative and quantitative analysis of pharmaceutical substances.
3. Responsibilities
- Analytical Chemist: Performs calibration activities, records data, and ensures instruments are within specifications.
- QA Personnel: Reviews calibration results and ensures compliance with regulatory requirements.
- AMD Head: Approves calibration reports and ensures timely execution of schedule.
4. Accountability
The AMD Section Head is accountable for ensuring all UV-Visible spectrophotometers are calibrated as per schedule and documentation is reviewed and approved as per GMP guidelines.
5. Procedure
5.1 Calibration Frequency
- Perform calibration monthly and after major maintenance or relocation of the instrument.
- Record all calibration activities in the UV Calibration Logbook (Annexure-1).
5.2 Wavelength Accuracy Check
- Use Holmium Perchlorate solution or a certified UV wavelength standard.
- Scan spectrum between 200 nm and 600 nm using a 1 cm quartz cell.
- Verify the presence of characteristic peaks (e.g., 241.15 nm, 279.4 nm, 287.15 nm).
- Acceptance Criteria: ±1.0 nm from certified reference values.
5.3 Photometric Accuracy Check
- Prepare potassium dichromate solution in 0.005M H2SO4 at 60, 40, 20, and 10 ppm.
- Measure absorbance at 235, 257, 313, and 350 nm against the solvent blank.
- Acceptance Criteria: Absorbance values must be within ±0.01 AU of expected values for each concentration.
5.4 Stray Light Test
- Use a 1.2% KCl solution and scan at 200 nm.
- Acceptance Criteria: Absorbance should be greater than 2.0 AU.
5.5 Resolution Test
- Use Toluene in Hexane and scan from 260–280 nm.
- Calculate resolution using absorbance at 269 nm and 266 nm.
- Acceptance Criteria: Ratio A269/A266 ≥ 1.5.
5.6 Documentation
- Record all observations, spectra, and absorbance values in Annexure-2.
- Fill calibration checklist (Annexure-3) and attach with report.
- Obtain QA review and approval before using instrument for analysis.
5.7 Handling Out of Calibration (OOC)
- If any parameter is out of acceptance criteria, stop use of the instrument immediately.
- Log deviation and initiate investigation as per deviation SOP.
- Perform maintenance and recalibrate before resuming use.
6. Abbreviations
- UV: Ultraviolet
- AU: Absorbance Unit
- QA: Quality Assurance
- OOC: Out of Calibration
- AMD: Analytical Method Development
7. Documents
- UV Calibration Logbook – Annexure-1
- Calibration Raw Data Sheet – Annexure-2
- Calibration Checklist – Annexure-3
8. References
- ICH Q2(R1) – Validation of Analytical Procedures
- USP General Chapter <857> – Ultraviolet-Visible Spectroscopy
- 21 CFR Part 211 – GMP for Finished Pharmaceuticals
9. SOP Version
Version: 2.0
10. Approval Section
| Prepared By | Checked By | Approved By | |
|---|---|---|---|
| Signature | |||
| Date | |||
| Name | Neha Mehta | Priya Nair | Dr. Harshita Goyal |
| Designation | AMD Analyst | QA Reviewer | Head – Analytical Method Development |
| Department | Analytical Method Development | Quality Assurance | Analytical Method Development |
11. Annexures
Annexure-1: UV Calibration Logbook
Includes calibration dates, performed by, instrument ID, and remarks for traceability.
Annexure-2: Calibration Raw Data Sheet
Scanned spectra, absorbance readings, calculation tables, and peak listings for each parameter tested.
Annexure-3: Calibration Checklist
| Parameter | Test Performed | Within Limits | Remarks |
|---|---|---|---|
| Wavelength Accuracy | Yes | Yes | Complies |
| Photometric Accuracy | Yes | Yes | Complies |
| Stray Light | Yes | Yes | Complies |
| Resolution | Yes | Yes | Complies |
Revision History:
| Revision Date | Revision No. | Details | Reason | Approved By |
|---|---|---|---|---|
| 01/06/2025 | 2.0 | Added resolution check and updated annexures | Periodic SOP review | Dr. Harshita Goyal |
| 10/08/2022 | 1.0 | Initial release | New SOP | QA Head |