SOP Guide for Pharma

Analytical Method Development: SOP for Calibration of UV-Visible Spectrophotometer – V 2.0

Analytical Method Development: SOP for Calibration of UV-Visible Spectrophotometer – V 2.0

Standard Operating Procedure for Calibration of UV-Visible Spectrophotometer in Analytical Method Development


Department Analytical Method Development
SOP No. SOP/AMD/315/2025
Supersedes SOP/AMD/315/2022
Page No. Page 1 of 11
Issue Date 01/06/2025
Effective Date 03/06/2025
Review Date 01/06/2026

1. Purpose

The purpose of this SOP is to provide detailed instructions for the calibration of UV-Visible spectrophotometers to ensure accurate wavelength, photometric, and stray light performance for reliable analytical testing in method development.

2. Scope

This SOP applies

to all UV-Visible spectrophotometers used in the Analytical Method Development (AMD) department for both qualitative and quantitative analysis of pharmaceutical substances.

3. Responsibilities

  • Analytical Chemist: Performs calibration activities, records data, and ensures instruments are within specifications.
  • QA Personnel: Reviews calibration results and ensures compliance with regulatory requirements.
  • AMD Head: Approves calibration reports and ensures timely execution of schedule.

4. Accountability

The AMD Section Head is accountable for ensuring all UV-Visible spectrophotometers are calibrated as per schedule and documentation is reviewed and approved as per GMP guidelines.

5. Procedure

5.1 Calibration Frequency

  1. Perform calibration monthly and after major maintenance or relocation of the instrument.
  2. Record all calibration activities in the UV Calibration Logbook (Annexure-1).

5.2 Wavelength Accuracy Check

  1. Use Holmium Perchlorate solution or a certified UV wavelength standard.
  2. Scan spectrum between 200 nm and 600 nm using a 1 cm quartz cell.
  3. Verify the presence of characteristic peaks (e.g., 241.15 nm, 279.4 nm, 287.15 nm).
  4. Acceptance Criteria: ±1.0 nm from certified reference values.

5.3 Photometric Accuracy Check

  1. Prepare potassium dichromate solution in 0.005M H2SO4 at 60, 40, 20, and 10 ppm.
  2. Measure absorbance at 235, 257, 313, and 350 nm against the solvent blank.
  3. Acceptance Criteria: Absorbance values must be within ±0.01 AU of expected values for each concentration.

5.4 Stray Light Test

  1. Use a 1.2% KCl solution and scan at 200 nm.
  2. Acceptance Criteria: Absorbance should be greater than 2.0 AU.

5.5 Resolution Test

  1. Use Toluene in Hexane and scan from 260–280 nm.
  2. Calculate resolution using absorbance at 269 nm and 266 nm.
  3. Acceptance Criteria: Ratio A269/A266 ≥ 1.5.

5.6 Documentation

  1. Record all observations, spectra, and absorbance values in Annexure-2.
  2. Fill calibration checklist (Annexure-3) and attach with report.
  3. Obtain QA review and approval before using instrument for analysis.

5.7 Handling Out of Calibration (OOC)

  1. If any parameter is out of acceptance criteria, stop use of the instrument immediately.
  2. Log deviation and initiate investigation as per deviation SOP.
  3. Perform maintenance and recalibrate before resuming use.

6. Abbreviations

  • UV: Ultraviolet
  • AU: Absorbance Unit
  • QA: Quality Assurance
  • OOC: Out of Calibration
  • AMD: Analytical Method Development

7. Documents

  1. UV Calibration Logbook – Annexure-1
  2. Calibration Raw Data Sheet – Annexure-2
  3. Calibration Checklist – Annexure-3

8. References

  • ICH Q2(R1) – Validation of Analytical Procedures
  • USP General Chapter <857> – Ultraviolet-Visible Spectroscopy
  • 21 CFR Part 211 – GMP for Finished Pharmaceuticals

9. SOP Version

Version: 2.0

10. Approval Section

Prepared By Checked By Approved By
Signature
Date
Name Neha Mehta Priya Nair Dr. Harshita Goyal
Designation AMD Analyst QA Reviewer Head – Analytical Method Development
Department Analytical Method Development Quality Assurance Analytical Method Development

11. Annexures

Annexure-1: UV Calibration Logbook

Includes calibration dates, performed by, instrument ID, and remarks for traceability.

Annexure-2: Calibration Raw Data Sheet

Scanned spectra, absorbance readings, calculation tables, and peak listings for each parameter tested.

Annexure-3: Calibration Checklist

Parameter Test Performed Within Limits Remarks
Wavelength Accuracy Yes Yes Complies
Photometric Accuracy Yes Yes Complies
Stray Light Yes Yes Complies
Resolution Yes Yes Complies

Revision History:

Revision Date Revision No. Details Reason Approved By
01/06/2025 2.0 Added resolution check and updated annexures Periodic SOP review Dr. Harshita Goyal
10/08/2022 1.0 Initial release New SOP QA Head
Exit mobile version