  • Standard Test Procedures (STP)
  • GMP Audit Findings
    • Non-Compliance
      • SOP Absence
      • Revision Control
      • Non-Adherence
      • Poor Writing
      • Training Failure
      • Data Integrity Gaps
      • Mismatch Between SOPs and Practice
      • Critical Operations
      • Regulatory Change
      • Emergency Changes
      • Third-Party SOPs
      • Inspection Readiness
      • Deviation Handling
      • CAPA Integration
      • Validation Alignment
      • System Integration
      • Uncontrolled Distribution
      • SOP Implementation Gaps
      • SOP Alignment with Validation
      • Change Control Linkage
  • Toggle search form

Analytical Method Development: SOP for Calibration of UV-Visible Spectrophotometer – V 2.0

Posted on By

Analytical Method Development: SOP for Calibration of UV-Visible Spectrophotometer – V 2.0

Standard Operating Procedure for Calibration of UV-Visible Spectrophotometer in Analytical Method Development


Department Analytical Method Development
SOP No. SOP/AMD/343/2025
Supersedes SOP/AMD/343/2022
Page No. Page 1 of 10
Issue Date 01/06/2025
Effective Date 03/06/2025
Review Date 01/06/2026

1. Purpose

This SOP outlines the step-by-step procedure for the calibration of UV-Visible spectrophotometers to ensure accurate and reproducible results in analytical testing. Calibration ensures the instrument complies with regulatory and pharmacopoeial requirements

for absorbance accuracy, wavelength accuracy, stray light, and baseline flatness.

2. Scope

This SOP applies to all UV-Visible spectrophotometers used for analytical method development, validation, and routine quality control within the organization.

3. Responsibilities

  • Analytical Chemist: Performs calibration and records data in the calibration logbook.
  • QA Representative: Reviews and verifies the calibration data for compliance.
  • Instrument Supervisor: Ensures timely calibration, troubleshooting, and maintenance.

4. Accountability

The Head of Analytical Method Development is accountable for ensuring the calibration of all UV-Visible spectrophotometers is performed as per this SOP and that the records are maintained appropriately.

See also  Analytical Method Development: SOP for Linearity Evaluation in Analytical Methods - V 2.0

5. Procedure

5.1 Calibration Frequency

  1. Perform calibration quarterly or after any major maintenance.
  2. Calibration should also be performed if abnormal results or instrument errors are observed.

5.2 Equipment and Materials Required

  • UV-Visible Spectrophotometer
  • Certified Holmium Oxide Filter or Solution
  • 0.005 M Potassium Dichromate Solution
  • Distilled Water (Blank)
  • Quartz Cuvettes
  • Validated calibration logbook (Annexure-1)

5.3 Calibration Parameters and Acceptance Criteria

  1. Wavelength Accuracy: Use holmium oxide filter/solution. Acceptable variation: ±1 nm from certified value.
  2. Absorbance Accuracy: Prepare 0.005 M potassium dichromate and measure at 235 nm, 257 nm, 313 nm, and 350 nm. Compare with acceptance limits as per pharmacopeial standards.
  3. Stray Light: Use potassium chloride solution. Absorbance at 200 nm should be ≥2.0.
  4. Baseline Flatness: Scan 200–400 nm with distilled water. Fluctuation in absorbance should not exceed ±0.01 AU.

5.4 Calibration Steps

  1. Switch ON the UV-Visible spectrophotometer and allow warm-up for 30 minutes.
  2. Clean cuvettes and ensure they are free from scratches and fingerprints.
  3. Zero the instrument with blank (distilled water).
  4. Carry out wavelength calibration using the certified holmium oxide standard.
  5. Record the peak maxima and compare with certified values.
  6. Prepare and measure absorbance of potassium dichromate solution at specified wavelengths.
  7. Record stray light absorbance using 1% potassium chloride at 200 nm.
  8. Run a full scan for baseline flatness from 200 to 400 nm with blank cuvette.
  9. Record all results in the instrument calibration logbook.
See also  Analytical Method Development: Wavelength Scanning and Selection SOP - V 2.0

5.5 Documentation

  1. Fill Annexure-1 with all calibration readings and analyst signature.
  2. Submit for QA verification and sign-off.
  3. Paste one copy in the instrument area and another in the QA documentation room.

5.6 Out of Calibration (OOC) Handling

  1. If any parameter fails, immediately label instrument as “OUT OF ORDER”.
  2. Initiate deviation form and perform root cause analysis.
  3. Perform recalibration after corrective action and document accordingly.

