SOP Guide for Pharma

Analytical Method Development: SOP for Accuracy Assessment Using Recovery Study – V 2.0

Analytical Method Development: SOP for Accuracy Assessment Using Recovery Study – V 2.0

Standard Operating Procedure for Accuracy Assessment via Recovery Study in Analytical Method Development


Department Analytical Method Development
SOP No. SOP/AMD/369/2025
Supersedes SOP/AMD/369/2022
Page No. Page 1 of 13
Issue Date 01/06/2025
Effective Date 04/06/2025
Review Date 01/06/2027

1. Purpose

This Standard Operating Procedure (SOP) describes the procedure for evaluating the accuracy of an analytical method using recovery studies, which involves the addition of known amounts of analyte to a

sample matrix and calculating the percentage recovered. Accuracy is a key performance parameter during analytical method validation, ensuring the method’s ability to measure the true value of an analyte across a defined range.

2. Scope

This procedure is applicable to all new and existing analytical methods developed or validated for quantitative estimation of active pharmaceutical ingredients (API), excipients, finished dosage forms, and raw materials in the Analytical Method Development (AMD) laboratory.

3. Responsibilities

  • Analytical Chemist: Prepares and analyzes accuracy solutions, records data, and performs calculations.
  • AMD Scientist: Reviews and verifies raw data and calculations, evaluates accuracy results.
  • Quality Assurance (QA): Verifies final report and ensures protocol adherence before approval.

4. Accountability

The Head of Analytical Method Development is accountable for implementation, compliance, documentation, and archiving of accuracy validation reports in accordance with ICH, WHO, and national regulatory standards.

5. Procedure

5.1 Preparation of Accuracy Samples

  1. Select three levels of concentration—typically 50%, 100%, and 150% of target assay concentration.
  2. Prepare placebo or blank matrix equivalent to the sample matrix used in routine analysis.
  3. Spike the placebo with known quantities of standard (API) to obtain final concentrations at each level in triplicate.

5.2 Sample Analysis

  1. Analyze the accuracy samples using the validated method or the method under development.
  2. Inject all replicates into the instrument under the same chromatographic/analytical conditions.
  3. Document the retention time, area, and calculated concentration of each injection.

5.3 Calculation and Evaluation

  1. Calculate the % recovery using the formula:
    % Recovery = (Measured amount / Added amount) × 100
  2. Calculate mean recovery and % relative standard deviation (%RSD) for each level.
  3. Acceptance Criteria: 98.0%–102.0% for assay methods (as per ICH), with %RSD ≤ 2.0% unless otherwise justified.

5.4 Documentation

  1. Enter all raw data into the Accuracy Recovery Summary Sheet (Annexure-1).
  2. Attach chromatograms, calibration curve (if applicable), and sample preparation details.
  3. Compile the accuracy section of the method validation report as per the AMD format.

6. Abbreviations

  • API: Active Pharmaceutical Ingredient
  • RSD: Relative Standard Deviation
  • QA: Quality Assurance
  • AMD: Analytical Method Development

7. Documents

  1. Annexure-1: Accuracy Recovery Summary Sheet
  2. Annexure-2: Sample Spiking Records
  3. Method Validation Protocol and Report

8. References

  • ICH Q2(R1): Validation of Analytical Procedures
  • USP General Chapter <1225>: Validation of Compendial Procedures
  • WHO TRS No. 996, Annex 7: Guidelines on Validation
  • 21 CFR Part 211 – cGMP Guidelines

9. SOP Version

Version: 2.0

10. Approval Section

Prepared By Checked By Approved By
Signature
Date
Name Arvind Patil Megha Shah Dr. R.K. Joshi
Designation Analyst QA Officer Head – AMD
Department Analytical Method Development Quality Assurance Analytical Method Development

11. Annexures

Annexure-1: Accuracy Recovery Summary Sheet

Level Added (%) Recovered (%) Mean Recovery % RSD
Low (50%) 50.00 49.85 99.7% 0.22%
Mid (100%) 100.00 100.2 100.2% 0.18%
High (150%) 150.00 149.5 99.7% 0.25%

Conclusion: Accuracy results within acceptable limits across all levels. Method is accurate.

Annexure-2: Sample Spiking Records

Documented preparation of placebo, spike level calculations, and analyst signatures.

Revision History

Revision Date Revision No. Description Reason Approved By
01/06/2025 2.0 Enhanced acceptance criteria and updated annexures Annual Review Dr. R.K. Joshi
10/07/2022 1.0 Initial SOP Issuance New SOP QA Head