Analytical Method Development: Conductivity Method Development – V 2.0

Posted on By

Analytical Method Development: Conductivity Method Development – V 2.0

SOP for Development of Conductivity-Based Analytical Methods in AMD Laboratory


Department Analytical Method Development
SOP No. SOP/AMD/066/2025
Supersedes SOP/AMD/066/2022
Page No. Page 1 of 14
Issue Date 19/05/2025
Effective Date 20/05/2025
Review Date 19/05/2026

1. Purpose

This SOP defines the systematic procedure for the development and validation of analytical methods based on electrical conductivity. These methods are used for evaluating the ionic strength, purity, and quality of pharmaceutical substances and formulations.

2. Scope

This SOP is applicable to the Analytical Method Development (AMD) department for both qualitative and quantitative analysis of electrolytes, ionic excipients, and API solutions. It is especially applicable for water analysis, buffer systems, and ionic APIs.

3. Responsibilities

  • Analytical Chemist: Responsible for conducting method development, performing measurements, documenting results, and preparing validation reports.
  • Reviewer: Ensures test data integrity, verifies instrument calibration, and checks report completeness.
  • QA Officer: Reviews compliance with GMP and regulatory requirements.
  • Head – AMD: Approves the validated conductivity method for routine and regulatory use.

4. Accountability

The Head of Analytical Method Development is accountable for accuracy, regulatory compliance, and timely approval of the developed method.

See also  Analytical Method Development: Gravimetric Method Development for Purity - V 2.0

5. Procedure

5.1 Principle of Conductivity Measurement

  1. Conductivity is the ability of a solution to conduct electric current, which is directly proportional to the concentration and mobility of ions present.
  2. Measurements are expressed in microSiemens per centimeter (µS/cm) or milliSiemens per centimeter (mS/cm).
  3. Conductivity readings vary with temperature and must be corrected using a temperature compensation factor or by measuring at standard conditions (25°C).

5.2 Instrumentation and Calibration

  1. Use a calibrated conductivity meter with digital temperature compensation, preferably equipped with:
    • Temperature sensor (integrated or external)
    • Glass or graphite conductivity cell (cell constant known)
  2. Calibrate the instrument daily using certified standard solutions (e.g., 84 µS/cm, 1413 µS/cm, 12.88 mS/cm).
  3. Record instrument ID, calibration date, and standard lot number in Annexure-1: Instrument Calibration Log.

5.3 Sample Preparation

  1. Prepare solutions using conductivity-grade or deionized water.
  2. Avoid use of glassware previously exposed to ionic detergents or residues.
  3. Degas the sample if needed to remove CO₂ interference.
  4. Ensure temperature equilibration before measurement.
  5. Document preparation and any dilution factor in Annexure-2: Sample Preparation Log.

5.4 Measurement Procedure

  1. Turn on the conductivity meter and allow to stabilize as per manufacturer’s guidelines.
  2. Rinse probe with deionized water and blot dry with lint-free tissue.
  3. Immerse the probe into the sample ensuring adequate coverage of electrodes.
  4. Allow the temperature reading to stabilize at 25°C before recording the conductivity.
  5. Repeat the measurement in triplicate and calculate the average.
  6. Record all readings in Annexure-3: Conductivity Observation Sheet.
See also  Analytical Method Development: Risk Assessment in Analytical Method Development - V 2.0

5.5 Calculation

Average Conductivity (µS/cm) = (Reading 1 + Reading 2 + Reading 3) / 3

If required, apply dilution correction and report in terms of original concentration.

5.6 Method Optimization

  1. Evaluate:
    • Effect of temperature variation on conductivity
    • Stability of sample over time
    • Concentration-dependent linearity
  2. Develop calibration curve using standard solutions to verify instrument response.
  3. Document optimization studies in Annexure-4: Method Optimization Record.

5.7 Method Validation

  1. Specificity: Confirm that only ionic species influence conductivity readings.
  2. Linearity: Validate from low to high concentrations (R² ≥ 0.999).
  3. Accuracy: Perform spike recovery studies at 80%, 100%, and 120% concentration levels.
  4. Precision: Assess repeatability and intermediate precision (RSD ≤ 2.0%).
  5. Robustness: Evaluate variation in sample volume, temperature ±2°C, and electrode rinsing method.
  6. Summarize all findings in Annexure-5: Validation Summary Sheet.

