Standard Operating Procedure for Calibration of UV-Visible Spectrophotometer in Analytical Method Development
| Department | Analytical Method Development |
|---|---|
| SOP No. | SOP/AMD/343/2025 |
| Supersedes | SOP/AMD/343/2022 |
| Page No. | Page 1 of 10 |
| Issue Date | 01/06/2025 |
| Effective Date | 03/06/2025 |
| Review Date | 01/06/2026 |
1. Purpose
This SOP outlines the step-by-step procedure for the calibration of UV-Visible spectrophotometers to ensure accurate and reproducible results in analytical testing. Calibration ensures the instrument complies with regulatory and pharmacopoeial requirements
2. Scope
This SOP applies to all UV-Visible spectrophotometers used for analytical method development, validation, and routine quality control within the organization.
3. Responsibilities
- Analytical Chemist: Performs calibration and records data in the calibration logbook.
- QA Representative: Reviews and verifies the calibration data for compliance.
- Instrument Supervisor: Ensures timely calibration, troubleshooting, and maintenance.
4. Accountability
The Head of Analytical Method Development is accountable for ensuring the calibration of all UV-Visible spectrophotometers is performed as per this SOP and that the records are maintained appropriately.
5. Procedure
5.1 Calibration Frequency
- Perform calibration quarterly or after any major maintenance.
- Calibration should also be performed if abnormal results or instrument errors are observed.
5.2 Equipment and Materials Required
- UV-Visible Spectrophotometer
- Certified Holmium Oxide Filter or Solution
- 0.005 M Potassium Dichromate Solution
- Distilled Water (Blank)
- Quartz Cuvettes
- Validated calibration logbook (Annexure-1)
5.3 Calibration Parameters and Acceptance Criteria
- Wavelength Accuracy: Use holmium oxide filter/solution. Acceptable variation: ±1 nm from certified value.
- Absorbance Accuracy: Prepare 0.005 M potassium dichromate and measure at 235 nm, 257 nm, 313 nm, and 350 nm. Compare with acceptance limits as per pharmacopeial standards.
- Stray Light: Use potassium chloride solution. Absorbance at 200 nm should be ≥2.0.
- Baseline Flatness: Scan 200–400 nm with distilled water. Fluctuation in absorbance should not exceed ±0.01 AU.
5.4 Calibration Steps
- Switch ON the UV-Visible spectrophotometer and allow warm-up for 30 minutes.
- Clean cuvettes and ensure they are free from scratches and fingerprints.
- Zero the instrument with blank (distilled water).
- Carry out wavelength calibration using the certified holmium oxide standard.
- Record the peak maxima and compare with certified values.
- Prepare and measure absorbance of potassium dichromate solution at specified wavelengths.
- Record stray light absorbance using 1% potassium chloride at 200 nm.
- Run a full scan for baseline flatness from 200 to 400 nm with blank cuvette.
- Record all results in the instrument calibration logbook.
5.5 Documentation
- Fill Annexure-1 with all calibration readings and analyst signature.
- Submit for QA verification and sign-off.
- Paste one copy in the instrument area and another in the QA documentation room.
5.6 Out of Calibration (OOC) Handling
- If any parameter fails, immediately label instrument as “OUT OF ORDER”.
- Initiate deviation form and perform root cause analysis.
- Perform recalibration after corrective action and document accordingly.
6. Abbreviations
- QA: Quality Assurance
- OOC: Out of Calibration
- AU: Absorbance Unit
- nm: Nanometer
7. Documents
- Annexure-1: UV-Visible Spectrophotometer Calibration Log
- Annexure-2: Deviation Form Template
8. References
- USP General Chapter <857> – Ultraviolet-Visible Spectroscopy
- ICH Q2(R1) – Validation of Analytical Procedures
- Instrument Operation Manual
9. SOP Version
Version: 2.0
10. Approval Section
| Prepared By | Checked By | Approved By | |
|---|---|---|---|
| Signature | |||
| Date | |||
| Name | Sanjana Iyer | Rahul Deshmukh | Dr. Smita Narula |
| Designation | Analyst | QA Reviewer | Head – AMD |
| Department | Analytical Method Development | Quality Assurance | Analytical Method Development |
11. Annexures
Annexure-1: UV-Visible Spectrophotometer Calibration Log
| Date | Wavelength Accuracy | Absorbance Accuracy | Stray Light | Baseline Flatness | Remarks |
|---|---|---|---|---|---|
| 01/06/2025 | Within ±1 nm | Pass | 2.2 AU | ±0.008 AU | All parameters within limits |
Annexure-2: Deviation Form Template
This form captures date, deviation ID, description, probable cause, corrective action, and QA comments.
Revision History:
| Revision Date | Revision No. | Details | Reason | Approved By |
|---|---|---|---|---|
| 01/06/2025 | 2.0 | Stray light and baseline flatness sections revised | Annual review and regulatory update | Dr. Smita Narula |
| 15/07/2022 | 1.0 | Initial Release | New SOP | QA Head |