Analytical Method Development: Colorimetric Method Development – V 2.0
SOP for Colorimetric Method Development in Analytical Method Development Laboratory
Department |
Analytical Method Development |
SOP No. |
SOP/AMD/051/2025 |
Supersedes |
SOP/AMD/051/2022 |
Page No. |
Page 1 of 14 |
Issue Date |
19/05/2025 |
Effective Date |
20/05/2025 |
Review Date |
19/05/2026 |
1. Purpose
The purpose of this SOP is to outline the standardized approach for developing, optimizing, and validating colorimetric methods for the qualitative and quantitative determination of analytes in pharmaceutical substances.
Colorimetric methods use specific reagents to develop color based on chemical reactions, which is then measured spectrophotometrically to infer concentration.
2. Scope
This SOP is applicable to the Analytical Method Development (AMD) laboratory for developing colorimetric procedures for analysis of APIs, excipients, intermediates, and dosage forms where applicable, in compliance with pharmacopeial and ICH guidelines.
3. Responsibilities
- Analytical Chemist: Responsible for developing and executing the method, selecting reagents, measuring absorbance, and recording observations.
- Method Reviewer: Validates the accuracy and repeatability of the developed method, verifies absorbance readings and calculations.
- QA: Ensures the method and documentation comply with GMP, ICH, and pharmacopeial requirements.
- Head – AMD: Provides approval for final method release and oversees its incorporation into regulatory documents.
4. Accountability
The Head of the Analytical Method Development department is accountable for ensuring the suitability, reproducibility, and regulatory compliance of all colorimetric methods developed for pharmaceutical testing.
5. Procedure
5.1 Preliminary Assessment
- Review available pharmacopeial and scientific literature for existing colorimetric procedures relevant to the analyte or chemical class.
- Assess the chemical functionality of the target analyte to identify potential reagents that produce color upon reaction.
- Determine the purpose of the colorimetric method:
- Limit test (qualitative)
- Assay or content determination (quantitative)
- Record findings and justification in Annexure-1: Preliminary Method Review Sheet.
5.2 Reagent Selection and Reaction Optimization
- Identify reagents known to form chromogenic complexes with the target analyte. Examples include:
- 1,10-phenanthroline for iron
- Nessler’s reagent for ammonia
- Diphenylcarbazone for mercury
- Prepare multiple test solutions with varying concentrations of analyte and reagent under controlled conditions.
- Determine the optimal reagent-to-analyte molar ratio based on the intensity of color and reaction stability.
- Evaluate different reaction times, pH values, and buffer systems.
- Measure absorbance at multiple wavelengths to identify λmax.
- Document the optimized reaction conditions in Annexure-2: Optimization Worksheet.
5.3 Standard Preparation and Calibration Curve
- Prepare standard stock solutions of the analyte using a certified reference material (CRM).
- Prepare a series of dilutions (5–7 points) spanning the expected concentration range.
- React each dilution with the color reagent under optimized conditions.
- Measure absorbance at the determined λmax using a UV-Vis spectrophotometer.
- Plot the absorbance vs concentration to generate the calibration curve.
- Ensure linearity with correlation coefficient (R²) ≥ 0.999.
- Record data in Annexure-3: Calibration Curve Log.
5.4 Sample Analysis
- Prepare test samples according to formulation or API-specific instructions (e.g., dissolution, dilution, filtration).
- Treat test samples with the same reagents and under identical conditions as standards.
- Measure the absorbance and determine analyte concentration using the calibration curve.
- Repeat analysis in triplicate to ensure precision.
- Document findings in Annexure-4: Sample Analysis Record.
5.5 Method Validation
- Conduct validation as per ICH Q2(R1):
- Specificity: Demonstrate that color development is analyte-specific and unaffected by excipients or solvents.
- Linearity: Confirm R² ≥ 0.999 over working range.
- Accuracy: Perform spike recovery studies at 80%, 100%, and 120% levels (acceptable recovery 98–102%).
- Precision: Determine intra-day and inter-day variability (RSD ≤ 2.0%).
- LOD & LOQ: Calculate based on signal-to-noise ratio of 3.3 and 10 respectively.
- Robustness: Vary reaction time, reagent concentration, and pH to observe stability of absorbance.
- Solution Stability: Test absorbance at intervals over 6 hours.
- Record validation data in Annexure-5: Validation Summary Report.
5.6 Final Method Documentation
- Prepare detailed method SOP including:
- Objective and applicability
- Reagent composition and preparation
- Sample preparation instructions
- Reaction procedure
- Instrumentation details
- System suitability and acceptance criteria
- Submit method to QA for review and approval prior to official use.
6. Abbreviations
- SOP: Standard Operating Procedure
- CRM: Certified Reference Material
- UV-Vis: Ultraviolet-Visible Spectrophotometry
- ICH: International Council for Harmonisation
- RSD: Relative Standard Deviation
7. Documents
- Preliminary Method Review Sheet – Annexure-1
- Optimization Worksheet – Annexure-2
- Calibration Curve Log – Annexure-3
- Sample Analysis Record – Annexure-4
- Validation Summary Report – Annexure-5
8. References
- USP <851> – Spectrophotometry and Light Scattering
- ICH Q2(R1) – Validation of Analytical Procedures
- IP, Ph. Eur., BP – General Colorimetric Tests
- AOAC Official Methods for Colorimetric Analysis
9. SOP Version
Version: 2.0
10. Approval Section
|
Prepared By |
Checked By |
Approved By |
Signature |
|
|
|
Date |
|
|
|
Name |
|
|
|
Designation |
|
|
|
Department |
|
|
|
11. Annexures
Annexure-1: Preliminary Method Review Sheet
Analyte |
Literature Source |
Color Reagent |
Remarks |
Prepared By |
Iron (III) |
USP |
1,10-Phenanthroline |
Forms orange-red complex |
Rajesh Kumar |
Annexure-2: Optimization Worksheet
Trial |
Reagent Volume (mL) |
pH |
λmax (nm) |
Observation |
1 |
2.0 |
4.5 |
510 |
Stable complex |
Annexure-3: Calibration Curve Log
Concentration (ppm) |
Absorbance |
5 |
0.135 |
10 |
0.270 |
20 |
0.540 |
30 |
0.810 |
40 |
1.080 |
Annexure-4: Sample Analysis Record
Sample |
Absorbance |
Calculated Conc. (ppm) |
Analyst |
API-A |
0.540 |
20.0 |
Sunita Reddy |
Annexure-5: Validation Summary Report
Parameter |
Criteria |
Result |
Status |
Precision |
RSD ≤ 2% |
1.2% |
Pass |
Accuracy |
98–102% |
99.6% |
Pass |
Linearity |
R² ≥ 0.999 |
0.9995 |
Pass |
Revision History:
Revision Date |
Revision No. |
Details |
Reason |
Approved By |
04/05/2025 |
2.0 |
Expanded validation, annexures and buffer optimization data |
Audit Preparedness |
|