6. Abbreviations

  • SOP: Standard Operating Procedure
  • AMD: Analytical Method Development
  • RSD: Relative Standard Deviation
  • µS/cm: MicroSiemens per centimeter
  • mS/cm: MilliSiemens per centimeter

7. Documents

  1. Instrument Calibration Log – Annexure-1
  2. Sample Preparation Log – Annexure-2
  3. Conductivity Observation Sheet – Annexure-3
  4. Method Optimization Record – Annexure-4
  5. Validation Summary Sheet – Annexure-5
See also  Analytical Method Development: Total Ash and Acid Insoluble Ash Method Development - V 2.0

8. References

  • USP <645> – Water Conductivity
  • EP 2.2.38 – Conductivity
  • ICH Q2(R1) – Validation of Analytical Procedures
  • IS 3025 (Part 14): Method of Measurement of Conductivity

9. SOP Version

Version: 2.0

10. Approval Section

Prepared By Checked By Approved By
Signature
Date
Name
Designation
Department

11. Annexures

Annexure-1: Instrument Calibration Log

Instrument ID Standard Used Measured Value Acceptance Limit Status Checked By
COND-021 1413 µS/cm 1412 µS/cm 1410–1416 Pass Sunita Reddy

Annexure-2: Sample Preparation Log

Sample Code Solvent Dilution Temp Prepared By
API-2025 Water 1:5 25°C Rajesh Kumar

Annexure-3: Conductivity Observation Sheet

Trial Conductivity (µS/cm) Temperature Average Status
1 142.3 25.1°C 142.4 Pass
2 142.5 24.9°C
3 142.4 25.0°C

Annexure-4: Method Optimization Record

Parameter Variation Observation Conclusion
Temperature ±2°C ±3.5% change in reading Use temperature compensation

Annexure-5: Validation Summary Sheet

Parameter Criteria Result Status
Accuracy 98–102% 99.1% Pass
Precision RSD ≤ 2.0% 0.85% Pass
Linearity R² ≥ 0.999 0.9993 Pass

Revision History:

Revision Date Revision No. Details Reason Approved By
04/05/2025 2.0 Expanded optimization and validation for high-conductivity buffers Annual Review
Analytical Method Development V 2.0 Tags:Analytical Method Development SOP, SOP for analytical calculations, SOP for analytical documentation, SOP for analytical instrument calibration, SOP for analytical instrument qualification, SOP for analytical method change control, SOP for analytical method deviation handling, SOP for analytical method documentation practices, SOP for analytical method lifecycle management, SOP for analytical method optimization, SOP for analytical method risk assessment, SOP for analytical method training and competency, SOP for analytical method transfer protocol, SOP for analytical method validation, SOP for forced degradation studies, SOP for HPLC method development, SOP for limit of detection (LOD) determination, SOP for limit of quantitation (LOQ) determination, SOP for method accuracy testing, SOP for method linearity assessment, SOP for method precision evaluation, SOP for method reproducibility, SOP for method robustness testing, SOP for method transfer, SOP for method verification, SOP for qualification of working standards, SOP for reference standard handling, SOP for specificity testing in analytical methods, SOP for stability indicating methods, SOP for system suitability testing

Post navigation

Previous Post: Aerosol: SOP for Verifying Utility System Performance During Process Changes – V 2.0
Next Post: Biosimilars: SOP for Inoculation of Bioreactors – V 2.0

Standard Operating Procedures V 1.0

  • Aerosols
  • Analytical Method Development
  • Bioequivalence Bioavailability Study
  • Capsule Formulation
  • Clinical Studies
  • Creams
  • Data Integrity
  • Dental Dosage Forms
  • Drug Discovery
  • Environment, Health and Safety
  • Formulation Development
  • Gels
  • Good Distribution Practice
  • Good Warehousing Practices
  • In-Process Control
  • Injectables
  • Liquid Orals
  • Liposome and Emulsion Formulations
  • Lotions
  • Lyophilized Products
  • Maintenance Dept.
  • Medical Devices
  • Metered-Dose Inhaler
  • Microbiology Testing
  • Nanoparticle Formulation
  • Nasal Spray Formulations
  • Nebulizers
  • Ocular (Eye) Dosage Forms
  • Ointments
  • Otic (Ear) Dosage Forms
  • Pharmacovigilance
  • Powder & Granules
  • Purchase Departments
  • Quality Assurance
  • Quality Control
  • Raw Material Stores
  • Regulatory Affairs
  • Tablet Manufacturing
  • Rectal Dosage Forms
  • Transdermal Patches
  • Vaginal Dosage Forms
  • Validations and Qualifications

Read SOPs in your Language:

 - 
Bengali
 - 
bn
English
 - 
en
Gujarati
 - 
gu
Hindi
 - 
hi
Malayalam
 - 
ml
Marathi
 - 
mr
Punjabi
 - 
pa
Tamil
 - 
ta
Telugu
 - 
te

NEW! Revised SOPs – V 2.0

  • Aerosols V 2.0
  • Analytical Method Development V 2.0
  • API Manufacturing V 2.0
  • BA-BE Studies V 2.0
  • Biosimilars V 2.0
  • Capsules V 2.0
  • Creams V 2.0
  • Elixers V 2.0
  • Ointments V 2.0
  • Raw Material Warehouse V 2.0
  • Tablet Manufacturing V2.0

New Publication: A must for All.

Copyright © 2025 SOP Guide for Pharma.

Powered by PressBook WordPress theme

Go to mobile version