6. Abbreviations

  • QA: Quality Assurance
  • OOC: Out of Calibration
  • AU: Absorbance Unit
  • nm: Nanometer

7. Documents

  1. Annexure-1: UV-Visible Spectrophotometer Calibration Log
  2. Annexure-2: Deviation Form Template

8. References

  • USP General Chapter <857> – Ultraviolet-Visible Spectroscopy
  • ICH Q2(R1) – Validation of Analytical Procedures
  • Instrument Operation Manual

9. SOP Version

Version: 2.0

10. Approval Section

Prepared By Checked By Approved By
Signature
Date
Name Sanjana Iyer Rahul Deshmukh Dr. Smita Narula
Designation Analyst QA Reviewer Head – AMD
Department Analytical Method Development Quality Assurance Analytical Method Development

11. Annexures

Annexure-1: UV-Visible Spectrophotometer Calibration Log

Date Wavelength Accuracy Absorbance Accuracy Stray Light Baseline Flatness Remarks
01/06/2025 Within ±1 nm Pass 2.2 AU ±0.008 AU All parameters within limits

Annexure-2: Deviation Form Template

This form captures date, deviation ID, description, probable cause, corrective action, and QA comments.

Revision History:

Revision Date Revision No. Details Reason Approved By
01/06/2025 2.0 Stray light and baseline flatness sections revised Annual review and regulatory update Dr. Smita Narula
15/07/2022 1.0 Initial Release New SOP QA Head
See also  Analytical Method Development: SOP for Apparatus Suitability and RPM Optimization - V 2.0
Analytical Method Development V 2.0 Tags:Analytical Method Development SOP, SOP for analytical calculations, SOP for analytical documentation, SOP for analytical instrument calibration, SOP for analytical instrument qualification, SOP for analytical method change control, SOP for analytical method deviation handling, SOP for analytical method documentation practices, SOP for analytical method lifecycle management, SOP for analytical method optimization, SOP for analytical method risk assessment, SOP for analytical method training and competency, SOP for analytical method transfer protocol, SOP for analytical method validation, SOP for forced degradation studies, SOP for HPLC method development, SOP for limit of detection (LOD) determination, SOP for limit of quantitation (LOQ) determination, SOP for method accuracy testing, SOP for method linearity assessment, SOP for method precision evaluation, SOP for method reproducibility, SOP for method robustness testing, SOP for method transfer, SOP for method verification, SOP for qualification of working standards, SOP for reference standard handling, SOP for specificity testing in analytical methods, SOP for stability indicating methods, SOP for system suitability testing

Post navigation

Previous Post: Elixir Department: SOP for Sample Retention Procedure – V 2.0
Next Post: Analytical Method Development: SOP for Qualification of HPLC System – V 2.0

Standard Operating Procedures V 1.0

  • Aerosols
  • Analytical Method Development
  • Bioequivalence Bioavailability Study
  • Capsule Formulation
  • Clinical Studies
  • Creams
  • Data Integrity
  • Dental Dosage Forms
  • Drug Discovery
  • Environment, Health and Safety
  • Formulation Development
  • Gels
  • Good Distribution Practice
  • Good Warehousing Practices
  • In-Process Control
  • Injectables
  • Liquid Orals
  • Liposome and Emulsion Formulations
  • Lotions
  • Lyophilized Products
  • Maintenance Dept.
  • Medical Devices
  • Metered-Dose Inhaler
  • Microbiology Testing
  • Nanoparticle Formulation
  • Nasal Spray Formulations
  • Nebulizers
  • Ocular (Eye) Dosage Forms
  • Ointments
  • Otic (Ear) Dosage Forms
  • Pharmacovigilance
  • Powder & Granules
  • Purchase Departments
  • Quality Assurance
  • Quality Control
  • Raw Material Stores
  • Regulatory Affairs
  • Tablet Manufacturing
  • Rectal Dosage Forms
  • Transdermal Patches
  • Vaginal Dosage Forms
  • Validations and Qualifications

Read SOPs in your Language:

 - 
Bengali
 - 
bn
English
 - 
en
Gujarati
 - 
gu
Hindi
 - 
hi
Malayalam
 - 
ml
Marathi
 - 
mr
Punjabi
 - 
pa
Tamil
 - 
ta
Telugu
 - 
te

NEW! Revised SOPs – V 2.0

  • Aerosols V 2.0
  • Analytical Method Development V 2.0
  • API Manufacturing V 2.0
  • BA-BE Studies V 2.0
  • Biosimilars V 2.0
  • Capsules V 2.0
  • Creams V 2.0
  • Elixers V 2.0
  • Gels V 2.0
  • Injectables V 2.0
  • Ointments V 2.0
  • Raw Material Warehouse V 2.0
  • Tablet Manufacturing V2.0

New Publication: A must for All.

Copyright © 2026 SOP Guide for Pharma.

Powered by PressBook WordPress theme

Go to mobile version