Analytical Method Development: Peak Purity Evaluation using PDA – V 2.0
SOP for Peak Purity Assessment Using PDA Detector in HPLC Method Development
Department |
Analytical Method Development |
SOP No. |
SOP/AMD/073/2025 |
Supersedes |
SOP/AMD/073/2022 |
Page No. |
Page 1 of 14 |
Issue Date |
19/05/2025 |
Effective Date |
20/05/2025 |
Review Date |
19/05/2026 |
1. Purpose
This SOP describes the procedure for evaluating peak purity using a Photodiode Array (PDA) detector during HPLC method development. Peak purity assessment is essential to confirm that an analyte peak is not co-eluted with impurities or degradants and that it represents a single chemical entity.
2. Scope
This SOP applies to the Analytical Method Development (AMD) department and is applicable to the evaluation of drug substances, finished dosage forms, degradation studies, and stability samples using PDA-equipped HPLC systems.
3. Responsibilities
- Analytical Chemist: Performs PDA data acquisition, spectral comparison, and documentation of purity results.
- Reviewer: Validates spectral homogeneity and confirms peak purity match parameters are within acceptance criteria.
- QA Officer: Ensures that chromatographic data including PDA spectra and reports meet regulatory and data integrity requirements.
- Head – AMD: Reviews and approves the finalized peak purity reports and method suitability for validation.
4. Accountability
The Head of Analytical Method Development is accountable for verifying that peak purity results are scientifically sound and used to support method specificity and selectivity claims.
5. Procedure
5.1 Instrument and Software Requirements
- Use a qualified HPLC system equipped with a PDA detector and validated chromatography software (e.g., Empower, Chromeleon).
- Ensure baseline noise and drift are within specification before sample analysis.
- Verify detector calibration using holmium oxide filter or certified wavelength standards.
5.2 Sample Preparation
- Prepare standard API solution and test sample in appropriate diluent at a concentration that ensures optimal peak intensity without saturation.
- Filter through 0.45 µm or 0.22 µm PVDF or nylon syringe filters.
- Document all preparations in Annexure-1: Sample Preparation Log.
5.3 Chromatographic Conditions
- Inject blank, standard, and sample under the selected method conditions.
- Set PDA acquisition parameters:
- Wavelength range: 200–400 nm
- Sampling rate: ≥ 10 Hz
- Resolution: 1.2 nm or better
- Ensure the PDA trace is enabled and active on chromatographic method file.
- Perform triplicate injections for peak reproducibility and purity confirmation.
5.4 Spectral Evaluation
- Use chromatography software to compare UV spectra across:
- Peak apex
- Peak leading edge (start)
- Peak trailing edge (end)
- Calculate match angle or similarity index using software algorithm (e.g., Apex vs Start and Apex vs End).
- Acceptance criteria:
- Purity angle must be less than purity threshold
- Similarity index > 990 (or > 99%) depending on software
- No major spectral deviation across the peak
- Document observations in Annexure-2: Peak Purity Evaluation Log.
5.5 Interference and Co-Elution Check
- Overlay UV spectra of the analyte peak and known impurities.
- Inject placebo and degradation samples to verify spectral uniqueness.
- Confirm that no shoulders, split peaks, or non-matching spectra exist under method conditions.
- Document and attach PDA spectral overlays in Annexure-3: PDA Overlay Report.
5.6 Reporting and Interpretation
- If peak passes spectral similarity test and purity angle is below threshold, report as “Spectrally Pure”.
- If results fail, investigate possible co-elution, matrix interference, or system suitability issues.
- Use peak purity result to support specificity data in validation report or regulatory filing.
- Final interpretation to be entered in Annexure-4: Peak Purity Certificate.
6. Abbreviations
- PDA: Photodiode Array
- HPLC: High-Performance Liquid Chromatography
- API: Active Pharmaceutical Ingredient
- UV: Ultraviolet
- SOP: Standard Operating Procedure
7. Documents
- Sample Preparation Log – Annexure-1
- Peak Purity Evaluation Log – Annexure-2
- PDA Overlay Report – Annexure-3
- Peak Purity Certificate – Annexure-4
8. References
- ICH Q2(R1) – Validation of Analytical Procedures
- USP General Chapter <621> – Chromatography
- FDA Guidance on Analytical Procedures and Method Validation
- Software Manuals for Empower / Chromeleon / OpenLab
9. SOP Version
Version: 2.0
10. Approval Section
|
Prepared By |
Checked By |
Approved By |
Signature |
|
|
|
Date |
|
|
|
Name |
|
|
|
Designation |
|
|
|
Department |
|
|
|
11. Annexures
Annexure-1: Sample Preparation Log
Sample Code |
Concentration |
Diluent |
Filtered |
Prepared By |
XYZ-API |
100 µg/mL |
Mobile Phase |
Yes |
Sunita Reddy |
Annexure-2: Peak Purity Evaluation Log
Injection |
Match Angle |
Purity Threshold |
Result |
Status |
STD-01 |
2.45 |
3.10 |
Pass |
Pure |
Annexure-3: PDA Overlay Report
Overlay ID |
Comparison Type |
Observation |
Status |
PDA-001 |
Apex vs End |
Similarity: 99.5% |
Pass |
Annexure-4: Peak Purity Certificate
Sample ID |
Peak RT (min) |
Detector |
Status |
Analyst |
XYZ-API |
3.24 |
PDA |
Spectrally Pure |
Rajesh Kumar |
Revision History:
Revision Date |
Revision No. |
Details |
Reason |
Approved By |
04/05/2025 |
2.0 |
Added overlay comparison and PDA threshold parameters |
Annual Review Update |
|